Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches

Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches — Temperature control units.

The Unistat GL 570w Dynamisches is a water-cooled, CO2-refrigerant (GL) temperature control unit with magnetically coupled stainless-steel pump, covering -50 to 200 °C with 48 kW heating and 400 V 3~ 50 Hz supply. Its 196 l/min pump and Pilot ONE controller distinguish it within the Unistat 5xx series for high-flow, high-pressure-drop external circuits.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat GL 570w Dynamisches Current —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 48 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -50...200 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be 1–4 bar differential, max 6 bar, 5–20 °C

Insufficient differential pressure prevents rated cooling; excessive pressure can damage the water circuit. Rated cooling capacities are valid only at 20 °C room, 15 °C cooling-water inlet and 1 bar differential; higher cooling-water temperatures reduce capacity. Insulate internal cooling-water pipes if water is below condensation point.

Technical Data Sheet, page 2
Mounting
Do not install in ATEX zone; unit is IP20 and EMC Class A

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. IP20 protection and EMC Class A Group 1 mean it must be in a dry, protected industrial location; it may cause interference in residential areas. Ambient must stay 5–40 °C.

Operation Manual, Section 1.3.3 / page 15-16; Technical Data Sheet, page 2
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger is reached the controller releases maximum cooling capacity instead of switching off all poles; this requires the compressor to be set to Permanently ON. Configure the overtemperature protection for the application and test its functionality.

Operation Manual, Section 1.7.2 / page 19-20

Selection & Integration Q&A

When should I choose the Unistat GL 570w Dynamisches over the air-cooled 510w/520w or the 525 Dynamisches?

Choose the 570w when you need water cooling (no air-conditioning load) and CO2 refrigerant (GL), with 48 kW heating and 196 l/min pump for high-flow circuits. Air-cooled siblings suit sites without cooling water; the 525 differs in cooling capacity and noise.

What are the requirements for the cooling-water loop and electrical supply when integrating this unit?

Cooling water: 1–4 bar differential (min 1 bar, max 6 bar), 5–20 °C intake, G3/4 male connections, compatible with copper/stainless/EPDM. Electrical: 400 V 3~ 50 Hz, cable without plug below 63 A, ±5 %/±2 % tolerances not combined.

What mounting and environment limits apply from the manual?

Install upright only, not in ATEX zones, ambient 5–40 °C, IP20 dry location, EMC Class A industrial. Leveling feet/parking brakes engaged; allow 2 h acclimation after cold/warm move. Expansion vessel unlocked for closed, locked for open applications.

What maintenance and accessories are required beyond base delivery?

Base delivery excludes plug (below 63 A), mating hoses/adapters for M38x1.5 and G3/4, and heat transfer fluid (min 8.9 l). Periodic hose checks, cooling-water shutoff overnight, and refrigerant work by approved contractors. Drip tray required by local regulations.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat GL 570w Technical Data Sheet
Datasheet ·98,3 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF