Peter Huber Kältemaschinenbau Unichiller 040T
Peter Huber Kältemaschinenbau Unichiller 040T — Recirculating chillers with Noise level 69 dB.
The Unichiller 040T is a free-standing, air-cooled recirculating chiller with a single-stage R-449A refrigerant circuit and a fixed-speed pump (type B, G3/4 male). It covers -10 to 40 °C and is rated for 400 V 3~ 50 Hz with a 3x10 A fuse. It serves externally closed process loops requiring up to 69 dB(A) acoustic tolerance.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 040T Current | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 69 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...40 °C |
| Weight | 167 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-449A (A1, H280) (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The 040T is the base 40-size model in the Unichiller series; siblings 030T, 030Tw, 030Tw-H and 040T-H differ in size, water-cooling ('w') or high-temperature ('-H') configuration. The 040T is air-cooled with a single-stage R-449A circuit (1.9 kg, 2.7 t CO2e).
- Pump type B is fixed-speed (not adjustable) with G3/4 male connections; minimum filling capacity 3.5 l and expansion volume 3.5 l.
- Controller is Pilot ONE with Ethernet, USB host/device and RS232; external Pt100 sensor input available. Protection IP20, weight 167 kg.
Behavior and process limits
- Temperature range -10 to 40 °C; cooling capacity is rated at 20 °C room temperature and derates as ambient rises. Ambient limit 5–40 °C.
- Intended only for externally closed applications. Models with 'P' in the name are for high pressure drop; the 040T has no 'P' designation.
- Safety class I/NFL per DIN 12876; Process Safety mode can release maximum cooling on overtemperature trigger instead of all-pole switch-off.
Installation and operation
- Requires 400 V 3~ 50 Hz with 3x10 A fuse; delivered with cable but no plug (three-phase <63 A). Voltage/frequency tolerances must be met simultaneously; -5% voltage with +2% frequency is not allowed.
- Must be installed indoors, not in ATEX zones, with illuminance ≥300 lx and altitude ≤2000 m. Wall clearance 50–200 cm per side for maintenance. Leveling feet/parking brakes must be set before start-up.
- Refrigerant charge vs room volume must be checked; cooling water supply (if water-cooled variant) must be shut off even overnight. EMC class A group 1 for industrial environments.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance and airflow. Plan the footprint accordingly; the unit weighs 167 kg and must sit on level ground with leveling feet or parking brakes engaged before start-up.
The chiller is a three-phase device drawing less than 63 A, so it ships with a cable but without a plug; the customer must fit a suitable plug. The electrical installation must be sized for a 3x10 A fuse. Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency combination is explicitly not allowed.
The unit is designed for externally closed applications. The hydraulic circuit must be closed; open vessels or atmospheric baths are outside the intended use. Minimum filling capacity is 3.5 l and expansion volume 3.5 l, so the external loop volume and expansion allowance must be planned accordingly.
The unit must not be installed or started inside an ATEX zone (not explosion-protected). It is for indoor use only, at illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must stay within 5–40 °C; cooling capacity is rated at 20 °C and derates at higher ambient. IP20 requires a dry, protected location.
Selection & Integration Q&A
The 040T is the base 40-size air-cooled model for standard temperature control down to -10 °C. Choose 030T for lower capacity needs; 030Tw for water-cooled (61 dB(A)); 040T-H for higher-temperature duty. The 040T has no 'P' high-pressure-drop designation.
Pump connections are G3/4 male, so hoses/fittings must match. Electrical supply is 400 V 3~ 50 Hz with 3x10 A fuse; the unit comes with cable but no plug, so the customer must provide one. External Pt100 sensor can be connected to the controller.
Install indoors only, not in ATEX zones, at 5–40 °C ambient, ≥300 lx, ≤2000 m altitude. Maintain 50–200 cm side clearance for service. Engage leveling feet/parking brakes before start-up. Cooling capacity derates above 20 °C room temperature.
Refrigerant circuit work requires certified contractors; leak checks apply per 1.9 kg R-449A (2.7 t CO2e). Shut off cooling water even overnight. A mains plug is not included (three-phase <63 A). External Pt100 sensor and process hoses are not part of base delivery.