Peter Huber Kältemaschinenbau Unichiller GL 035w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 035w

Peter Huber Kältemaschinenbau Unichiller GL 035w — Recirculating chillers.

The Unichiller GL 035w is a water-cooled recirculating chiller with a pressure pump (speed not variable) and Pilot ONE controller, operating from -20 to 40 °C. It is the water-cooled GL variant of the Unichiller series, distinguished from air-cooled 030T/040T siblings by its G1/2 cooling-water circuit (1–6 bar pressure). It serves externally closed process loops requiring heat rejection without an air-cooled condenser.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller GL 035w Current —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Process Limits
Cooling water must stay within 1–6 bar and 5–20 °C

The cooling-water circuit requires a minimum differential pressure of 1 bar to reject heat and must never exceed 6 bar, so a pressure-reducing valve is needed on high-pressure facility supplies. Intake water above 20 °C reduces cooling power and may require higher flow; below 5 °C is outside the design range.

Unichiller GL 035w Datasheet, Specifications and Operation
Mounting
Indoor site with 50–200 cm side clearance and ≥300 lx lighting

The chiller is IP20 and must be installed indoors, outside any ATEX zone, at altitudes up to 2000 m and with illuminance of at least 300 lx. Free-standing models need 50–200 cm wall clearance on either side for maintenance access. Leveling feet or parking brakes must be engaged before startup to prevent movement.

Unichiller Operation Manual, Installation
Cooling & Media
Shut off cooling water during overnight downtime

Even short downtimes such as overnight require the cooling-water supply to be shut off to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation, and pipes must be insulated if condensation forms.

Unichiller Operation Manual, Installation and Operation
Control & I/O
Configure power-failure response for the process risk

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles, which must be matched to the process hazard analysis.

Unichiller Operation Manual, Setup Mode and Safety
Maintenance
Refrigerant work only by approved refrigeration contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-744 (CO2) with GWP 1, classified A1/H280, UN 2857; room volume and charge must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Maintenance and Installation

Selection & Integration Q&A

When should I choose the water-cooled GL 035w instead of an air-cooled sibling such as the 030T or 040T?

Choose the GL 035w when facility cooling water (5–20 °C, 1–6 bar differential) is available and condenser heat cannot be rejected to room air. Air-cooled 030T/040T siblings emit 61–69 dB(A) fan noise and require air flow; the GL variant avoids that but depends on water supply and water quality compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

What is required to connect the GL 035w to the process loop and the cooling-water supply?

Process side uses the G1 1/4 AG male pump connection; cooling water uses G1/2 male connections with 1 bar minimum differential and 6 bar maximum pressure. An external Pt100 sensor can be connected for process control. The 460 V 3~ 60 Hz supply needs a user-supplied plug or hardwiring plus UL 489 branch protection.

What installation environment and duty limits apply to the GL 035w?

Install indoors only, outside ATEX zones, at 5–40 °C ambient, up to 2000 m altitude, with at least 300 lx lighting and 50–200 cm side clearance for maintenance. IP20 requires a dry location. Shut off cooling water even overnight, and protect electrical lines below the unit from condensation.

What maintenance and accessory items are not included with the base unit?

The base delivery includes a cable but no plug for three-phase supply below 63 A; branch circuit protection per UL 489 and any mating connectors are site-provided. Refrigerant-circuit service must be performed by approved refrigeration contractors. No consumables or spare parts are listed in the supplied documents; consult the manual's maintenance section for intervals.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 035w Datasheet
Datasheet ·30,4 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF