Peter Huber Kältemaschinenbau Unistat 527w Dynamisches
Peter Huber Kältemaschinenbau Unistat 527w Dynamisches — Temperature control units.
The Unistat 527w Dynamisches is a water-cooled, free-standing temperature control unit in the Unistat 5xx series, rated for -55 to 250 °C with 12 kW heating and a magnetically coupled stainless-steel pump. Its 'w' suffix denotes water cooling, distinguishing it from air-cooled siblings; the 'Dynamisches' designation targets dynamic temperature ramps in externally closed or open circuits.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 527w Dynamisches Current | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -55...250 °C |
| Weight | 448 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 880 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 5xx chassis with casters and leveling feet; water-cooled ('w') condenser circuit, requiring a site cooling water supply.
- Sibling models in the same series (510w, 515w, 520w, 525) share the 'Dynamisches' dynamic-duty concept; the 527w is the largest of the listed water-cooled dynamic variants, with a 448 kg mass and 12 kW heating capacity.
- Pump is magnetically coupled, stainless steel, with adjustable speed; pump connection is M38x1.5 male with Y-shaped bore (60°) per DIN 3863, mating to ball-socket fittings and sleeve nuts per DIN 3870.
- Controller is Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB (host and device) and RS232; external Pt100 sensor connection.
Working range and media behavior
- Temperature range -55 to 250 °C; safety class III / FL, so flammable heat transfer fluids must have a fire point ≥ 65 °C, and ethanol use is restricted to supervised operation.
- Cooling capacity is rated at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure; capacity drops as cooling water temperature rises, and flow rate may increase.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used. Insulate cooling water pipes if water temperature can fall below the condensation point.
- Minimum filling capacity 8.2 l; expansion tank 19.9 l. For externally closed applications the expansion vessel must remain unlocked; for externally open (bath) applications it must be locked.
Installation, control and operation
- Requires 400 V 3~ 50 Hz supply with max current 31.5 A and 3x32 A fuse; delivered with cable but without plug (three-phase <63 A). Leakage current >3.5 mA.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; combinations such as -5% voltage with +2% frequency are not allowed.
- Cooling water: 5–20 °C intake, 1–4 bar differential pressure, max 6 bar static pressure; connection G3/4 male; consumption 880 l/h at 15 °C, flow 0 °C.
- Ambient operating range 5–40 °C; protection IP20 (no water protection). Must not be installed in an ATEX zone.
- Pilot ONE offers Process Safety mode (releases maximum cooling at OT trigger instead of switching off; requires compressor set to Permanently ON) and configurable power-failure response (Off/Standby or ON/Temperature control active).
- Refrigerant R-452A, GWP 2141, charge 2.9 kg, CO2 equivalent 6.2 t.
Integration & Operating Notes
The unit is designed for cooling water intake between 5 °C and 20 °C; warmer water is outside the intended operating range and reduces cooling capacity. Differential pressure must be at least 1 bar and no more than 4 bar, with static pressure not exceeding 6 bar. Consumption is 880 l/h at 15 °C water and 0 °C flow. Insulate cooling water pipes if water temperature can fall below the condensation point. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
The unit must not be installed or started up within an ATEX zone; explosion protection requires an Ex px cabinet with separate ATEX annex. IP20 rating means no protection against water ingress, so the location must be dry. Ambient temperature must remain between 5 °C and 40 °C. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. The power-failure response can be configured via Pilot ONE as Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power outage, with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable.
Selection & Integration Q&A
Choose the 527w when you need the highest heating capacity (12 kW) and the largest dynamic water-cooled unit in the listed series, with a 448 kg free-standing chassis. The 510w, 515w and 520w are smaller water-cooled dynamic variants; the 525 is air-cooled (no 'w' suffix) and suits sites without a cooling water supply. Select the 527w when cooling water is available and the process demands the full -55 to 250 °C range at high heating load.
Provide cooling water at 5–20 °C with 1–4 bar differential pressure and max 6 bar static pressure, connected via G3/4 male thread. Consumption is 880 l/h at 15 °C water and 0 °C flow. The pump connection is M38x1.5 male with Y-shaped bore (60°) per DIN 3863; use ball-socket fittings and sleeve nuts per DIN 3870. Insulate pipes if water can fall below the condensation point.
Ambient must be 5–40 °C, IP20 (dry location only), and the unit must not be installed in an ATEX zone. Leveling feet and parking brakes must be set before commissioning. Allow 2 hours acclimation after moving between cold and warm environments. The 400 V 3~ 50 Hz supply must be within ±5% voltage only with ±2% frequency; -5% voltage with +2% frequency is not allowed. Leakage current >3.5 mA.
The unit is delivered with a cable but without a plug; a suitable three-phase plug must be provided on site. A drip tray for condensation water / heat transfer fluid must be installed below the unit. Cooling water hoses and fittings per DIN 3863/3870 are not included. Periodic checks of hoses and connections are required; cooling water should be shut off even overnight. Refrigerant-circuit work requires approved contractors.