No photo available yet
Peter Huber Kältemaschinenbau Unistat P510 Dynamisches
Peter Huber Kältemaschinenbau Unistat P510 Dynamisches — Temperature control units with Noise level 60 dB.
The Unistat P510 Dynamisches is a free-standing, air-cooled refrigerated heating circulator with a -50 to 250 °C working range, 6 kW heating and a magnetically coupled stainless-steel pump. Its 'P' designation targets applications with high pressure drop, distinguishing it from the w (water-cooled) and standard siblings in the Unistat 5xx series. It serves single-device temperature control of closed or open process loops in industrial environments.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P510 Dynamisches Current | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Noise level | 60 dB(A) (sound pressure level) |
| Control output | POKO ONE (potential-free contact) |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...250 °C |
| Weight | 232 kg (net) |
| Current consumption | 14 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s (max) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 5xx platform: 232 kg net, 560 x 754 x 1457 mm installation envelope, IP20, air-cooled refrigeration circuit with R-452A (A1, H280, GWP 2141).
- The 'P' suffix marks the high-pressure-drop modification; siblings Unistat 510w, 515w and 520w are water-cooled ('w') variants, and the Unistat 525 is a different capacity step. Choose the P510 when the process loop imposes significant resistance and when air cooling is acceptable.
- Pump: magnetically coupled, stainless steel, speed-adjustable, M30x1.5 male connection (DIN 3863 ball socket / DIN 3870 sleeve nut mating).
- Control: Pilot ONE controller with POKO ONE potential-free contact output, 0.01 K display resolution, Pt100 external sensor input, Ethernet/USB host and device/RS232.
Behavior under load and media
- Temperature range -50...250 °C from one device; heating capacity 6 kW.
- Cooling capacity is strongly temperature-dependent: 5.3 kW at 20 °C but only 0.25 kW at -50 °C, so low-temperature duty must be sized at the actual set point.
- Pump delivery is 119 l/min at zero head and 33 l/min at 2.5 bar; verify the external pressure drop against this curve before specifying flow.
- Maximum viscosity 50 mm²/s; safety class III / FL, so flammable heat transfer fluids require a fire point ≥ 65 °C and, for ethanol, supervised operation only.
- Process Safety mode releases maximum cooling at the OT trigger instead of switching off all poles; it requires the compressor set to Permanently ON.
Installation, connection and operation
- 400 V 3~ 50 Hz supply, 14 A max current, 3x16 A fusing; three-phase units below 63 A ship with cable but without a plug. Voltage/frequency tolerance ±5% / ±2%, but not simultaneously in the unfavorable combination.
- Leakage current exceeds 3.5 mA: earthing and RCD selection must account for this. EMC Class A Group 1 — industrial environment only.
- Air-cooled condenser requires free airflow; cooling capacity is rated at 20 °C ambient and falls as ambient rises. Permissible ambient 5...40 °C.
- Externally closed applications: expansion vessel unlocked. Externally open (bath) applications: expansion vessel locked. Locking on a closed loop causes overpressure damage.
- Power-interruption behavior selectable: Off/Standby (default) or automatic restart; Power Failure Auto resumes temperature control if it was active.
- Transport upright and emptied only; casters are not for transport and carry only 25% of total mass. Allow ~2 h acclimation after a cold/warm move before power-on.
Integration & Operating Notes
The unit is delivered with a cable but without a plug for three-phase supplies below 63 A, so the customer must fit a suitable plug. Plan the feeder for 14 A max current and 3x16 A fusing, and verify that voltage and frequency deviations do not occur simultaneously in the unfavorable combination (-5% voltage with +2% frequency is not allowed).
Leakage current exceeds 3.5 mA, which affects protective earthing design and residual-current device choice; a standard 30 mA RCD may nuisance-trip. The electrical installer must select and document the protective measure accordingly.
Cooling capacity falls from 5.3 kW at 20 °C to 0.25 kW at -50 °C. A process requiring significant heat removal near the cold end must be checked against the capacity at its real operating temperature, otherwise the unit will not reach or hold set point.
The pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar. High-pressure-drop processes (the reason to choose the P modification) must be evaluated against this curve; if the required flow at the loop's pressure drop is below the process demand, a different pump or piping change is needed.
The manufacturer explicitly requires respecting space requirements. The refrigeration machine is air-cooled, so the bay must allow heat rejection and service access; an undersized enclosure will raise ambient temperature and reduce cooling capacity, which is rated at 20 °C room temperature.
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit, and national legislation may prescribe it. The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet with its own rating plate.
In Process Safety mode the controller releases maximum cooling at the overtemperature trigger instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; without that setting the protective behavior is not available.
Only transport the unit upright and emptied; horizontal transport damages the compressor and a filled unit risks thermal fluid overflow. Casters, if present, carry only 25% of total mass symmetrically and are not suitable for transport — use an industrial truck. Allow about 2 hours acclimation after a cold/warm move before power-on.
Selection & Integration Q&A
Pick the P510 when the process loop has high pressure drop (the 'P' modification) and air cooling is acceptable in the plant room. Choose a 'w' sibling when cooling water is available and lower ambient heat rejection is needed; choose the 520w only if its higher capacity step is required.
The pump connection is M30x1.5 male with DIN 3863 ball socket and DIN 3870 sleeve nut mating, so hoses and fittings must match that standard. The unit ships with a cable but without a plug for the 400 V three-phase supply; the site plug and the drip tray are customer-supplied.
Permissible ambient is 5...40 °C, and cooling capacity is rated at 20 °C and drops as ambient rises, so keep the room cool. The unit is EMC Class A Group 1 for industrial environments, must not sit in an ATEX zone, and needs the full 560 x 754 x 1457 mm bay with airflow clearance.
Plan periodic hose and connection inspection with replacement as needed, cleaning of condenser fins and the dirt trap, and heat transfer fluid inspection/replacement. Refrigerant work requires approved contractors. The site plug, drip tray and heat transfer fluid are not included in base delivery.