Peter Huber Kältemaschinenbau Unichiller 040T-H
Peter Huber Kältemaschinenbau Unichiller 040T-H — Recirculating chillers with Noise level 69 dB.
The Unichiller 040T-H is a 400 V three-phase, air-cooled recirculating chiller with a 2 kW heater and a -10 to 100 °C working range, controlled by the Pilot ONE. Its 'H' suffix denotes the heating function, distinguishing it from the unheated 040T sibling. It serves externally closed cooling circuits requiring both heat removal and active temperature elevation.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 040T-H Current | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 69 dB(A) |
| Heating capacity | 2 kW |
| Maximum delivery flow rate | 26 l/min (pressure pump max delivery capacity) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 171 kg |
| Current consumption | 9 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
The 040T-H is a free-standing, air-cooled chiller with an integrated 2 kW heater, rated for a -10 to 100 °C temperature range. Within the Unichiller series, the 'H' designation marks the heated variant: the base 040T lacks the 2 kW heating stage, while the 030Tw/030Tw-H are quieter (61 dB(A)) but smaller-capacity models; the 040T-H shares the 69 dB(A) noise level of the 030T and 040T. The unit is classified safety class III/FL per DIN 12876 with adjustable overtemperature protection, and is intended exclusively for externally closed applications. The refrigerant is R-449A (GWP 1397, CO₂ equivalent 2.7 t), and the refrigeration machine is CFC/HCFC-free.
Hydraulic and Thermal Behavior
The circulation pump (type B, non-adjustable speed) delivers a maximum of 26 l/min and 8 l/min at 2.0 bar, so the external circuit pressure drop must be checked against this curve. Minimum filling capacity is 3.5 l, and the atmospheric open expansion tank with digital level indicator requires fluid-level monitoring. Cooling capacity is rated at 20 °C room temperature and derates at higher ambient temperatures; the permissible ambient range is 5 to 40 °C. The 2 kW heater enables active temperature elevation within the -10 to 100 °C envelope.
Installation, Control and Operation
The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 connection for an external sensor. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. The unit requires a 400 V 3~ 50 Hz supply with a 3×10 A fuse; three-phase devices under 63 A are delivered with a cable but without a plug. Voltage and frequency tolerances must be met simultaneously, and the combination of -5% voltage with +2% frequency is not allowed. The pump connection is G3/4 male, and maximum tightening torques must not be exceeded with connections counterheld. A wall clearance of 50 to 200 cm per side is required for maintenance, and the unit must be used indoors at ≥300 lx and up to 2000 m altitude. Leveling feet must be activated and parking brakes applied before start-up. The unit is EMC class A, group 1 (EN55011), intended for industrial environments.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to permit maintenance access to the refrigeration circuit and condenser. Insufficient clearance also restricts air flow around the air-cooled condenser, which can reduce cooling capacity and cause overheating.
The 400 V 3~ 50 Hz supply must be protected by a 3×10 A fuse. Because the unit is delivered with a cable but without a plug (three-phase devices under 63 A), the customer must supply and fit the site plug. Voltage and frequency tolerances must be satisfied simultaneously; a -5% voltage with +2% frequency combination is explicitly not permitted.
The unit is designed exclusively for externally closed applications. The pump delivers 26 l/min maximum and 8 l/min at 2.0 bar, so the external loop pressure drop must be verified against this curve. Minimum filling capacity is 3.5 l, and the atmospheric open expansion tank requires fluid-level monitoring via the digital indicator.
In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This prevents runaway temperatures in exothermic processes. The power-failure response is separately configurable as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with a CO₂ equivalent of 2.7 t, which may trigger periodic leak-check obligations under F-gas rules. Cooling water supply (where applicable) must be shut off even for short downtimes such as overnight.
The unit may only be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must remain within 5 to 40 °C. Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient temperatures, so capacity must be derated for warm rooms. The unit must not be installed or started inside an ATEX zone.
Selection & Integration Q&A
Choose the 040T-H when the process requires active heating up to 100 °C in addition to cooling; the base 040T lacks the 2 kW heater. If noise is critical, the 030Tw/030Tw-H at 61 dB(A) are quieter than the 69 dB(A) of the 040T-H, but they are smaller-capacity models.
The customer must provide a 400 V 3~ 50 Hz supply with 3×10 A fusing and fit a site plug (cable delivered without plug). Hoses and fittings must match the G3/4 male pump connection, and the external loop plus expansion volume must hold at least 3.5 l. The loop pressure drop must be checked against the pump curve (8 l/min at 2.0 bar).
Install indoors only, at 5–40 °C ambient and ≥300 lx illuminance, below 2000 m altitude, and never inside an ATEX zone. Allow 50–200 cm wall clearance per side for maintenance and condenser air flow. Activate leveling feet and parking brakes before start-up. Cooling capacity is rated at 20 °C room temperature and derates at higher ambient.
Refrigerant circuit work must be performed by approved refrigeration contractors; R-449A charge has a CO₂ equivalent of 2.7 t, which may trigger F-gas leak checks. The site plug is not included. Fluid level must be monitored via the digital indicator, and cooling water supply (if applicable) must be shut off even overnight.