Peter Huber Kältemaschinenbau Unistat 545w Dynamisches
Peter Huber Kältemaschinenbau Unistat 545w Dynamisches — Temperature control units.
The Unistat 545w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -55 to 250 °C. It is the largest of the Unistat 5xxw series, distinguished by 24 kW heating and 35 kW cooling with Thermooil. It serves externally closed or open circuits requiring high heating and cooling capacity with precise Pt100 control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 545w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 479 kg |
| Current consumption | 58 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1630 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (w) free-standing Unistat 5xx series; no suffix indicates water cooling, unlike air-cooled siblings.
- Magnetically coupled stainless-steel pump with M38x1.5 AG male connections; pump speed adjustable.
- Pilot ONE controller with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB host/device, RS232.
- Safety class III/FL with adjustable overtemperature protection per DIN 12876; external Pt100 sensor input.
- Dimensions 730x804x1738 mm, net weight 479 kg; IP20 enclosure.
Behavior under load and media
- Temperature range -55 to 250 °C; heating capacity 24 kW.
- Cooling capacity 35 kW with Thermooil from 20 to 250 °C; drops to 1.5 kW with ethanol at -50 °C.
- Pump delivers max 196 l/min and max 2.5 bar; at 2.0 bar delivery is 86 l/min, so external circuit pressure drop must be checked.
- Suitable for externally closed and externally open applications; expansion vessel must be unlocked for closed and locked for open applications.
- Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off; requires compressor set to Permanently ON.
Installation, connection and control
- Requires 400 V 3~ 50 Hz supply, max current 58 A, fuse 3x63 A; three-phase devices below 63 A are supplied with cable but without plug.
- Cooling water: G3/4 male connection, 1630 l/h at 15 °C, min differential pressure 1 bar, max differential pressure 4 bar, max static pressure 6 bar; intake temperature 5–20 °C.
- Ambient operating range 5–40 °C; unit must not be installed in an ATEX zone; IP20 prohibits wet or dusty environments.
- Leakage current >3.5 mA requires protective earthing and RCD selection.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; combinations such as -5% voltage with +2% frequency are not allowed.
- Transport upright only; allow ~2 hours acclimation when moving between cold and warm environments before power-on.
- Leveling feet must be activated and parking brakes engaged before commissioning; casters are not for transport.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible water may be used.
- Refrigerant R-452A (A1, H280), GWP 2141, charge 3.6 kg, CO2 equivalent 7.7 t; work on coolant circuit by approved refrigeration contractors only.
Integration & Operating Notes
The unit draws up to 58 A per phase; upstream protection and cable cross-section must be sized for this. Three-phase devices below 63 A are supplied with cable but without a plug, so the final connection must be provided on site by an electrician.
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed. Site power quality must be verified before connection to avoid invalidating warranty and safe operation.
Out-of-range cooling water reduces cooling capacity and risks condensation; pipes must be insulated if water temperature can fall below the dew point. The circuit requires 1630 l/h at 15 °C with 1–4 bar differential pressure and max 6 bar static pressure.
Without an Ex px cabinet the unit is not protected against explosion. Site selection must exclude ATEX zones; operation in an Ex px cabinet requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.
Horizontal transport damages the compressor; only an emptied unit may be moved. When changing from a cold to a warm environment (or vice versa), allow about 2 hours acclimation before turning on to prevent condensation-induced electrical faults.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active.
After a power outage or when switching on, the response can be set via Pilot ONE: 'Off/Standby' (default, temperature control started only by manual input) or 'ON/Temperature control active' (temperature control always started after turning on). Choose according to whether unattended restart is acceptable for the process.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Refrigerant R-452A has GWP 2141 and 7.7 t CO2 equivalent, which affects leak detection, F-gas reporting and room ventilation obligations.
Selection & Integration Q&A
Choose the 545w when you need the highest heating capacity (24 kW) and cooling capacity (35 kW with Thermooil) in the 5xxw series, or when your process requires the larger pump delivery (196 l/min max). Smaller siblings like the 510w or 525 have lower capacities and may suffice for less demanding loads; the 545w is the largest water-cooled model in this series.
Cooling water: G3/4 male connection, 1630 l/h at 15 °C, 1–4 bar differential pressure, max 6 bar static, intake 5–20 °C. Electrical: 400 V 3~ 50 Hz, max 58 A, 3x63 A fuse; cable supplied without plug, final connection by electrician. Leakage current >3.5 mA requires protective earthing and RCD selection.
Ambient 5–40 °C, IP20 enclosure (no wet or dusty environments), not in ATEX zones. Transport upright only; allow 2 hours acclimation before power-on. Leveling feet must be activated and parking brakes engaged before commissioning. Weight 479 kg and dimensions 730x804x1738 mm require floor loading and access route verification.
Periodic checks: hoses and connections for leaks/quality, cleaning liquefier fins, cleaning strainer/dirt trap, heat transfer fluid inspection/replacement, function test of emergency stop. Cooling water supply must be shut off even overnight. Not included: final power plug (cable supplied without plug), drip tray (must be installed by responsible body), and any external Pt100 sensor or interface cables.