Peter Huber Kältemaschinenbau Unistat 520w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 520w Dynamisches

Peter Huber Kältemaschinenbau Unistat 520w Dynamisches — Temperature control units with Noise level 61 dB.

Water-cooled (w) free-standing Unistat 520w with magnetically coupled stainless-steel pump, covering -55 to 250 °C with 6 kW heating and 400 V three-phase supply. It sits between the 515w and 525w in the series, distinguished by water cooling and dynamic performance. Used for externally closed or open temperature control of reactors and process equipment.

Noise level
61 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 520w Dynamisches Current 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 620w Dynamisches 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 61 dB(A)
Control output POKO ONE
Expansion volume 16 l
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -55...250 °C
Dimensions, Weight & Mechanical Design
Weight 210 kg
Electrical, Power & Drive Specifications
Current consumption 14 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 492 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Electrical
Leakage current above 3.5 mA affects RCD and earthing design

The datasheet explicitly warns that leakage current exceeds 3.5 mA, so protective earthing must be reliable and residual current protective devices must be selected with this leakage in mind; a standard 30 mA RCD may not be appropriate without engineering review.

Unistat 520w Technical Data, page 2
Mounting
Unit must not be installed in an ATEX zone

The operation manual states the temperature control unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet with a separate ATEX annex, and only the model/serial number on the Ex px cabinet rating plate is suitable. Site selection must exclude hazardous areas.

Unistat Operation Manual — Unistat 5, Section 1.3.3
Mounting
Transport upright only and allow 2 h acclimation before power-on

The manual requires upright transport and storage only; horizontal transport damages the compressor. Only an emptied unit may be moved. When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Casters are not for transport and carry only 25% of total mass symmetrically.

Unistat Operation Manual, pages 22 and 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON in the controller configuration; otherwise the safety function will not deliver cooling under exothermic conditions.

Unistat Operation Manual — Unistat 5, Section 1.7.2 and page 19
Maintenance
Shut off cooling water even for overnight downtime

The manual directs that the cooling water supply be shut off even for shorter downtimes such as overnight, to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the 520w over the 515w or 525w siblings?

The 520w is the water-cooled dynamic unit between the 515w and 525w in capacity. Choose it when the process heat load and -55 to 250 °C range fit its 6 kW heating and cooling curve, and when water cooling is available; the 525w offers higher capacity, the 515w lower. The 525 (non-w) is air-cooled and avoids a cooling-water circuit entirely.

What cooling-water supply and piping must be provided for the 520w?

Provide G1/2 male connections with 5–20 °C water at 1–6 bar differential pressure and 492 l/h at 15 °C/0 °C flow. Insulate pipes if below dew point. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

What are the site and duty-cycle limits for installing the 520w?

Ambient 5–40 °C, IP20, upright transport only, 2 h acclimation after temperature change, no ATEX zone. Leveling feet or parking brakes must be set before operation. The unit emits EN55011 Class A Group 1 disturbances, so EMC measures may be needed near sensitive equipment.

What maintenance and accessories are required beyond the base delivery?

The base unit ships with cable but no plug for the 400 V three-phase supply; a suitable plug must be provided on site. Open applications need an isolation sleeve accessory. Periodic checks of hoses and connections, cooling-water shutoff for overnight downtime, and refrigerant work by approved contractors are required; no consumables are included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 520w Technical Data
Datasheet ·74,0 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF