Peter Huber Kältemaschinenbau Unichiller P075w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P075w

Peter Huber Kältemaschinenbau Unichiller P075w — Recirculating chillers.

The Unichiller P075w is a water-cooled recirculating chiller with a 400 V three-phase supply, designed for externally closed circuits and rated for -20 to 40 °C. Its 'P' designation marks it for high-pressure-drop applications, and the E2 pump delivers up to 130 l/min at 5.7 bar. It serves as a central cooling source for process loops requiring stable temperature control under significant hydraulic resistance.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller P075w Current —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 050w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E2
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 301 kg
Electrical, Power & Drive Specifications
Current consumption 10 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 240 l/h (cooling water at 15°C, flow -20°C)
cooling water at 15°C, flow -10°C: 270 l/h · cooling water at 15°C, flow 0°C: 330 l/h
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and Series Positioning

The Unichiller P075w is a free-standing, water-cooled recirculating chiller in the Unichiller series. Unlike the air-cooled 030T/030Tw siblings (69/61 dB(A)), this model uses a cooling water circuit with G1/2 male connections and requires a site supply of 5–20 °C water at 0.5–6 bar differential pressure. The 'P' in the model name specifically denotes suitability for applications with high pressure drop, distinguishing it from standard variants. The refrigerant is R-449A (GWP 1397, charge 0.95 kg, 1.3 t CO2e), and the unit is classified I/NFL per DIN 12876 for externally closed applications.

Hydraulic and Thermal Behavior

The E2 pump is fixed-speed (not adjustable) and delivers up to 130 l/min at a maximum 5.7 bar, but flow falls sharply as external pressure drop rises—system resistance must be checked against the pump curve. Cooling water consumption depends on the required flow temperature: 330 l/h at 0 °C, 270 l/h at -10 °C, and 240 l/h at -20 °C, all referenced to 15 °C cooling water. Warmer cooling water reduces cooling capacity and may increase the required flow rate. The unit has a 48 l expansion volume and requires a minimum fill of 18 l; the external loop must be closed and sized accordingly.

Installation, Control and Operation

The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto), and Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached—critical for exothermic processes. The unit is IP20 and must be installed indoors in a dry environment at 5–40 °C ambient, with at least 300 lx illuminance and a wall clearance of 50–200 cm per side for maintenance. Leveling feet or parking brakes must be engaged before startup, and cooling water must be shut off even for overnight downtimes.

Integration & Operating Notes

Electrical
Three-phase supply must be sized for 3x10 A with no plug included

The 400 V 3~ 50 Hz supply requires a 3x10 A fuse or breaker, and the unit is delivered with a cable but without a plug for currents below 63 A. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not permitted—verify site power quality before commissioning.

Unichiller P075w Datasheet, Technical Specifications
Cooling & Media
Cooling water must stay within 5–20 °C and 0.5–6 bar differential

Inlet water below 5 °C or above 20 °C is outside the design envelope; pressure below 0.5 bar risks insufficient flow, while above 6 bar may damage the circuit. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water without aggressive additives may be used. Insulate pipes where condensation can form.

Unichiller P075w Datasheet, Installation Requirements
Process Limits
External circuit must be closed and pump pressure drop checked

The unit is designed only for externally closed applications; open baths or unsealed loops are not permitted. The E2 pump delivers up to 130 l/min at 5.7 bar, but flow drops sharply at high pressure, so calculate the external circuit pressure drop to ensure the required flow is achievable. Minimum fill volume is 18 l and expansion volume is 48 l.

Unichiller P075w Datasheet, Technical Specifications; Operation Manual, Function Description

Selection & Integration Q&A

When should I choose the Unichiller P075w over the air-cooled 030T/030Tw siblings?

Choose the P075w when the process loop has high pressure drop (the 'P' designation) and a cooling water supply is available. The air-cooled 030T/030Tw are simpler to install but have lower pump pressure capability and produce 61–69 dB(A) noise; the P075w is water-cooled and requires 5–20 °C cooling water at 0.5–6 bar.

What are the requirements for the cooling water supply and the external process loop?

Cooling water: 5–20 °C inlet, 0.5–6 bar differential, G1/2 male connection, compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Process loop: externally closed only, minimum fill 18 l, expansion volume 48 l, pump connection G1 1/4 AG male, and pressure drop must be checked against the E2 pump curve (max 130 l/min at 5.7 bar).

What maintenance and service items are required, and what is not included in the base delivery?

Refrigerant circuit work must be done by approved contractors; shut off cooling water even overnight. The unit is delivered with a cable but without a plug (for <63 A), so a suitable plug must be provided on site. No external Pt100 sensor is included unless specified; the Pilot ONE interfaces (Ethernet, USB, RS232) are built-in. Regular maintenance includes checking cooling water quality and pipe insulation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P075w Datasheet
Datasheet ·92,2 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF