Peter Huber Kältemaschinenbau Unichiller 050w-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 050w-H

Peter Huber Kältemaschinenbau Unichiller 050w-H — Recirculating chillers with Noise level 62 dB.

The Unichiller 050w-H is a water-cooled recirculating chiller with a 2 kW heater, a fixed-speed E1 pump, and a -20 to 100 °C working range, controlled by the Pilot ONE. It is the 460 V 3~ 60 Hz variant of the Unichiller series, distinguished from air-cooled siblings by its G1/2 cooling-water connection and 6 bar maximum water pressure. It serves externally closed process loops requiring both cooling and active heating within a 48 l expansion volume.

Noise level
62 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 050w-H Current 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Noise level 62 dB(A)
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 290 kg
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 137 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 167 л/час · at 15°C, flow 0°C: 216 l/h
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Mounting
Maintain 50–200 cm clearance per side for service access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access to the refrigerant circuit and pump. Insufficient clearance blocks service work and may prevent proper airflow around the unit.

Unichiller Operation Manual, Section on Installation (page 25
Electrical
Provide UL 489 branch protection and a 3x16 A circuit

The 460 V 3~ 60 Hz supply must be protected by an external branch circuit breaker according to UL 489. The unit draws up to 3x16 A; the cable is supplied without a plug, so the installer must provide the final connection and ensure voltage/frequency tolerances of ±5% and ±2% are not exceeded simultaneously in opposite directions.

Unichiller 050w-H Technical Datasheet, page 2
Cooling & Media
Cooling water must stay within 5–20 °C and 0.4–6 bar

The water-cooled condenser requires intake water between 5 °C and 20 °C and a differential pressure of at least 0.4 bar but not more than 6 bar. Cooling capacity drops if water temperature rises above 15 °C, and flow rate may need to be increased. Shut off cooling water even for overnight downtime to prevent condensation and corrosion.

Unichiller 050w-H Technical Datasheet, page 2; Operation Manual, page 28
Process Limits
External loop must be closed and hold at least 18 l

The unit is designed only for externally closed applications. The external circuit must have a minimum filling capacity of 18 l and must not exceed the 48 l expansion volume. Exceeding the expansion volume or using an open loop will cause overflow or incorrect temperature control.

Unichiller Operation Manual, page 34; Datasheet, page 2
Safety
Configure power-failure response for process safety

The Pilot ONE allows power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. Choose the mode that matches the process risk before commissioning.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280) with 0.85 kg and 1.2 t CO2 equivalent; room size and charge must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, pages 15, 17

Selection & Integration Q&A

When should I choose the Unichiller 050w-H over the air-cooled Unichiller 030T or 040T siblings?

Choose the 050w-H when a water-cooled condenser is preferred or required, e.g., to avoid room heat rejection or when cooling water is available. Air-cooled siblings like 030T/040T are simpler to install but reject heat to the room and have different noise levels (69 dB(A) vs 62 dB(A) here).

What are the requirements for the cooling water supply and the process loop connections?

Cooling water: G1/2 male, 5–20 °C, 0.4–6 bar differential, flow per datasheet (e.g., 240 l/h at 15 °C water/20 °C flow). Process loop: G1 1/4 AG male pump connections, externally closed, min 18 l fill, max 48 l expansion. Use compatible materials (copper, stainless 1.4401, EPDM, PTFE, etc.).

What installation environment and duty-cycle limits apply to the 050w-H?

Indoors only, ambient 5–40 °C, ≥300 lx illuminance, max 2000 m altitude, 50–200 cm side clearance. Not for ATEX zones. Leveling feet/parking brakes must be set before operation. Shut off cooling water for even short downtimes. Power-failure response configurable.

What maintenance is required and what is not included in the base delivery?

Refrigerant circuit work only by approved contractors. No specific maintenance intervals are given in the digests. Base delivery includes cable but no plug for the 460 V 3~ supply; external UL 489 branch protection must be provided by the installer. No accessories or consumables are listed as included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 050w-H Technical Datasheet
Datasheet ·92,3 КБ ·2 pp.
View PDF