Peter Huber Kältemaschinenbau Unichiller 050w-H
Peter Huber Kältemaschinenbau Unichiller 050w-H — Recirculating chillers with Noise level 62 dB.
The Unichiller 050w-H is a water-cooled recirculating chiller with a 2 kW heater, a fixed-speed E1 pump, and a -20 to 100 °C working range, controlled by the Pilot ONE. It is the 460 V 3~ 60 Hz variant of the Unichiller series, distinguished from air-cooled siblings by its G1/2 cooling-water connection and 6 bar maximum water pressure. It serves externally closed process loops requiring both cooling and active heating within a 48 l expansion volume.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 050w-H Current | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 62 dB(A) |
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 290 kg |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 137 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 167 л/час · at 15°C, flow 0°C: 216 l/h |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled condenser with G1/2 male cooling-water connections; requires 5–20 °C intake water and 0.4–6 bar differential pressure.
- Fixed-speed E1 pump with G1 1/4 AG male process connections; pump speed is not adjustable.
- 48 l expansion volume and 18 l minimum fill volume define the external loop capacity.
- Differs from air-cooled siblings (Unichiller 030T, 030Tw, 030Tw-H, 040T) by using a water circuit instead of fan-cooled condenser; noise level is 62 dB(A).
Behavior under load and media
- Temperature range -20 to 100 °C with 2 kW heating capacity; cooling capacity depends on cooling-water temperature and flow.
- Cooling-water consumption varies with flow temperature: 216 l/h at 0 °C, 167 l/h at -10 °C, 137 l/h at -20 °C (all at 15 °C water).
- Datasheet notes cooling power drops as cooling-water temperature rises, and flow rate may need to increase.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; Process Safety mode releases maximum cooling capacity at the OT trigger instead of switching off.
Installation, control and operation
- Requires 460 V 3~ 60 Hz supply with 3x16 A fuse; external branch circuit protection per UL 489 mandatory; cable supplied without plug for <63 A three-phase devices.
- Pilot ONE controller with Ethernet, USB (Host and Device), and RS232 interfaces; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto.
- Must be installed indoors, 5–40 °C ambient, ≥300 lx illuminance, max 2000 m altitude, with 50–200 cm wall clearance per side for maintenance.
- Not explosion-protected; must not be installed in ATEX zones. Refrigerant R-449A (A1, H280), 0.85 kg, 1.2 t CO2 equivalent; room size and charge must be checked.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access to the refrigerant circuit and pump. Insufficient clearance blocks service work and may prevent proper airflow around the unit.
The 460 V 3~ 60 Hz supply must be protected by an external branch circuit breaker according to UL 489. The unit draws up to 3x16 A; the cable is supplied without a plug, so the installer must provide the final connection and ensure voltage/frequency tolerances of ±5% and ±2% are not exceeded simultaneously in opposite directions.
The water-cooled condenser requires intake water between 5 °C and 20 °C and a differential pressure of at least 0.4 bar but not more than 6 bar. Cooling capacity drops if water temperature rises above 15 °C, and flow rate may need to be increased. Shut off cooling water even for overnight downtime to prevent condensation and corrosion.
The unit is designed only for externally closed applications. The external circuit must have a minimum filling capacity of 18 l and must not exceed the 48 l expansion volume. Exceeding the expansion volume or using an open loop will cause overflow or incorrect temperature control.
The Pilot ONE allows power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. Choose the mode that matches the process risk before commissioning.
Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280) with 0.85 kg and 1.2 t CO2 equivalent; room size and charge must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
Selection & Integration Q&A
Choose the 050w-H when a water-cooled condenser is preferred or required, e.g., to avoid room heat rejection or when cooling water is available. Air-cooled siblings like 030T/040T are simpler to install but reject heat to the room and have different noise levels (69 dB(A) vs 62 dB(A) here).
Cooling water: G1/2 male, 5–20 °C, 0.4–6 bar differential, flow per datasheet (e.g., 240 l/h at 15 °C water/20 °C flow). Process loop: G1 1/4 AG male pump connections, externally closed, min 18 l fill, max 48 l expansion. Use compatible materials (copper, stainless 1.4401, EPDM, PTFE, etc.).
Indoors only, ambient 5–40 °C, ≥300 lx illuminance, max 2000 m altitude, 50–200 cm side clearance. Not for ATEX zones. Leveling feet/parking brakes must be set before operation. Shut off cooling water for even short downtimes. Power-failure response configurable.
Refrigerant circuit work only by approved contractors. No specific maintenance intervals are given in the digests. Base delivery includes cable but no plug for the 460 V 3~ supply; external UL 489 branch protection must be provided by the installer. No accessories or consumables are listed as included.