Peter Huber Kältemaschinenbau Unistat 527w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 527w Dynamisches

Peter Huber Kältemaschinenbau Unistat 527w Dynamisches — Temperature control units.

The Unistat 527w Dynamisches is a water-cooled, free-standing temperature control unit in the Unistat 5xx series, rated for -55 to 250 °C with 12 kW heating and a magnetically coupled stainless-steel pump. Its 'w' suffix denotes water cooling, distinguishing it from air-cooled siblings; the 'Dynamisches' designation targets dynamic temperature ramps in externally closed or open circuits.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 527w Dynamisches Current —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 448 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 880 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and media behavior

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be 5–20 °C with 1–4 bar differential pressure

The unit is designed for cooling water intake between 5 °C and 20 °C; warmer water is outside the intended operating range and reduces cooling capacity. Differential pressure must be at least 1 bar and no more than 4 bar, with static pressure not exceeding 6 bar. Consumption is 880 l/h at 15 °C water and 0 °C flow. Insulate cooling water pipes if water temperature can fall below the condensation point. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Unistat 527w Technical Data, page 2
Mounting
Do not install in an ATEX zone; IP20 offers no water protection

The unit must not be installed or started up within an ATEX zone; explosion protection requires an Ex px cabinet with separate ATEX annex. IP20 rating means no protection against water ingress, so the location must be dry. Ambient temperature must remain between 5 °C and 40 °C. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.

Unistat Operation Manual — Unistat 5, Unistat Pilot ONE, Section 1.3.3 and Section 1.4
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power-failure response can be configured via Pilot ONE as Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power outage, with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable.

Unistat Operation Manual — Unistat 5, Unistat Pilot ONE, Section 1.7.2 and Section 1.8.1

Selection & Integration Q&A

When should I choose the Unistat 527w Dynamisches over the smaller 510w, 515w, 520w or 525 Dynamisches models?

Choose the 527w when you need the highest heating capacity (12 kW) and the largest dynamic water-cooled unit in the listed series, with a 448 kg free-standing chassis. The 510w, 515w and 520w are smaller water-cooled dynamic variants; the 525 is air-cooled (no 'w' suffix) and suits sites without a cooling water supply. Select the 527w when cooling water is available and the process demands the full -55 to 250 °C range at high heating load.

What cooling water supply and piping must be provided for the 527w, and what fittings mate with the pump connection?

Provide cooling water at 5–20 °C with 1–4 bar differential pressure and max 6 bar static pressure, connected via G3/4 male thread. Consumption is 880 l/h at 15 °C water and 0 °C flow. The pump connection is M38x1.5 male with Y-shaped bore (60°) per DIN 3863; use ball-socket fittings and sleeve nuts per DIN 3870. Insulate pipes if water can fall below the condensation point.

What are the site and duty-cycle limits for installing the 527w?

Ambient must be 5–40 °C, IP20 (dry location only), and the unit must not be installed in an ATEX zone. Leveling feet and parking brakes must be set before commissioning. Allow 2 hours acclimation after moving between cold and warm environments. The 400 V 3~ 50 Hz supply must be within ±5% voltage only with ±2% frequency; -5% voltage with +2% frequency is not allowed. Leakage current >3.5 mA.

What maintenance and accessories are required beyond the base delivery?

The unit is delivered with a cable but without a plug; a suitable three-phase plug must be provided on site. A drip tray for condensation water / heat transfer fluid must be installed below the unit. Cooling water hoses and fittings per DIN 3863/3870 are not included. Periodic checks of hoses and connections are required; cooling water should be shut off even overnight. Refrigerant-circuit work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 527w Technical Data
Datasheet ·86,0 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF