No photo available yet
Peter Huber Kältemaschinenbau Unistat P620w Dynamisches
Peter Huber Kältemaschinenbau Unistat P620w Dynamisches — Temperature control units.
The Unistat P620w Dynamisches is a water-cooled, high-pressure-drop circulating temperature control unit with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 12 kW heating. Its 'P' designation targets applications with high pressure drop, distinguishing it from the standard 5xx/6xx siblings. It serves as a dynamic heating/cooling source for externally closed process circuits requiring three-phase power and facility cooling water.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P620w Dynamisches Current | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled ('w') free-standing Unistat with a magnetically coupled, stainless-steel pump and adjustable pump speed; the 'P' suffix marks the variant intended for applications with high pressure drop.
- Sibling models in the series (Unistat 510w, 515w, 520w, 525 Dynamisches) differ in configuration; this card covers only the P620w Dynamisches.
- Controller is Pilot ONE with POKO ONE control output; interfaces include Ethernet, USB (host and device) and RS232; external Pt100 sensor input.
- Safety class III / FL (flammable fluid rating); refrigerant R-1270 (A3, H220), a flammable natural refrigerant.
Working envelope and behavior
- Temperature range -60 to 200 °C with 12 kW heating capacity; display resolution 0.01 K.
- Cooling capacity with ethanol falls from 12 kW at -20 °C to 5.6 kW at -40 °C and 1.4 kW at -60 °C, so process design must account for reduced cooling at the low end.
- Maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, limiting fluid choice at low temperatures.
- Ambient operating range 5–40 °C; IP20 protection requires a dry indoor location.
Installation, connection and control
- Requires 400 V 3~ 50 Hz supply; three-phase units below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%.
- Cooling water connection G3/4 male; minimum differential pressure 1 bar, maximum cooling water pressure 6 bar; cooling water inlet temperature 5–20 °C, with pipe insulation required below that.
- Pump connection M38x1.5 male; mating hoses/adapters must match this thread.
- Pilot ONE controller supports overtemperature protection settings, Delta T limiter, Pt100 monitoring, auto/expert control modes, and temperature control programs; power-interruption behavior is configurable (Off/Standby or automatic restart).
Integration & Operating Notes
Cooling water is connected via G3/4 male and must be supplied at 1–4 bar differential pressure and no more than 6 bar, with inlet temperature between 5 °C and 20 °C; lower inlet temperatures require insulating the cooling water pipes. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be suitable to avoid corrosion and fouling.
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s; exceeding it can overload the magnetically coupled pump and reduce flow. Fluid selection at low temperatures must respect this limit, and for FL-rated fluids the fire point must be ≥65 °C, with ethanol permitted only under supervised operation.
The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on that cabinet's rating plate is suitable. The refrigerant R-1270 (A3, H220) is flammable, which affects site safety classification and handling procedures.
After a power outage or switch-on, the Pilot ONE can be set to 'Off/Standby' (default, manual start required) or 'ON/Temperature control active' (automatic restart with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable for the process; Process Safety mode additionally requires the compressor set to Permanently ON.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed; work on the coolant circuit must be carried out by approved refrigeration/air-conditioning contractors.
Selection & Integration Q&A
Pick the P620w when the application has high pressure drop (the 'P' designation) and needs the -60 to 200 °C range with 12 kW heating. The 510w/520w are different capacity points in the series; confirm the required cooling capacity at your lowest process temperature, since the P620w delivers only 1.4 kW at -60 °C with ethanol.
A 400 V 3~ 50 Hz supply with a site-fitted plug (unit ships with cable, no plug below 63 A) and voltage/frequency within ±5%/±2%. Cooling water via G3/4 male at 1–4 bar differential and max 6 bar, inlet 5–20 °C. Process hoses must match the M38x1.5 male pump connection.
Install indoors, dry, IP20, ambient 5–40 °C, never in an ATEX zone. Transport and store upright only; allow ~2 h acclimation after temperature changes before power-on. Activate leveling feet/parking brakes, and install a drip tray below the unit as required by local regulations.
Periodically inspect/replace heat transfer fluid and hoses, clean liquefier fins and the dirt trap, and test the emergency stop. The power plug, cooling-water hoses, heat transfer fluid and drip tray are not part of base delivery; refrigerant-circuit work requires approved contractors.