Peter Huber Kältemaschinenbau Unichiller P050w-H
Peter Huber Kältemaschinenbau Unichiller P050w-H — Recirculating chillers with Noise level 62 dB.
The Unichiller P050w-H is a water-cooled recirculating chiller with a 5 kW cooling capacity at 20 °C and 100 °C, covering a -20 to 100 °C working range. Its 'P' designation and E2 pump suit externally closed loops with high pressure drop, distinguishing it from the T-series siblings. It serves as a central cooling source for laboratory and pilot-scale processes.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P050w-H Current | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 62 dB(A) |
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 307 kg |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 5 kW (at 100°C) at 20°C: 5 kW |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 137 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 167 л/час · at 15°C, flow 0°C: 216 l/h |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The P050w-H is a water-cooled, free-standing recirculating chiller in the Unichiller series. Unlike the air-cooled 030T/040T siblings (69 dB(A)), this water-cooled variant is rated at 62 dB(A) and requires a G1/2 male cooling-water connection. The 'P' in the model name marks it as particularly suitable for applications with high pressure drops, per the operation manual. The E2 pump is fixed-speed (not adjustable) and delivers up to 130 l/min at 5.7 bar maximum delivery pressure, with flow falling sharply at higher pressures. The 48 l expansion volume and 18 l minimum fill volume define the hydraulic circuit sizing.
Behavior and media
Cooling capacity is 5 kW at both 20 °C and 100 °C. The unit is designed for externally closed applications only. Cooling-water consumption depends on the required flow temperature: 216 l/h at 0 °C, 167 l/h at -10 °C, and 137 l/h at -20 °C, all at a 15 °C cooling-water intake. As cooling-water temperature rises, cooling power drops and the required flow rate may increase. The cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. The refrigerant is R-449A (A1, H280).
Control and operation
The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. In Process Safety mode, maximum cooling capacity is released when the overtemperature trigger value is reached, rather than switching off all poles. The unit is rated safety class III/FL and IP20, and is classified Class A, Group 1 for electromagnetic disturbance to EN55011.
Integration & Operating Notes
The chiller requires a 400 V 3~ 50 Hz supply with a 3x16 A fuse. Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% and frequency tolerance ±2%, but these must not be combined in opposite directions.
The cooling-water connection is G1/2 male. Minimum differential pressure is 0.4 bar and maximum pressure is 6 bar, so a pressure reducer may be needed on the building water line. Intake temperature must be between 5 °C and 20 °C; lower temperatures require insulating the cooling-water pipes. Shut off the cooling-water supply even for short downtimes such as overnight.
The unit is IP20 and must be installed in a dry indoor location protected from water and dust, at an illuminance of at least 300 lx and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance. Leveling feet must be activated and parking brakes applied before start-up. Electrical lines directly below the unit must be protected against liquid ingress from condensation.
The unit is designed for externally closed applications and must not be installed or started up inside an ATEX zone because it is not explosion-protected. Refrigerant charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air. Ambient temperature must stay between 5 °C and 40 °C.
The Pilot ONE allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto, which determines whether the process restarts automatically after an outage. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.
Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The cooling-water circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) dictate compatible water quality to avoid corrosion and fouling. Maximum tightening torques for pump connections must not be exceeded and connections must be counterheld.
Selection & Integration Q&A
Choose the P050w-H for high-pressure-drop loops (the 'P' designation) or where water cooling and lower noise (62 dB(A) vs 69 dB(A)) are needed. The 030T/040T are air-cooled alternatives; select them only if a cooling-water supply is unavailable.
A G1/2 male connection with 5–20 °C intake, 0.4–6 bar differential pressure, and a pressure reducer if line pressure exceeds 6 bar. Insulate pipes if intake is below 5 °C. The unit is water-cooled and requires this supply for operation.
Indoor use only, IP20, ambient 5–40 °C, illuminance ≥300 lx, altitude ≤2000 m, and 50–200 cm wall clearance per side. Not for ATEX zones. Shut off cooling water even for overnight downtime. Leveling feet and parking brakes must be set before start-up.
A suitable three-phase plug must be procured (cable supplied, plug not). Refrigerant circuit service only by approved contractors. No consumables or cables are listed as included beyond the power cable. Cooling-water circuit compatibility with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM must be verified.