Peter Huber Kältemaschinenbau Unichiller 060T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 060T-H

Peter Huber Kältemaschinenbau Unichiller 060T-H — Recirculating chillers.

The Unichiller 060T-H is an air-cooled recirculating chiller with a 2 kW heater and a C3 pump delivering up to 80 l/min at 5.6 bar, covering -20 to 100 °C. It is the 060T-H modification of the Unichiller series, distinguished by its higher capacity and heating function. It serves externally closed process loops requiring precise temperature control with Pilot ONE electronics.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 060T-H Current —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Heating capacity 2 kW
Minimum fill volume 5 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 235 kg
Electrical, Power & Drive Specifications
Current consumption 14 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and positioning

Behavior under load

Installation and operation

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for service access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate air flow for the air-cooled condenser. Insufficient clearance can lead to reduced cooling performance and difficult service access.

Unichiller Operation Manual, Section on Installation
Electrical
Plug not included for three-phase units under 63 A

The chiller is delivered with a power cable but without a plug. The customer must provide and install a suitable three-phase plug (e.g., CEE 16 A) matching the site receptacle, and ensure the supply is protected by a 3x16 A breaker.

Unichiller 060T-H Datasheet, Technical Specifications
Process Limits
Only for externally closed loops; not for open baths

The unit is designed exclusively for externally closed applications. Using it with an open bath or unsealed circuit will cause fluid loss, poor temperature control, and potential damage from moisture ingress.

Unichiller Operation Manual, Section on Selection
Process Limits
Do not install in ATEX zones; unit is not explosion-protected

The chiller must not be installed or started up inside an ATEX zone because it lacks explosion protection. Additionally, refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Section on Installation
Maintenance
Refrigerant circuit work only by certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280) with a CO2 equivalent of 3.1 t, requiring proper handling and leak checks for F-gas compliance.

Unichiller Operation Manual, Section on Maintenance
Operation
Configure power-failure response to match process risk

The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, consider Process Safety mode which releases maximum cooling on overtemperature trigger instead of switching off all poles.

Unichiller Operation Manual, Section on Operation

Selection & Integration Q&A

When should I choose the Unichiller 060T-H instead of a smaller sibling like the 030T or 040T?

Choose the 060T-H when you need higher cooling capacity, a 2 kW heater, and a C3 pump (80 l/min, 5.6 bar) for larger or higher-pressure-drop loops. The 030T/040T are lower-capacity alternatives; the 060T-H also includes heating, unlike the non-H variants.

What are the requirements for the electrical supply and plug for the 060T-H?

It requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse. The unit is delivered with a cable but no plug, so you must provide a suitable three-phase plug. Voltage and frequency tolerances must be met simultaneously; -5% voltage with +2% frequency is not allowed.

What installation clearances and ambient conditions are mandatory for the 060T-H?

Install indoors with 50–200 cm clearance on either side for maintenance, ambient 5–40 °C, illuminance ≥300 lx, and altitude up to 2000 m. Leveling feet or parking brakes must be engaged before start-up. Do not install in ATEX zones.

What maintenance and consumables are required, and what is not included in the base delivery?

Refrigerant circuit work must be by certified contractors; regular checks of fluid level and expansion tank are needed. The base delivery excludes the three-phase plug and thermal fluid. No specific maintenance intervals are given in the provided documents; follow the manual's maintenance section.