Peter Huber Kältemaschinenbau Unistat 510w Dynamisches
Peter Huber Kältemaschinenbau Unistat 510w Dynamisches — Temperature control units with Noise level 60 dB.
The Unistat 510w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 250 °C. Within the Unistat 5 series it is the base 510w modification, distinguished from the 515w/520w/525w siblings by its 6 kW heating capacity and 60 dB(A) noise level. It serves precise temperature control of externally closed or open circuits in laboratory and pilot-scale processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 510w Dynamisches Current | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 60 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...250 °C |
| Weight | 180 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 355 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (w) free-standing Unistat 5-series unit; magnetically coupled stainless-steel pump with M30x1.5 male connection and adjustable speed.
- Pilot ONE controller with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB (Host and Device) and RS232; external Pt100 sensor input.
- Safety class III / FL; IP20 protection; 180 kg; 400 V 3~ 50 Hz supply.
- Within the series, the 510w is the lowest-capacity water-cooled 5-series model; siblings 515w, 520w and 525w offer higher performance (520w/525w at 61 dB(A), 525 at 56 dB(A)).
Behavior under load and media
- Working range -50 to 250 °C. Cooling power falls steeply at low temperature: 5.3 kW at 0 °C with ethanol down to 0.25 kW at -50 °C, so low-temperature duties must be sized against the curve, not the nominal rating.
- Heating capacity 6 kW. Pump delivers max. 112 l/min and 54 l/min at 1.0 bar; external circuit pressure drop must be matched to this curve.
- Cooling water circuit: G1/2 male connection, 355 l/h consumption at 15 °C water/0 °C flow, minimum 1 bar differential pressure, maximum 6 bar; inlet water 5–20 °C. Wetted materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Refrigerant R-452A (A1, H280), charge equivalent to 3.1 t CO2.
Installation, control and operation
- Requires 400 V three-phase supply with 3x16 A fuse; delivered with cable but without plug (three-phase devices below 63 A). Leakage current >3.5 mA, so protective earthing and RCD selection must account for it.
- Voltage tolerance ±5% and frequency ±2%, not to be combined in the unfavorable direction.
- Ambient 5–40 °C; IP20 requires dry indoor location. Unit must not be installed in an ATEX zone. Transport and store upright only; allow ~2 h acclimation after cold/warm transfer before power-on.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications. Process Safety mode requires compressor set to Permanently ON. Power Failure Auto can resume temperature control automatically after outage.
- Cooling water pipes must be insulated if water temperature can fall below dew point. Shut off cooling water even for overnight downtimes.
Integration & Operating Notes
The cooling water circuit requires an inlet temperature between 5 and 20 °C, at least 1 bar differential pressure and no more than 6 bar. Consumption is 355 l/h at 15 °C water with 0 °C flow. Pipes must be insulated if water temperature can drop below the dew point to prevent condensation. Wetted materials (copper, 1.4401, MS, PA, PPE, PTFE, EPDM) constrain water quality.
Although the range extends to -50 °C, available cooling power falls from 5.3 kW at 0 °C with ethanol to 0.25 kW at -50 °C. Low-temperature duties must be sized against this curve, and the external circuit pressure drop must be checked against the pump curve (max. 112 l/min, 54 l/min at 1.0 bar).
The unit must be transported and stored upright; horizontal transport damages the compressor. Only an emptied unit may be moved. Casters are not for transport and carry only 25% of total mass symmetrically. Leveling feet must be activated and parking brakes engaged before commissioning. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation-induced electrical faults.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and hose connections must be periodically checked for leaks and quality and replaced as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Pick the 510w when 6 kW heating and the -50…250 °C range with 5.3 kW cooling at 0 °C cover the duty. Move to 515w/520w/525w when higher heating or cooling capacity is needed; note 520w/525w are 61 dB(A) and 525 is 56 dB(A), so noise also differs.
G1/2 male connection, inlet water 5–20 °C, 1–6 bar differential pressure, 355 l/h at 15 °C water/0 °C flow. Insulate pipes if below dew point. Wetted materials include copper, 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible.
Ambient 5–40 °C, IP20 dry indoor only, not in an ATEX zone. Transport upright and emptied; allow ~2 h acclimation after cold/warm transfer. Leveling feet and parking brakes must be set. Shut off cooling water overnight. Process Safety mode requires compressor Permanently ON.
No plug is supplied for the 400 V three-phase cable (devices below 63 A). A drip tray for condensation/heat transfer fluid must be provided by the installer. Hoses and adapters for M30x1.5 pump and G1/2 cooling water are not included. Periodic hose inspection and cooling-water shutoff are required.