Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches
Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches — Temperature control units.
The Unistat GL 570w Dynamisches is a water-cooled, CO2-refrigerant (GL) temperature control unit with magnetically coupled stainless-steel pump, covering -50 to 200 °C with 48 kW heating and 400 V 3~ 50 Hz supply. Its 196 l/min pump and Pilot ONE controller distinguish it within the Unistat 5xx series for high-flow, high-pressure-drop external circuits.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat GL 570w Dynamisches Current | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 48 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...200 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 100 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The GL suffix denotes CO2 (R-744) refrigerant; 'w' denotes water cooling, so this unit rejects heat to a site cooling-water loop rather than an air-cooled condenser. Sibling models 510w/515w/520w/525 Dynamisches are air-cooled or differ in cooling capacity; the 570w is the water-cooled high-capacity variant.
- Pump is magnetically coupled, stainless steel, with M38x1.5 male connections and adjustable speed; cooling-water ports are G3/4 male.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; controller is Pilot ONE with POKO ONE control output, Ethernet/USB/RS232 interfaces and Pt100 external sensor input.
Behavior under load and media
- Temperature range -50 to 200 °C; heating capacity 48 kW. Pump delivers max 196 l/min at 2.5 bar, falling to 86 l/min at 2.0 bar, so external circuit pressure drop must be matched to the pump curve.
- Rated cooling capacities apply only at 20 °C room, 15 °C cooling-water inlet and 1 bar differential; higher cooling-water temperatures reduce available cooling.
- Cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible water may be used. If cooling water is below the condensation point, internal cooling-water pipes must be insulated.
- Minimum filling capacity 8.9 l; expansion tank 19.9 l. Viscosity limit 100 mm²/s.
Installation, control and operation
- Requires 400 V 3~ 50 Hz supply; three-phase devices below 63 A are delivered with cable but without plug, so installer must provide plug or hard-wired connection. Voltage tolerance ±5 % and frequency ±2 % must not combine in the same direction.
- EMC Class A Group 1 per EN55011: industrial environment only; may cause interference in residential areas. IP20, ambient 5–40 °C, dry protected location.
- Cooling water: 1–4 bar differential (min 1 bar, max 6 bar), 5–20 °C intake. Shut off cooling water even for overnight downtime.
- Expansion vessel must remain unlocked for externally closed applications; locked for externally open (bath) applications. Process Safety mode requires compressor set to Permanently ON.
- Power Failure Auto can resume temperature control after outage if active; configure per unattended-operation policy.
Integration & Operating Notes
Insufficient differential pressure prevents rated cooling; excessive pressure can damage the water circuit. Rated cooling capacities are valid only at 20 °C room, 15 °C cooling-water inlet and 1 bar differential; higher cooling-water temperatures reduce capacity. Insulate internal cooling-water pipes if water is below condensation point.
The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. IP20 protection and EMC Class A Group 1 mean it must be in a dry, protected industrial location; it may cause interference in residential areas. Ambient must stay 5–40 °C.
In Process Safety mode, when the overtemperature trigger is reached the controller releases maximum cooling capacity instead of switching off all poles; this requires the compressor to be set to Permanently ON. Configure the overtemperature protection for the application and test its functionality.
Selection & Integration Q&A
Choose the 570w when you need water cooling (no air-conditioning load) and CO2 refrigerant (GL), with 48 kW heating and 196 l/min pump for high-flow circuits. Air-cooled siblings suit sites without cooling water; the 525 differs in cooling capacity and noise.
Cooling water: 1–4 bar differential (min 1 bar, max 6 bar), 5–20 °C intake, G3/4 male connections, compatible with copper/stainless/EPDM. Electrical: 400 V 3~ 50 Hz, cable without plug below 63 A, ±5 %/±2 % tolerances not combined.
Install upright only, not in ATEX zones, ambient 5–40 °C, IP20 dry location, EMC Class A industrial. Leveling feet/parking brakes engaged; allow 2 h acclimation after cold/warm move. Expansion vessel unlocked for closed, locked for open applications.
Base delivery excludes plug (below 63 A), mating hoses/adapters for M38x1.5 and G3/4, and heat transfer fluid (min 8.9 l). Periodic hose checks, cooling-water shutoff overnight, and refrigerant work by approved contractors. Drip tray required by local regulations.