Peter Huber Kältemaschinenbau Unichiller P050w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P050w

Peter Huber Kältemaschinenbau Unichiller P050w — Recirculating chillers with Noise level 62 dB.

The Unichiller P050w is a water-cooled recirculating chiller with a Pilot ONE controller, covering -20 to 40 °C and requiring a 400 V 3~ 50 Hz supply. Its 'P' designation marks it for high-pressure-drop external circuits, distinguishing it from T-series siblings. It serves closed-loop process cooling where heat loads fall as setpoint drops.

Noise level
62 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller P050w Current 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Noise level 62 dB(A)
Expansion volume 48 l
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E2
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 300 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 137 л/час (at 15°C, flow -20°C)
at 15°C, flow -10°C: 167 л/час · at 15°C, flow 0°C: 216 l/h
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The P050w is a water-cooled, free-standing chiller in the Unichiller series. Unlike the T-series siblings (030T, 030Tw, 030Tw-H, 040T), the 'P' in its model name designates it for applications requiring high pressure drops, per the operation manual. The unit is IP20, weighs 300 kg, and measures 740x1160x1050 mm. It uses refrigerant R-449A (A1, H280) with a 0.85 kg charge (1.2 t CO2 equivalent). The pump is a fixed-speed E2 type with G1 1/4 AG male connections; cooling water connections are G1/2 male.

Thermal and hydraulic behavior

Cooling capacity falls steeply with setpoint: 5 kW at +20 °C, dropping to 0.8 kW at -20 °C (at +20 °C ambient). The pump delivers up to 130 l/min, but only 20 l/min at 5.5 bar, so system pressure drop must be matched to the pump curve. Cooling water consumption depends on flow temperature: 216 l/h at 0 °C, 167 l/h at -10 °C, and 137 l/h at -20 °C, all at 15 °C intake. The unit requires a minimum fill volume of 18 l and provides 48 l expansion volume; it is intended for externally closed applications only.

Installation, control and operation

The Pilot ONE controller offers Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off. The unit must be installed indoors at 5–40 °C ambient, with 50–200 cm wall clearance per side for maintenance, and leveling feet or parking brakes engaged before operation. Electrical lines below the unit must be protected from condensate.

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ with 3x10 A fusing and a site-fitted plug

The chiller is delivered with a cable but without a plug for three-phase devices below 63 A. The electrical contractor must provide a suitable mating plug and protection rated 3x10 A. Voltage tolerance is ±5 % only with simultaneous frequency tolerance of ±2 %; certain voltage/frequency combinations are not permitted and could damage the unit.

Unichiller P050w Datasheet, page 2
Cooling & Media
Cooling water must stay within 5–20 °C and 0.4–6 bar differential

Intake temperature outside 5–20 °C degrades cooling power and risks condensation; pipes must be insulated if the water can drop below the dew point. Differential pressure must be at least 0.4 bar and must not exceed 6 bar. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Unichiller P050w Datasheet, page 2
Process Limits
External circuit must be closed and sized for at least 18 l fill

The unit is intended for externally closed applications only; open-loop or atmospheric circuits are not permitted. The external circuit volume must be planned with a minimum filling capacity of 18 l and the 48 l expansion volume in mind. The 'P' designation means the unit suits high-pressure-drop loops, but the pump curve (130 l/min max, 20 l/min at 5.5 bar) must still be checked against actual system resistance.

Unichiller Operation Manual, page 34; Datasheet page 1

Selection & Integration Q&A

When should I choose the P050w over the T-series siblings like the 030T or 040T?

Choose the P050w when your external circuit has high pressure drop, since the 'P' designation marks it for such applications. The T-series siblings are not specified for high pressure drop. Also compare cooling capacity at your setpoint: the P050w delivers 5 kW at +20 °C but only 0.8 kW at -20 °C.

What are the requirements for the cooling water supply and the process loop connections?

Cooling water: G1/2 male connections, 5–20 °C intake, 0.4–6 bar differential pressure, and consumption of 216 l/h at 0 °C flow (137 l/h at -20 °C). Process loop: G1 1/4 AG male pump connections, externally closed circuit, minimum fill 18 l, 48 l expansion volume. Only cooling water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

What are the mounting and ambient limits for the P050w?

Install indoors only, at 5–40 °C ambient, up to 2000 m altitude, with illuminance at least 300 lx. Not permitted in ATEX zones. Allow 50–200 cm wall clearance per side for maintenance. Engage leveling feet or parking brakes before operation. Protect electrical lines below the unit from condensate.

What maintenance and service items are required, and what is not included in the base delivery?

Refrigerant circuit work must be done by approved refrigeration contractors. Shut off cooling water even for overnight downtimes. The base delivery includes a cable but no plug for the 400 V 3~ supply; a suitable plug must be provided on site. No other accessories or consumables are listed in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P050w Datasheet
Datasheet ·92,3 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF