Peter Huber Kältemaschinenbau Unichiller 060T-H
Peter Huber Kältemaschinenbau Unichiller 060T-H — Recirculating chillers.
The Unichiller 060T-H is an air-cooled recirculating chiller with a 2 kW heater and a C3 pump delivering up to 80 l/min at 5.6 bar, covering -20 to 100 °C. It is the 060T-H modification of the Unichiller series, distinguished by its higher capacity and heating function. It serves externally closed process loops requiring precise temperature control with Pilot ONE electronics.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 060T-H Current | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Heating capacity | 2 kW |
| Minimum fill volume | 5 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 235 kg |
| Current consumption | 14 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and positioning
- The 060T-H is a larger-capacity sibling of the 030T/040T models, with a 2 kW heater and a C3 pump (80 l/min, 5.6 bar max).
- It is designed for externally closed applications only; models with 'P' in the name are preferred for high pressure-drop loops.
- Air-cooled condenser using R-449A refrigerant (2.2 kg, CO2 equivalent 3.1 t).
- Pump connection is G1 1/4 male; minimum fill volume 5 l, expansion volume 7.5 l.
Behavior under load
- Temperature range -20 to 100 °C; cooling capacity rated at 20 °C room temperature and derated at higher ambient.
- Pump flow decreases as external pressure drop rises; verify the 80 l/min / 5.6 bar curve against the loop resistance.
- Overtemperature protection per DIN 12876 is adjustable; Process Safety mode can release maximum cooling on OT trigger.
Installation and operation
- Requires 400 V 3~ 50 Hz supply with 3x16 A fusing; cable supplied without plug (under 63 A).
- Voltage and frequency tolerances must be met simultaneously; -5% voltage with +2% frequency is not allowed.
- Ambient 5–40 °C, indoor use only, illuminance ≥300 lx, altitude up to 2000 m.
- Wall clearance 50–200 cm per side for maintenance; leveling feet/parking brakes must be set before start-up.
- Pilot ONE controller with Ethernet, USB (host/device), RS232; external Pt100 sensor input.
- Power-failure response configurable (OFF/Standby, ON, Power Failure Auto).
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate air flow for the air-cooled condenser. Insufficient clearance can lead to reduced cooling performance and difficult service access.
The chiller is delivered with a power cable but without a plug. The customer must provide and install a suitable three-phase plug (e.g., CEE 16 A) matching the site receptacle, and ensure the supply is protected by a 3x16 A breaker.
The unit is designed exclusively for externally closed applications. Using it with an open bath or unsealed circuit will cause fluid loss, poor temperature control, and potential damage from moisture ingress.
The chiller must not be installed or started up inside an ATEX zone because it lacks explosion protection. Additionally, refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.
Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280) with a CO2 equivalent of 3.1 t, requiring proper handling and leak checks for F-gas compliance.
The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, consider Process Safety mode which releases maximum cooling on overtemperature trigger instead of switching off all poles.
Selection & Integration Q&A
Choose the 060T-H when you need higher cooling capacity, a 2 kW heater, and a C3 pump (80 l/min, 5.6 bar) for larger or higher-pressure-drop loops. The 030T/040T are lower-capacity alternatives; the 060T-H also includes heating, unlike the non-H variants.
It requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse. The unit is delivered with a cable but no plug, so you must provide a suitable three-phase plug. Voltage and frequency tolerances must be met simultaneously; -5% voltage with +2% frequency is not allowed.
Install indoors with 50–200 cm clearance on either side for maintenance, ambient 5–40 °C, illuminance ≥300 lx, and altitude up to 2000 m. Leveling feet or parking brakes must be engaged before start-up. Do not install in ATEX zones.
Refrigerant circuit work must be by certified contractors; regular checks of fluid level and expansion tank are needed. The base delivery excludes the three-phase plug and thermal fluid. No specific maintenance intervals are given in the provided documents; follow the manual's maintenance section.