Peter Huber Kältemaschinenbau Unistat 610w Dynamisches
Peter Huber Kältemaschinenbau Unistat 610w Dynamisches — Temperature control units with Noise level 62 dB.
The Unistat 610w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -60 to 200 °C. It is distinguished within the Unistat 5xx/6xx series by its 6 kW heating capacity and 26 l expansion volume, serving closed or open external circuits requiring precise dynamic temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 610w Dynamisches Current | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 62 dB(A) |
| Control output | POKO ONE |
| Expansion volume | 26 l |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Weight | 360 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 400 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unistat 610w Dynamisches is a water-cooled (w suffix) free-standing unit in the Unistat 6xx family, using refrigerant R-452A (2.55 kg charge, 5.5 t CO2e) and a magnetically coupled stainless-steel pump with M30x1.5 male connections. Compared to sibling models such as the Unistat 510w (60 dB(A)) and 520w (61 dB(A)), this 610w delivers 6 kW heating and a 26 l expansion volume, with a noise level of 62 dB(A). The Pilot ONE controller with POKO ONE output provides Ethernet, USB host/device, and RS232 interfaces, plus a Pt100 external sensor input.
Thermal behavior and media constraints
Cooling capacity is strongly temperature-dependent: 2.6 kW at -40 °C falls to only 0.05 kW at -60 °C, so usable load near the lower limit is minimal. The published capacities are valid only at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure. The pump delivers 82 l/min at zero head but 29 l/min at 1.25 bar, so external circuit pressure drop directly reduces flow. Heat transfer fluid viscosity must not exceed 50 mm²/s, and the unit is rated safety class III/FL with adjustable overtemperature protection per DIN 12876.
Installation, connection and control
Electrical supply is 400 V 3~ 50 Hz with max current 15.5 A and 3x16 A fuse; three-phase units below 63 A are supplied with cable but without plug. Cooling water must be 5–20 °C, 1–6 bar differential pressure, with G1/2 male connections; consumption is 400 l/h at 15 °C. The unit is IP20 and must operate between 5 and 40 °C ambient, excluding ATEX zones. The Pilot ONE controller supports process safety mode (requires compressor permanently ON), power failure auto-restart, and adjustable overtemperature limits. For externally closed applications the expansion vessel must remain unlocked; for open applications it must be locked.
Integration & Operating Notes
The unit draws up to 15.5 A per phase and requires a 3x16 A fuse; three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug and breaker must be provided on site. Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction (e.g., -5% voltage with +2% frequency is not allowed).
Cooling water intake temperature must be between 5 °C and 20 °C, with a minimum differential pressure of 1 bar and maximum 6 bar. Consumption is 400 l/h at 15 °C. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate pipes if condensation can occur.
The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. With IP20 protection, it must be operated between 5 °C and 40 °C ambient and cannot be placed in wet or hot environments. A drip tray for condensation water or heat transfer fluid must be installed below the unit.
In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power failure auto setting can resume temperature control automatically after an outage if it was active.
Selection & Integration Q&A
Choose the 610w when you need 6 kW heating and a 26 l expansion volume for larger external circuits; the 510w and 520w have lower heating capacities and smaller expansion volumes. The 610w also has a higher noise level (62 dB(A)) than the 510w (60 dB(A)) and 520w (61 dB(A)).
Cooling water must be 5–20 °C, 1–6 bar differential pressure, with G1/2 male connections and 400 l/h consumption at 15 °C. Pump connections are M30x1.5 male with Y-shaped bore per DIN 3863; use ball socket and sleeve nut per DIN 3863/3870 for leak-free hoses.
The unit is IP20 and must operate between 5 and 40 °C ambient, excluding ATEX zones. It must be transported and stored upright only, and allowed to acclimate for about 2 hours when moving between cold and warm environments before power-on. Leveling feet must be activated before commissioning.
Periodically check hoses and connections for leaks and replace as needed; shut off cooling water even overnight. The unit is supplied with cable but without a plug, so a suitable plug must be provided on site. A drip tray must be installed below the unit. Refrigerant-circuit work requires approved contractors.