Peter Huber Kältemaschinenbau Unichiller P100w
Peter Huber Kältemaschinenbau Unichiller P100w — Recirculating chillers with Noise level 58 dB.
The Unichiller P100w is a water-cooled recirculating chiller with a 400 V three-phase supply, rated for -20 to 40 °C and built for externally closed circuits. Its 'P' designation targets high-pressure-drop applications, delivering up to 130 l/min at 5.7 bar. It serves as a central cooling source for laboratory or process loops requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P100w Current | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 58 dB(A) |
| Expansion volume | 48 l |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 311 kg |
| Current consumption | 13 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 240 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 270 л/час · at 15°C, flow 0°C: 330 l/h |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and Series Positioning
The Unichiller P100w is a water-cooled, three-phase recirculating chiller in the Unichiller series. Unlike air-cooled siblings such as the Unichiller 030T (69 dB(A)) and 030Tw (61 dB(A)), this model uses a cooling water circuit with G1/2 male connections and consumes 330 l/h at 15 °C water inlet and 0 °C flow. The 'P' in the model name indicates suitability for applications with high pressure drops, per the operation manual. The unit weighs 311 kg and has IP20 protection, an atmospheric open expansion tank (48 l), and a minimum fill volume of 18 l.
Behavior Under Load and Media
Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet, and 0.5 bar differential pressure. Cooling water consumption rises as flow temperature decreases: 330 l/h at 0 °C, 270 l/h at -10 °C, and 240 l/h at -20 °C. The pump delivers up to 130 l/min at 5.7 bar, but flow falls sharply at high pressure (e.g., 20 l/min at 5.5 bar), so external circuit pressure drop must be checked. The unit is intended only for externally closed applications; the expansion tank is open to atmosphere. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, restricting water quality.
Installation, Control, and Operation
The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. The unit requires 400 V 3~ 50 Hz supply with a 3x16 A fuse; three-phase devices under 63 A are delivered with cable but without plug. Cooling water pressure must be between 0.5 and 6 bar, and inlet temperature between 5 and 20 °C. Installation requires 50–200 cm wall clearance per side, indoor use only, and protection of electrical lines below the unit against condensate.
Integration & Operating Notes
The unit draws up to 13 A per phase and requires a 3x16 A fuse. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%; opposite-sign deviations are not permitted. The customer must provide a mating plug because the unit ships with cable only.
The chiller is designed for cooling water intake between 5 °C and 20 °C. Higher inlet temperatures reduce capacity, and published performance assumes 15 °C inlet and 0.5 bar differential pressure. Pipes must be insulated where condensation can occur.
The unit is intended only for externally closed applications; the expansion tank is open to atmosphere. The pump delivers 130 l/min at 5.7 bar but only 20 l/min at 5.5 bar, so the external loop's pressure drop must be matched to avoid insufficient flow.
Free-standing models require 50 to 200 cm clearance on either side for maintenance access. The unit must be installed indoors at illuminance of at least 300 lx and up to 2000 m altitude. Leveling feet or parking brakes must be engaged before startup.
The refrigerant circuit contains R-449A (1.03 kg, 1.4 t CO2 equivalent) and may only be serviced by approved refrigeration/air-conditioning contractors. Cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage.
The Pilot ONE allows power-failure response OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles, preventing runaway temperatures.
Selection & Integration Q&A
Choose the P100w for high-pressure-drop applications (P designation) and where a water-cooled circuit is available; it requires 330 l/h cooling water at 0 °C flow. Air-cooled 030T/030Tw suit sites without cooling water but have higher noise (69/61 dB(A)).
Cooling water: 5–20 °C inlet, 0.5–6 bar differential pressure, G1/2 male connections, and materials compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM. External loop must be closed, with pressure drop checked against the pump curve (130 l/min at 5.7 bar; 20 l/min at 5.5 bar).
Indoors only, 5–40 °C ambient, IP20, 50–200 cm wall clearance per side, illuminance ≥300 lx, altitude ≤2000 m. Not for ATEX zones. Shut off cooling water even overnight. Leveling feet/parking brakes must be engaged before operation.
Refrigerant circuit work only by certified contractors; R-449A charge 1.03 kg (1.4 t CO2e). The unit ships with cable but without a plug for three-phase <63 A. No external Pt100 sensor or process hoses are included; these must be sourced separately.