Peter Huber Kältemaschinenbau Unistat 625w Dynamisches
Peter Huber Kältemaschinenbau Unistat 625w Dynamisches — Temperature control units.
The Unistat 625w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -60 to 200 °C with 12 kW heating. Its water-cooled condenser requires 1–4 bar differential cooling water pressure and 5–20 °C intake, distinguishing it from air-cooled siblings. It serves dynamic process temperature control in externally closed or open circuits.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 625w Dynamisches Current | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled ('w') free-standing Unistat 6xx model; the 'Dynamisches' designation indicates the dynamic temperature control variant within the series.
- Pump: magnetically coupled, stainless steel, speed-adjustable; pump connection M38x1.5 male.
- Cooling water connection G3/4 male; cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; protection rating IP20.
- Sibling models in the series (510w, 515w, 520w, 525) differ in size/configuration; this 625w is the water-cooled unit requiring site cooling water.
Behavior under load and media
- Temperature range -60…200 °C; heating capacity 12 kW.
- Cooling power is strongly temperature-dependent: 15 kW at -20 °C falling to 1.7 kW at -60 °C (with ethanol), so low-temperature loads must be sized to the reduced capacity.
- Maximum permissible kinematic viscosity of heat transfer fluid: 100 mm²/s, limiting fluid choice at low temperatures.
- Minimum filling capacity 8.9 l; expansion tank capacity 19.9 l.
- For FL-rated flammable fluids, fire point must be ≥65 °C; ethanol use requires supervised operation.
Installation, control and operation
- Power supply 460 V 3~ 60 Hz; external branch circuit protection per UL 489 required; three-phase units under 63 A are delivered with cable but without plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
- Controller Pilot ONE with POKO ONE control output; interfaces Ethernet, USB (Host and Device), RS232; external Pt100 sensor connection; display resolution 0.01 K.
- Ambient temperature range 5…40 °C; unit must not be installed in an ATEX zone.
- Cooling water intake temperature 5…20 °C; pipes must be insulated if condensation conditions occur.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Process Safety mode releases maximum cooling at the OT trigger instead of switching off; requires compressor set to Permanently ON.
- Power Failure Auto can resume temperature control automatically after outage if active.
Integration & Operating Notes
The water-cooled condenser requires a cooling water supply with minimum 1 bar differential pressure and maximum 6 bar pressure. Insufficient differential pressure causes inadequate cooling; overpressure can damage the circuit. Confirm both values at the installation point before commissioning.
Achievable cooling power depends on the cooling water intake temperature, and condensation risk rises at low intake temperatures. Insulate pipes where condensation conditions occur. Operating outside 5–20 °C intake affects performance and reliability.
The unit requires a 460 V three-phase 60 Hz supply and external branch circuit protection according to UL 489. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed. Confirm site supply before ordering.
Because current consumption is below 63 A, the unit ships with a cable but no plug. The customer must provide a suitable plug and have the connection made by an electrician, matching the rating plate voltage and frequency.
The magnetically coupled stainless-steel pump is limited to a maximum kinematic viscosity of 100 mm²/s. Selecting a fluid that is too viscous at the lowest operating temperature can overload the pump and reduce circulation, so fluid choice must be checked against the -60 °C end of the range.
With safety class III / FL, any flammable heat transfer fluid used must have a fire point of at least 65 °C. If ethanol is used, only supervised operation is permitted. This constrains fluid selection and staffing for the application.
Without an Ex px cabinet the unit is not protected against explosion and must not be installed or started within an ATEX zone. Site selection must exclude hazardous areas; Ex px operation requires the separate ATEX annex and only the model/serial on the Ex px rating plate is suitable.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; configure this in the Pilot ONE before relying on the function for exothermic processes.
The Pilot ONE can be set to 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart after power interruption, with a brief INFO allowing cancellation). Power Failure Auto resumes temperature control if it was active. Select according to whether unattended restart is acceptable for the process.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as needed.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant R-452A has a GWP of 2141, which may trigger regulatory, reporting or leakage-control obligations at the site.
Selection & Integration Q&A
Choose the 625w when you need a water-cooled unit with 12 kW heating and -60…200 °C range and have a suitable cooling water supply (1–4 bar differential, 5–20 °C intake). If no cooling water is available, an air-cooled sibling such as the 525 Dynamisches avoids the water connection; the 520w is also water-cooled but a different size in the series.
Provide cooling water at 5–20 °C intake with minimum 1 bar differential and maximum 6 bar pressure, connected via G3/4 male. Electrical supply must be 460 V 3~ 60 Hz with external UL 489 branch protection; the unit ships with cable but no plug, so a suitable plug must be provided and connection made by an electrician.
Operate within 5…40 °C ambient. Do not install in an ATEX zone. Transport and store upright only, allow 2 hours acclimation after temperature changes before power-on, and activate leveling feet or parking brakes before commissioning. Insulate cooling water pipes where condensation occurs.
The unit ships without a plug; provide one. A drip tray for condensation water/heat transfer fluid must be installed below the unit. Shut off cooling water even overnight, periodically check hoses and connections, and have refrigerant-circuit work done by approved contractors. Heat transfer fluid is not included and must meet the 100 mm²/s viscosity limit.