Peter Huber Kältemaschinenbau Unichiller GL 035w
Peter Huber Kältemaschinenbau Unichiller GL 035w — Recirculating chillers.
The Unichiller GL 035w is a water-cooled recirculating chiller with a pressure pump (speed not variable) and Pilot ONE controller, operating from -20 to 40 °C. It is the water-cooled GL variant of the Unichiller series, distinguished from air-cooled 030T/040T siblings by its G1/2 cooling-water circuit (1–6 bar pressure). It serves externally closed process loops requiring heat rejection without an air-cooled condenser.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller GL 035w Current | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure pump, speed not variable |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled (GL) recirculating chiller; cooling water connections G1/2 male, pump connection G1 1/4 AG male.
- Pressure pump with fixed speed (no speed adjustment); external Pt100 sensor input for process-side control.
- Pilot ONE controller with Ethernet, USB (Host and Device) and RS232 interfaces.
- Sibling models 030T, 030Tw, 030Tw-H and 040T are air-cooled variants (noise 61–69 dB(A)); the GL 035w instead rejects heat to a cooling-water circuit, so it has no condenser fan noise figure and depends on facility water.
- Refrigerant R-744 (CO2), GWP 1, safety class A1/H280; unit UN 2857. Protection IP20; safety class I/NFL.
Behavior under load and media
- Temperature range -20…40 °C; ambient 5…40 °C.
- Cooling water intake must be 5…20 °C; as cooling water temperature rises, cooling power drops and required flow rate may increase.
- Cooling water circuit materials: copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM — only compatible water may be used.
- Adjustable overtemperature protection per DIN 12876; Process Safety mode releases maximum cooling capacity at the OT trigger instead of switching off.
Installation, connection and control
- Requires external branch circuit protection per UL 489; three-phase units under 63 A are supplied with cable but without plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%; certain combinations not allowed.
- Power-failure response configurable via Pilot ONE: OFF/Standby, ON, or Power Failure Auto.
- Must be installed indoors, illuminance ≥300 lx, altitude ≤2000 m, not in ATEX zone; wall clearance 50–200 cm per side for maintenance.
- Leveling feet/parking brakes must be set before operation; shut off cooling water even for overnight downtime.
Integration & Operating Notes
The cooling-water circuit requires a minimum differential pressure of 1 bar to reject heat and must never exceed 6 bar, so a pressure-reducing valve is needed on high-pressure facility supplies. Intake water above 20 °C reduces cooling power and may require higher flow; below 5 °C is outside the design range.
The chiller is IP20 and must be installed indoors, outside any ATEX zone, at altitudes up to 2000 m and with illuminance of at least 300 lx. Free-standing models need 50–200 cm wall clearance on either side for maintenance access. Leveling feet or parking brakes must be engaged before startup to prevent movement.
Even short downtimes such as overnight require the cooling-water supply to be shut off to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation, and pipes must be insulated if condensation forms.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles, which must be matched to the process hazard analysis.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-744 (CO2) with GWP 1, classified A1/H280, UN 2857; room volume and charge must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
Selection & Integration Q&A
Choose the GL 035w when facility cooling water (5–20 °C, 1–6 bar differential) is available and condenser heat cannot be rejected to room air. Air-cooled 030T/040T siblings emit 61–69 dB(A) fan noise and require air flow; the GL variant avoids that but depends on water supply and water quality compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Process side uses the G1 1/4 AG male pump connection; cooling water uses G1/2 male connections with 1 bar minimum differential and 6 bar maximum pressure. An external Pt100 sensor can be connected for process control. The 460 V 3~ 60 Hz supply needs a user-supplied plug or hardwiring plus UL 489 branch protection.
Install indoors only, outside ATEX zones, at 5–40 °C ambient, up to 2000 m altitude, with at least 300 lx lighting and 50–200 cm side clearance for maintenance. IP20 requires a dry location. Shut off cooling water even overnight, and protect electrical lines below the unit from condensation.
The base delivery includes a cable but no plug for three-phase supply below 63 A; branch circuit protection per UL 489 and any mating connectors are site-provided. Refrigerant-circuit service must be performed by approved refrigeration contractors. No consumables or spare parts are listed in the supplied documents; consult the manual's maintenance section for intervals.