Peter Huber Kältemaschinenbau Unichiller 060Tw-H
Peter Huber Kältemaschinenbau Unichiller 060Tw-H — Recirculating chillers.
The Unichiller 060Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a fixed-speed C3 pump, covering -20 to 100 °C for externally closed circuits. It is the 6 kW-class, water-cooled modification of the Unichiller series, distinguished from air-cooled siblings by its G1/2 cooling-water connection and 1–6 bar differential-pressure requirement.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 060Tw-H Current | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Heating capacity | 2 kW |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 210 kg |
| Current consumption | 13 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 380 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 580 л/час · at 15°C, flow 0°C: 730 l/h · at 15°C, flow 15°C: 870 l/h |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit (R-449A, GWP 1397, 1.6 t CO2e) with an atmospheric open expansion tank; intended only for externally closed applications.
- Fixed-speed pump type C3, not speed-adjustable, with G1 1/4 male pump connection; cooling water via G1/2 male.
- Within the series, the 060Tw-H is the higher-capacity water-cooled variant; the 030T/030Tw/040T siblings are air-cooled units (no cooling-water connection), so the 060Tw-H is chosen where a cooling-water loop exists and 6 kW-class duty is needed.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; Pilot ONE controller.
Behavior across the working range
- Cooling power falls with process temperature: 6 kW at 15 °C, 5 kW at 0 °C, 3.1 kW at -10 °C and 1.7 kW at -20 °C; the 2 kW heater covers the upper part of the -20…100 °C range.
- Cooling-water consumption depends on the process setpoint at 15 °C intake: 870 l/h at 15 °C, 730 l/h at 0 °C, 580 l/h at -10 °C and 380 l/h at -20 °C.
- Rising cooling-water temperature reduces cooling power and can increase the required flow rate, so the supply must be sized for worst-case water temperature, not nominal.
Installation, control and operation
- Requires 400 V 3~ 50 Hz, max. 13 A, 3x16 A fuse; three-phase units below 63 A are delivered with cable but without plug, so the site must provide the plug or hard-wiring.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%.
- Interfaces: Ethernet, USB host and device, RS232; external Pt100 sensor connection; power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles.
Integration & Operating Notes
Because current consumption is below 63 A, the 400 V 3~ 50 Hz unit is delivered with a cable only; the installer must provide a suitable plug or hard-wire the supply, and electrical lines directly below the water-cooled unit must be protected against liquid ingress from condensation.
Free-standing installation requires this side clearance for service access; the unit must be indoors at illuminance of at least 300 lx and up to 2000 m altitude, and must not be installed or started inside an ATEX zone because it is not explosion-protected.
The refrigerant charge (R-449A, 1.6 t CO2e) and room size must be checked at installation so the refrigerant limit per m³ of ambient air is not exceeded; this governs the minimum room volume for the installation site.
Leveling feet must be unscrewed/activated and caster parking brakes applied before the unit is put into operation, otherwise the unit can move during operation and cause crushing injuries.
The Pilot ONE allows OFF/Standby, ON or Power Failure Auto; for exothermic processes, Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles.
Selection & Integration Q&A
Pick the 060Tw-H when a cooling-water loop is available and 6 kW-class duty at 15 °C is needed; the 030T/030Tw/040T are air-cooled and have no cooling-water connection, so they suit sites without water supply but lower capacity.
A cooling-water supply at 5–20 °C intake with 1–6 bar differential pressure and G1/2 male connection, plus a 400 V 3~ 50 Hz supply with 3x16 A fuse; the unit arrives with cable but no plug, so the installer must fit a plug or hard-wire it.
Indoors only, 5–40 °C ambient, IP20, at least 300 lx illuminance and up to 2000 m altitude; keep 50–200 cm wall clearance per side, secure leveling feet and parking brakes, and protect electrical lines below the unit from condensation.
No plug is supplied for the three-phase cable, and no external Pt100 sensor or hoses/fittings are listed as included; refrigerant-circuit work must be done by approved refrigeration contractors, and cooling water must be shut off even overnight.