Peter Huber Kältemaschinenbau Unichiller P100w-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P100w-H

Peter Huber Kältemaschinenbau Unichiller P100w-H — Recirculating chillers with Noise level 58 dB.

The Unichiller P100w-H is a water-cooled recirculating chiller with a 2 kW heater, a fixed-speed E2 pump and a 48 l expansion volume, rated for -20 to 100 °C. Its 'P' designation marks it for high-pressure-drop process loops, and its Pilot ONE controller provides Ethernet, USB and RS232 interfaces. It serves closed secondary circuits requiring both cooling and active heating.

Noise level
58 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller P100w-H Current 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Noise level 58 dB(A)
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E2
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 321 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 240 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 270 л/час · at 15°C, flow 0°C: 330 l/h
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be sized for 15.5 A and fitted with a site plug

The 400 V three-phase supply needs a 3x16 A protection device and must tolerate ±5% voltage only with ±2% frequency, with certain combinations prohibited. The unit arrives with a cable but no plug, so a matching plug and connection point must be provided on site before commissioning.

Unichiller P100w-H Datasheet, page 2
Cooling & Media
Cooling water must stay between 5 and 20 °C at the inlet

Inlet water below the condensation point requires insulation of the internal cooling water pipes. Warmer water reduces cooling power and may increase the required flow rate, so the circuit must be designed for the actual site water temperature rather than nominal conditions.

Unichiller P100w-H Datasheet, page 2
Cooling & Media
Cooling water pressure must not exceed 6 bar

The G1/2 male connection requires a minimum differential pressure of 0.5 bar and tolerates a maximum of 6 bar. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only water compatible with these materials may be used to avoid corrosion and fouling.

Unichiller P100w-H Datasheet, page 2
Process Limits
Only externally closed circuits are permitted

The unit is designed for externally closed applications. The 'P' model designation indicates suitability for high pressure drop, but the system pressure drop must still be checked against the pump curve (130 l/min at 5.7 bar falling to 20 l/min at 5.5 bar).

Unichiller Operation Manual, page 34
Mounting
Leveling feet and parking brakes must be set before start-up

The 321 kg unit must be secured by unscrewing/activating the leveling feet and applying caster parking brakes before operation. A wall clearance of 50 to 200 cm per side is required for maintenance, and electrical lines directly below the unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, pages 25, 28, 31
Control & I/O
Power-failure response must be chosen for the process

The Pilot ONE can be configured for OFF/Standby, ON, or Power Failure Auto after a power outage. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge (GWP 1397) and room size must be checked at installation to stay within the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, pages 15, 17
Operation
Shut off cooling water even for overnight downtime

The cooling water supply must be closed even for short downtimes such as overnight to prevent water damage and corrosion during idle periods.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the P100w-H instead of the air-cooled Unichiller 030T or 040T siblings?

Choose the P100w-H when a water-cooled unit is acceptable and the process loop has high pressure drop, since the 'P' designation targets such applications. The 030T/040T are air-cooled with 69 dB(A) noise versus 58 dB(A) here, so site water availability and noise limits decide.

What cooling water supply and quality does the P100w-H require from adjacent plant?

A G1/2 male connection delivering 5–20 °C water at 0.5–6 bar differential pressure. Consumption is 330 l/h at 0 °C, 270 l/h at -10 °C and 240 l/h at -20 °C flow temperature. Water must be compatible with copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM.

What installation environment and clearances does the manual require?

Indoor use only, 5–40 °C ambient, IP20, illuminance ≥300 lx, altitude ≤2000 m, and no ATEX zone. Allow 50–200 cm wall clearance per side for maintenance, set leveling feet and parking brakes, and protect electrical lines below the unit from condensation.

What is not included in base delivery and what maintenance obligations apply?

The three-phase cable is supplied without a plug, so a site plug is needed. Refrigerant circuit work must be done by approved refrigeration contractors; the R-449A charge and room volume must be checked at installation. Cooling water must be shut off even overnight.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P100w-H Datasheet
Datasheet ·92,3 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF