Peter Huber Kältemaschinenbau Unistat GL 530w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat GL 530w Dynamisches

Peter Huber Kältemaschinenbau Unistat GL 530w Dynamisches — Temperature control units with Noise level 61 dB.

The Unistat GL 530w Dynamisches is a water-cooled, CO2 (R-744) refrigerant free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -50 to 200 °C. Within the Unistat 5xx Dynamisches series it is the water-cooled GL variant, distinguished from air-cooled siblings by its G3/4 cooling-water circuit. It serves dynamic tempering of externally closed or open process circuits requiring 6 kW heating and up to 10 kW cooling at 0 °C.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat GL 530w Dynamisches Current 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Heating capacity 6 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...200 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 620 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Electrical
Three-phase feed with restricted voltage/frequency tolerance

The unit requires 400 V 3~ 50 Hz and permits ±5% voltage only together with ±2% frequency; the datasheet explicitly forbids combinations such as -5% voltage with +2% frequency. The site distribution and any upstream transformer or generator must be checked against this combined envelope before connection, and hard-wiring must be performed by an electrician.

Datasheet, Technical Specifications, page 2; Operation Manual, Commissioning / power connection
Cooling & Media
Cooling water must be 5–20 °C at 1–4 bar differential

The water circuit needs a G3/4 male connection, intake temperature between 5 and 20 °C, minimum 1 bar differential pressure (datasheet also lists max. 4 bar differential) and no more than 6 bar static pressure, with 620 l/h consumption at 15 °C supply and 0 °C flow. Below-dew-point supply requires insulating the cooling-water pipes to prevent condensation.

Datasheet, Technical Specifications, page 1–2
Process Limits
Fluid viscosity above 100 mm²/s overloads the pump

The magnetically coupled stainless-steel pump is limited to a maximum kinematic viscosity of 100 mm²/s, so the heat transfer fluid must be selected so that its viscosity at the lowest operating temperature stays within this limit; otherwise circulation performance drops and the pump may be overloaded.

Datasheet, Technical Specifications, page 1
Process Limits
Expansion vessel lock state depends on open vs closed circuit

For an externally closed application the expansion vessel must remain unlocked — locking it causes overpressure damage. For an externally open (bath) application the expansion vessel must be locked to prevent fluid loss. The installation must therefore be classified before commissioning, and the isolating sleeve or shut-off valves set accordingly.

Operation Manual, Commissioning / Preparations for operation
Control & I/O
Process Safety needs compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, but this mode requires the compressor to be set to Permanently ON. The OT limit heating and OT expansion vessel values must be set and function-tested in setup mode before the process runs.

Operation Manual, Setup Mode / overtemperature protection
Control & I/O
Choose restart behavior for unattended operation

After power interruption the Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart with a short INFO window to cancel). Power Failure Auto resumes an active process after outage. The choice must match whether unattended restart is acceptable for the process.

Operation Manual, Normal Operation / power interruption behavior
Mounting
Transport upright and emptied; casters are not for transport

The unit must be transported and stored upright only — horizontal transport damages the compressor — and only after draining. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically; an industrial truck is required. Leveling feet must be activated and parking brakes engaged before commissioning, and about 2 hours acclimation is needed after a cold-to-warm move before power-on.

Operation Manual, Commissioning / transport and positioning
Mounting
No ATEX zone installation; drip tray required below unit

The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet. The responsible body must install a condensation-water / heat-transfer-fluid drip tray below the unit, and national legislation may prescribe it. Ambient temperature must stay between 5 and 40 °C.

Operation Manual, Section 1.3.3 and 1.4; Datasheet, ambient temperature
Maintenance
Shut cooling water off even overnight; inspect hoses periodically

The cooling water supply should be shut off even for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Hoses and hose connections must be checked periodically for leaks and quality and replaced as required; the hat-type strainer/dirt trap requires draining the cooling circuit for cleaning, and refrigerant-circuit work must be done by approved refrigeration/air-conditioning contractors.

Operation Manual, Service and Maintenance

Selection & Integration Q&A

When should I choose the GL 530w Dynamisches over the air-cooled 510w/520w or the 525 Dynamisches?

Pick the GL 530w when a CO2 (R-744) refrigerant and water cooling are required and a G3/4 cooling-water supply of 5–20 °C at 1–4 bar differential is available. Air-cooled 510w/520w siblings avoid water plumbing but reject heat to room air; the 525 Dynamisches is a different capacity point in the series. Choose by refrigerant, cooling medium and required cooling power at your lowest process temperature.

What must the process circuit and pump connection provide to match this unit?

The pump outlet is M38x1.5 male (bore shape Y, 60°, DIN 3863), so tempering hoses need compatible ball sockets and sleeve nuts. The pump delivers max. 196 l/min at max. 2.5 bar, with flow falling as pressure drop rises — size the external circuit against the pump curve. Minimum filling is 8.9 l with a 19.9 l expansion tank, and fluid viscosity must stay ≤100 mm²/s.

What site and duty-cycle limits apply before I can commission the unit?

Ambient must be 5–40 °C, the unit must not sit in an ATEX zone, and a drip tray is required below it. Transport upright and emptied; casters are not for transport. Allow ~2 h acclimation after a cold/warm move. Supply is 400 V 3~ 50 Hz with ±5% voltage only at ±2% frequency. Cooling water must be 5–20 °C, 1–4 bar differential, max. 6 bar.

What maintenance and accessories are needed beyond the base delivery?

Base delivery excludes cooling-water hoses and fittings (G3/4 male), process hoses with M38x1.5 ball sockets, heat transfer fluid, and the drip tray. Plan periodic hose/connection inspection and replacement, cooling-water shutoff for overnight downtime, strainer cleaning with circuit drain, and refrigerant work only by approved contractors. The optional gas detection sensor and processing unit are installed by the operating company.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat GL 530w Technical Data Sheet
Datasheet ·98,3 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF