Peter Huber Kältemaschinenbau Unichiller GL 075w
Peter Huber Kältemaschinenbau Unichiller GL 075w — Recirculating chillers.
The Unichiller GL 075w is a water-cooled recirculating chiller rated at 7.5 kW cooling capacity at 20 °C, operating from -20 to 40 °C with a fixed-speed pressure pump delivering up to 50 l/min at 3.5 bar. It is the 7.5 kW water-cooled modification in the Unichiller series, distinguished from air-cooled siblings by its G1/2 cooling-water circuit and 460 V 3~ 60 Hz supply. It serves externally closed process loops requiring stable temperature control with Pilot ONE electronics.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller GL 075w Current | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure pump, speed not variable |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The GL 075w is a water-cooled recirculating chiller in the Unichiller series. Unlike the air-cooled 030T/030Tw/040T siblings, heat is rejected to a cooling-water circuit via G1/2 male connections rather than to ambient air, so the unit requires a cooling-water supply with at least 1 bar differential pressure and no more than 6 bar supply pressure. The refrigerant is R-744 (A1, H280). The process loop connects through a G1 1/4 male pump connection; the pump is a fixed-speed pressure pump (not speed-variable) with maximum delivery 50 l/min and maximum delivery pressure 3.5 bar. Electrical supply is 460 V 3~ 60 Hz; the unit is IP20 and carries safety class I / NFL.
Behavior under load and media
Cooling power is rated at 7.5 kW at 20 °C. The datasheet states that higher cooling-water temperatures reduce cooling capacity and may require increased cooling-water flow, so the 7.5 kW figure is tied to the 20 °C condition. Cooling-water inlet temperature must be between 5 °C and 20 °C; below the condensation point, cooling-water pipes inside the unit must be insulated. The cooling-water circuit materials are copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water chemistry should be used. The process temperature range is -20 to 40 °C, and the unit is intended only for externally closed applications.
Control, installation and operation
The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, an external Pt100 sensor connection, and configurable power-failure response (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. Installation requires indoor use at illuminance of at least 300 lx and up to 2000 m altitude, with 50 to 200 cm wall clearance per side for maintenance on free-standing models. Leveling feet must be activated and parking brakes applied before startup. Electrical lines directly below the water-cooled unit must be protected against liquid ingress from condensation. The cooling-water supply must be shut off even for short downtimes such as overnight.
Integration & Operating Notes
The datasheet specifies a cooling-water intake temperature of 5 to 20 °C. Below the condensation point, cooling-water pipes inside the unit must be insulated. Higher cooling-water temperatures reduce cooling power below the 7.5 kW rating at 20 °C and may require increased cooling-water flow, so site water temperature must be checked against the required duty.
The cooling-water circuit requires a minimum differential pressure of 1 bar and tolerates a maximum supply pressure of 6 bar at the G1/2 male connections. Exceeding the maximum supply pressure risks damage to the cooling-water circuit components; insufficient differential pressure prevents adequate heat rejection.
The datasheet requires external branch circuit protection according to UL 489. Three-phase devices with current consumption below 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug or hardwired connection. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not permitted.
The manual restricts installation to indoors at illuminance of at least 300 lx and up to 2000 m altitude, with 50 to 200 cm wall clearance per side for maintenance on free-standing models. The unit must not be installed or started up inside an ATEX zone because it is not explosion-protected. Refrigerant charge and room size must be checked against the refrigerant limit per m³ of ambient air.
Because the unit is water-cooled, condensate can form. Electrical lines directly below the temperature control unit must be protected against liquid ingress. Pump connections must be counterheld and the maximum tightening torques in the manual table must not be exceeded during hose installation.
The manual requires shutting off the cooling-water supply to the unit even for shorter downtimes such as overnight, to prevent water damage and corrosion during idle periods. Before switching on, parking brakes must be activated at the casters (if any) and/or leveling feet unscrewed/activated to prevent the unit from moving.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. The chosen behavior determines whether the process restarts automatically after a power outage. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit uses R-744 refrigerant, classified A1 but carrying GHS hazard statement H280 (contains gas under pressure), so service personnel must be aware of the pressure hazard and proper handling procedures.
Selection & Integration Q&A
Choose the GL 075w when a 7.5 kW cooling capacity at 20 °C is needed and a cooling-water supply (1–6 bar, 5–20 °C inlet) is available, avoiding ambient heat rejection. The air-cooled 030T/030Tw/040T siblings suit sites without cooling water but have lower capacity and published noise levels of 61–69 dB(A).
Provide cooling water at 5–20 °C inlet, minimum 1 bar differential pressure and maximum 6 bar supply pressure, connected via G1/2 male fittings. The process loop connects to a G1 1/4 male pump connection; the fixed-speed pump delivers up to 50 l/min at up to 3.5 bar. Use water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors only, at illuminance of at least 300 lx and up to 2000 m altitude, with 50 to 200 cm wall clearance per side for maintenance on free-standing models. The unit is not explosion-protected and must not be placed in an ATEX zone. Check refrigerant charge against room volume per m³ of ambient air.
Refrigerant circuit work must be performed by approved refrigeration/air-conditioning contractors; R-744 is under pressure (H280). Shut off cooling water even overnight. Three-phase units below 63 A ship with cable but no plug, and external UL 489 branch protection must be supplied by the installer. No maintenance interval is stated in the provided documents.