Peter Huber Kältemaschinenbau Unistat 525w Dynamisches
Peter Huber Kältemaschinenbau Unistat 525w Dynamisches — Temperature control units with Noise level 61 dB.
The Unistat 525w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless steel pump and Pilot ONE controller, covering -55 to 250 °C. Within the Unistat 5xx Dynamisches series it is the water-cooled variant, distinguished from the air-cooled 525 Dynamisches and the smaller 510w/515w/520w by its cooling and pump capacity. It serves as a dynamic heating/cooling source for externally closed or open tempering circuits in process and laboratory systems.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 525w Dynamisches Current | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 61 dB(A) |
| Control output | POKO ONE |
| Expansion volume | 18 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 215 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 430 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (w) free-standing Unistat 5xx with 400 V 3~ 50 Hz supply, 3x16 A fuse, max. current 13.5 A; delivered with cable but without plug.
- Magnetically coupled stainless steel pump, speed-adjustable, max. delivery 79 l/min and max. delivery pressure 1.5 bar (43 l/min at 1.0 bar); pump connection M30x1.5 male, bore shape Y (60°) per DIN 3863.
- 6 kW heating; 10 kW cooling with thermooil, falling to 0.65 kW at -50 °C with ethanol. Cooling water connection G1/2 male, 430 l/h at 15 °C water / 0 °C flow, 1 bar differential pressure, max. 6 bar.
- Compared with siblings: the air-cooled Unistat 525 Dynamisches (56 dB(A)) needs no cooling water; the 510w/515w/520w are lower-capacity water-cooled variants. This 525w is the water-cooled unit at 61 dB(A).
Behavior across the working range
- Temperature range -55...250 °C with 0.01 K display resolution; Pt100 external sensor connection.
- Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling-water temperatures reduce available cooling.
- Safety class III / FL: for FL-rated flammable heat transfer fluid the fire point must be ≥65 °C; ethanol use only under supervision.
- Process Safety mode releases maximum cooling at the OT trigger instead of switching off all poles, and requires the compressor set to Permanently ON.
Installation, control and operation
- Pilot ONE controller with POKO ONE control output; interfaces Ethernet, USB host/device and RS232; touchscreen with LED temperature, flow-chart and status indicators.
- Ambient 5...40 °C, IP20, dry indoor location; unit must not be installed in an ATEX zone. Weight 215 kg, dimensions 540x604x1332 mm.
- Cooling water 5–20 °C, ≥1 bar differential, ≤6 bar; insulate cooling-water pipes when water can fall below the dew point. Circuit materials include copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM.
- Expansion vessel 18 l, minimum filling 5.1 l; vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Power-failure behavior selectable: Off/Standby (default) or automatic restart; Power Failure Auto resumes temperature control if it was active.
Integration & Operating Notes
The unit draws up to 13.5 A on 400 V 3~ 50 Hz and is delivered with a cable but without a plug, so the electrical contractor must size the feeder and breaker for 3x16 A and fit a suitable plug. Voltage tolerance is ±5% with ±2% frequency, and the combination of -5% voltage with +2% frequency is not permitted.
Supply cooling water at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar at the G1/2 male connection. Published cooling capacity assumes 15 °C inlet and 1 bar differential; warmer water lowers available cooling. Insulate the cooling-water pipes if the water can drop below the dew point, and use only water compatible with the copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM circuit materials.
The 215 kg unit must stand upright on level ground with leveling feet activated and caster brakes engaged; casters are not for transport and carry only 25% of total mass symmetrically. IP20 requires a dry indoor location, ambient 5–40 °C, and the responsible body must install a condensation-water/heat-transfer-fluid drip tray below the unit. Installation inside an ATEX zone is prohibited.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the vessel must be locked to prevent fluid loss. Minimum filling is 5.1 l with an 18 l expansion tank, so the heat-transfer-fluid charge and expansion allowance must be planned for the actual circuit volume.
Selection & Integration Q&A
Pick the 525w when a cooling-water supply of 5–20 °C at ≥1 bar is available and 61 dB(A) is acceptable; the air-cooled 525 Dynamisches (56 dB(A)) avoids water piping but rejects heat to the room. Choose 510w/515w/520w only if their lower capacity covers the load, since the 525w provides 10 kW cooling with thermooil and 6 kW heating.
Cooling water: G1/2 male, 5–20 °C, 1–6 bar differential, 430 l/h at 15 °C/0 °C flow, with insulated pipes if below dew point. Process: M30x1.5 male pump connection with bore shape Y (60°) per DIN 3863; hoses need a ball socket per DIN 3863 and sleeve nut per DIN 3870. Electrical: 400 V 3~ 50 Hz, 3x16 A, cable supplied without plug.
Install upright and level indoors, ambient 5–40 °C, IP20, outside ATEX zones, with a drip tray; allow ~2 h acclimation after cold/warm moves. Casters are not for transport and carry only 25% of mass. Shut off cooling water overnight. For FL fluids the fire point must be ≥65 °C; ethanol only under supervision.
The base unit ships with a power cable but no plug, and no process hoses or fittings. Plan periodic hose/connection inspection and replacement, cooling-water strainer cleaning, liquefier-fin cleaning, heat-transfer-fluid inspection/replacement, and refrigerant work by approved contractors. A drip tray and, if required, a gas-detection option must be engineered on site.