Peter Huber Kältemaschinenbau Unichiller 075w-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 075w-H

Peter Huber Kältemaschinenbau Unichiller 075w-H — Recirculating chillers.

The Unichiller 075w-H is a water-cooled recirculating chiller for externally closed applications, rated -20...100 °C with 2 kW heating and a fixed-speed E1 pump. It is distinguished within the series by its water-cooled condenser, requiring a 5...20 °C cooling water supply at 0.5...6 bar. It serves as a temperature-control source for closed process loops in indoor industrial settings.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller 075w-H Current —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 300 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 240 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 270 л/час · at 15°C, flow 0°C: 330 l/h
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must stay within 5–20 °C and 0.5–6 bar

The water-cooled condenser requires a facility cooling water supply of 5...20 °C intake temperature and a differential pressure of at least 0.5 bar, with 6 bar maximum. Below the condensation point the internal cooling water pipes must be insulated to prevent condensation and electrical risk.

Datasheet, Technical Specifications
Electrical
Provide a 3x16 A fused 400 V 3~ supply and a plug

The unit draws up to 12.5 A per phase on 400 V 3~ 50 Hz and is delivered with cable but without a plug. The installer must provide a suitable plug or hardwired connection and 3x16 A fusing. Voltage tolerance is ±5% only with simultaneous frequency tolerance ±2%.

Datasheet, Technical Specifications
Mounting
Keep 50–200 cm clearance per side and lock the casters

Free-standing installation requires a wall clearance of 50 to 200 cm on either side for maintenance access. Before start-up, parking brakes at the casters must be activated and/or leveling feet unscrewed so the 300 kg unit cannot move during operation.

Operation Manual, Installation and Commissioning
Process Limits
Use only with externally closed circuits; no ATEX zones

The chiller is designed for externally closed applications only and must not be installed or started inside an ATEX zone because it is not explosion-protected. Models with 'P' in the name are the ones suited to high pressure-drop loops; this model is not so designated.

Operation Manual, Function Description / Selection
Maintenance
Refrigerant circuit work only by approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains 0.95 kg R-449A (GWP 1397, CO2 equivalent 1.3 t), so leak checks and environmental reporting apply. Shut off cooling water even for overnight downtimes.

Operation Manual, Maintenance / Operation

Selection & Integration Q&A

When should I choose the 075w-H over the air-cooled Unichiller 030T/030Tw siblings?

Choose the 075w-H when a facility cooling water circuit (5–20 °C, 0.5–6 bar) is available and heat must not be rejected to room air; it also adds 2 kW heating. The 030T/030Tw are air-cooled and lack the water connection, so they suit sites without cooling water.

What adjacent system parts must be provided for the cooling water and process loop?

Provide a cooling water supply at 5–20 °C and 0.5–6 bar with G1/2 male mating connection, and a closed process loop with G1 1/4 AG male pump fittings. The 400 V 3~ supply needs 3x16 A fusing and a plug (not delivered). Use only water compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.

What mounting and environment limits apply from the manual?

Indoor use only, illuminance ≥300 lx, altitude ≤2000 m, ambient 5–40 °C, IP20. Allow 50–200 cm wall clearance per side for maintenance, activate caster brakes or leveling feet before start-up, and protect electrical lines below the unit against condensation ingress.

What maintenance and consumables are not included in base delivery?

The unit ships with cable but without a plug; the installer supplies the plug or hardwired connection. Refrigerant-circuit service must be by approved contractors. Shut off cooling water even overnight. Insulate internal cooling water pipes if water falls below the condensation point. No accessories or cables beyond the power cable are included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 075w-H Recirculating Chiller Datasheet
Datasheet ·92,2 КБ ·2 pp.
View PDF