Peter Huber Kältemaschinenbau Unistat 510w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 510w Dynamisches

Peter Huber Kältemaschinenbau Unistat 510w Dynamisches — Temperature control units with Noise level 60 dB.

The Unistat 510w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 250 °C. Within the Unistat 5 series it is the base 510w modification, distinguished from the 515w/520w/525w siblings by its 6 kW heating capacity and 60 dB(A) noise level. It serves precise temperature control of externally closed or open circuits in laboratory and pilot-scale processes.

Noise level
60 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 510w Dynamisches Current 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Control output POKO ONE
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...250 °C
Dimensions, Weight & Mechanical Design
Weight 180 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 355 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential pressure

The cooling water circuit requires an inlet temperature between 5 and 20 °C, at least 1 bar differential pressure and no more than 6 bar. Consumption is 355 l/h at 15 °C water with 0 °C flow. Pipes must be insulated if water temperature can drop below the dew point to prevent condensation. Wetted materials (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) constrain water quality.

Unistat 510w Technical Data, page 2
Process Limits
Cooling capacity collapses to 0.25 kW at -50 °C

Although the range extends to -50 °C, available cooling power falls from 5.3 kW at 0 °C with ethanol to 0.25 kW at -50 °C. Low-temperature duties must be sized against this curve, and the external circuit pressure drop must be checked against the pump curve (max. 112 l/min, 54 l/min at 1.0 bar).

Unistat 510w Technical Data, page 1
Mounting
Transport upright only and allow 2 h acclimation before power-on

The unit must be transported and stored upright; horizontal transport damages the compressor. Only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically. Leveling feet must be activated and parking brakes engaged before commissioning. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults.

Unistat Operation Manual, page 22 and page 29
Maintenance
Shut off cooling water overnight and inspect hoses periodically

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and hose connections must be periodically checked for leaks and quality and replaced as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.

Unistat Operation Manual, page 29; Unistat Operation Manual, page 19

Selection & Integration Q&A

When should I choose the Unistat 510w over the 515w, 520w or 525w siblings?

Pick the 510w when 6 kW heating and the -50…250 °C range with 5.3 kW cooling at 0 °C cover the duty. Move to 515w/520w/525w when higher heating or cooling capacity is needed; note 520w/525w are 61 dB(A) and 525 is 56 dB(A), so noise also differs.

What cooling water supply and connections does the 510w require from adjacent plant?

G1/2 male connection, inlet water 5–20 °C, 1–6 bar differential pressure, 355 l/h at 15 °C water/0 °C flow. Insulate pipes if below dew point. Wetted materials include copper, 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible.

What are the installation environment and duty-cycle limits from the manual?

Ambient 5–40 °C, IP20 dry indoor only, not in an ATEX zone. Transport upright and emptied; allow ~2 h acclimation after cold/warm transfer. Leveling feet and parking brakes must be set. Shut off cooling water overnight. Process Safety mode requires compressor Permanently ON.

What maintenance and accessories are not included in base delivery?

No plug is supplied for the 400 V three-phase cable (devices below 63 A). A drip tray for condensation/heat transfer fluid must be provided by the installer. Hoses and adapters for M30x1.5 pump and G1/2 cooling water are not included. Periodic hose inspection and cooling-water shutoff are required.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 510w Technical Data
Datasheet ·83,8 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF