Peter Huber Kältemaschinenbau Unistat 680w Dynamisches
Peter Huber Kältemaschinenbau Unistat 680w Dynamisches — Temperature control units.
The Unistat 680w Dynamisches is a water-cooled refrigerated heating circulator with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 96 kW heating capacity and a 93 l filling volume. It is the high-capacity member of the Unistat 5xx/6xx free-standing series, distinguished by DN50 PN16 pump and DN65 PN16 cooling-water flanges. It serves closed-loop temperature control of large process loads where cooling water is available.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat 680w Dynamisches Current | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Alarm output | optic, acoustic, relay |
| Control output | POKO ONE |
| Expansion volume | 230 l |
| Filling volume | 93 l |
| Heating capacity | 96 kW |
| Pump connection | DN50 PN16 (DIN EN 1092-1) |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -60...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | DN65 PN16 |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing, water-cooled (w) refrigerated heating circulator with a water-cooled cooling machine, electronic level indicator and automatic capacity adaptation.
- Pump: magnetically coupled, stainless steel, speed-adjustable, with DN50 PN16 (DIN EN 1092-1) connections; cooling-water circuit connects via DN65 PN16.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB host/device and RS232; external Pt100 sensor input.
- Within the Unistat 5xx/6xx Dynamisches family, the 680w is the larger-frame, water-cooled variant; siblings such as the 510w, 515w, 520w and 525 Dynamisches are smaller air- or water-cooled units. The 680w is chosen when heating capacity (96 kW) and filling volume (93 l, expansion volume 230 l) exceed what the smaller frames can support.
Working range and media behavior
- Temperature range -60 to 200 °C; heating capacity 96 kW. No cooling-capacity curve points were supplied, so cooling performance must be taken from the datasheet rating conditions: 20 °C room temperature, 15 °C cooling-water inlet and 3 bar differential pressure.
- Cooling water inlet must be 5–20 °C; higher inlet temperature reduces cooling power and may require higher cooling-water flow.
- Heat-transfer fluid must respect the FL safety class (III / FL): fire point ≥ 65 °C; ethanol only under supervision. Viscosity limit 50 mm²/s.
- Safety class III / FL; refrigerant R-1270 (A3, H220), GWP 0, UN 3358.
Installation, control and operation
- Requires 400 V 3~ 50 Hz supply; three-phase units below 63 A are delivered with cable but without plug, so final connection is by the installer/electrician. Voltage tolerance ±5% with simultaneous frequency tolerance ±2%.
- Ambient temperature 5–40 °C; IP20, so dry indoor location only; no ATEX zone installation.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Pilot ONE supports Process Safety mode (compressor Permanently ON) and configurable power-failure restart behavior (Off/Standby or ON/Temperature control active).
- Alarm output: optic, acoustic and relay.
Integration & Operating Notes
Cooling water inlet temperature must be between 5 °C and 20 °C, with minimum differential pressure 3 bar and maximum pressure 6 bar. Higher inlet temperature reduces cooling power and may require increased flow. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used; insulate pipes if below dew point.
Selection & Integration Q&A
Choose the 680w when the process needs 96 kW heating, a 93 l filling volume and DN50/DN65 flanged connections. The 510w, 515w, 520w and 525 Dynamisches are smaller-frame units; select them only when their lower capacity and smaller connections match the load.
Pump connections are DN50 PN16 (DIN EN 1092-1) with bore shape Y (60°) per DIN 3863 and ball socket/sleeve nut per DIN 3863/3870; cooling water is DN65 PN16. The unit needs a 400 V 3~ 50 Hz feed with final connection by an electrician, since three-phase units below 63 A ship with cable but no plug.
Install indoors, dry, IP20, at 5–40 °C ambient, outside any ATEX zone. Activate leveling feet and caster brakes, provide a drip tray, and allow about 2 hours acclimation after moving between cold and warm environments before power-on.
Heat-transfer fluid, cooling-water hoses and mating flanges, drip tray, and the final power plug are not included. Plan periodic hose inspection/replacement, cooling-water shutoff during downtimes, and refrigerant-circuit service only by approved contractors.