Peter Huber Kältemaschinenbau Unistat GL P550w Dynamisches
Peter Huber Kältemaschinenbau Unistat GL P550w Dynamisches — Temperature control units.
The Unistat GL P550w Dynamisches is a water-cooled, CO2 (R-744) refrigerated circulating thermostat with a magnetically coupled stainless-steel pump, covering -50 to 200 °C with 24 kW heating. Its high-pressure pump delivers 201 l/min at zero head and 42 l/min at 5.0 bar, distinguishing it from standard-pressure siblings for high-pressure-drop loops. It serves precise temperature control of externally closed industrial processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat GL P550w Dynamisches Current | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...200 °C |
| Weight | 799 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 3100 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing water-cooled unit (suffix 'w') using R-744 refrigerant (suffix 'GL'), 8.7 kg charge, GWP 1; 799 kg, 918x963x1771 mm.
- 'P' designates the high-pressure-drop pump variant: magnetically coupled stainless-steel pump, 201 l/min at 0 bar, 42 l/min at 5.0 bar, speed-adjustable.
- Siblings in the Unistat 5xx dynamic series (510w, 515w, 520w, 525) are air-cooled or standard-pressure alternatives; this model is selected when the external circuit has high pressure drop and CO2 refrigerant is required.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB host/device, RS232; external Pt100 sensor input.
Behavior under load and media
- Temperature range -50 to 200 °C; heating capacity 24 kW.
- Cooling water required at 5–20 °C, 1–4 bar differential, max 6 bar static; consumption 3100 l/h at 15 °C water/0 °C flow.
- Heat transfer fluid max kinematic viscosity 100 mm²/s; safety class III/FL requires fire point ≥65 °C for flammable fluids.
- Minimum filling 8.9 l; expansion tank 19.9 l.
Installation, connection and control
- 400 V 3~ 50 Hz, max current 61.5 A, fuse 3x63 A; delivered with cable, without plug.
- Cooling water connections G3/4 male; pump connections M38x1.5 AG male.
- IP20, ambient 5–40 °C; EMC Class A Group 1 (EN55011); leakage current >3.5 mA.
- Pilot ONE supports OT protection, Process Safety mode, Delta T limiter, program/clock functions; power-failure behavior configurable (Off/Standby or ON).
Integration & Operating Notes
The datasheet permits ±5% voltage only together with ±2% frequency, and explicitly forbids combinations such as -5% voltage with +2% frequency. Verify the plant supply quality before connection; the unit draws up to 61.5 A and requires a 3x63 A fuse.
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug. The customer must provide and fit the site plug, and hard-wired connection must be performed by an electrician with supply matching the rating plate.
When cooling water temperatures fall below the condensation point, the cooling water pipes inside the unit must be insulated to prevent condensation and corrosion. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
The pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar. Size the external circuit so its pressure drop at the required flow stays within the pump curve; otherwise the required flow cannot be achieved. Maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.
Only transport the temperature control unit in an upright position; horizontal transport damages the compressor. Only an emptied unit may be moved, as filled transport risks thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck.
When moving between cold and warm environments (or vice versa), allow the unit to acclimate for about 2 hours before turning it on to prevent condensation-induced electrical faults. Leveling feet must be unscrewed/activated and parking brakes engaged before commissioning.
In Process Safety mode, when the overtemperature trigger is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active under exothermic conditions.
After a power outage or when switching on, the Pilot ONE can be set to 'Off/Standby' (default, manual start) or 'ON/Temperature control active' (automatic restart with a few-second INFO to disable). Power Failure Auto resumes temperature control automatically if it was active before the outage.
Hoses and connections must be periodically checked for leaks and quality and replaced as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors; general maintenance personnel cannot service the refrigerant circuit.
The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. The installation room must stay within 5–40 °C ambient. EMC Class A Group 1 equipment is intended for industrial environments, not residential.
Selection & Integration Q&A
Choose the P550w when the external circuit has high pressure drop and you need CO2 refrigerant (GL) with water cooling (w). Its pump delivers 42 l/min at 5.0 bar. Siblings such as the 520w are air-cooled or standard-pressure alternatives for lower pressure-drop loops.
Cooling water: 5–20 °C, 1–4 bar differential, max 6 bar static, G3/4 male connections. Electrical: 400 V 3~ 50 Hz, max 61.5 A, 3x63 A fuse; cable supplied without plug. Voltage tolerance ±5% only with ±2% frequency.
Install upright only, on level ground with leveling feet activated and casters braked. Ambient 5–40 °C, no ATEX zone. Allow 2 h acclimation after temperature changes. Transport only when emptied; casters are not for transport.
Periodically check hoses and replace as needed; shut off cooling water during downtimes. Refrigerant work by approved contractors. Base delivery excludes the site power plug and a drip tray, which the responsible body must install.