Peter Huber Kältemaschinenbau Unichiller GL 050w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 050w

Peter Huber Kältemaschinenbau Unichiller GL 050w — Recirculating chillers.

The Unichiller GL 050w is a water-cooled recirculating chiller with a Pilot ONE controller, a fixed-speed pressure pump (max 3.5 bar, 50 l/min) and G1 1/4 male process connections, covering -20 to 40 °C. It delivers 20 kW cooling at 5 °C and is intended for externally closed industrial process loops.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBPump connectionPump typeSafety classTemperature range, °CPower supply
Unichiller GL 050w Current —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller 030T 69G3/4 AG maleBI / NFL-10–40460V 3~ 60Hz
Unichiller 030Tw 61G3/4 AG maleBI / NFL-20–40400V 3~ 50Hz
Unichiller 040T 69G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040T-H 69G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 040Tw 70G3/4 AG maleBI / NFL-10–40400V 3~ 50Hz
Unichiller 040Tw-H 70G3/4 AG maleBIII / FL-10–100400V 3~ 50Hz
Unichiller 050-H 62G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 050w-H 62G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 055T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 055T-H 74G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 055Tw —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 060T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 060Tw-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 070T-H 66G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 070Tw-H —G1 1/4 AG maleC3III / FL-10–100400V 3~ 50Hz
Unichiller 075w —G1 1/4 AG maleE1I / NFL-20–40400V 3~ 50Hz
Unichiller 075w-H —G1 1/4 AG maleE1III / FL-20–100400V 3~ 50Hz
Unichiller 100-H 58G1 1/4 AG maleE1III / FL-20–100460V 3~ 60Hz
Unichiller 100T —G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller 100T-H —G1 1/4 AG maleC3III / FL-20–100400V 3~ 50Hz
Unichiller 100Tw 65G1 1/4 AG maleC3I / NFL-20–40400V 3~ 50Hz
Unichiller GL 035w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075 —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 075w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40460V 3~ 60Hz
Unichiller GL 100w —G1 1/4 AG malePressure pump, speed notI / NFL-20–40400V 3~ 50Hz
Unichiller P050 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w 62G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P050w-H 62G1 1/4 AG maleE2III / FL-20–100—
Unichiller P075 61G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w —G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P075w-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100 —G1 1/4 AG maleE2I / NFL-20–40460V 3~ 60Hz
Unichiller P100-H —G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz
Unichiller P100w 58G1 1/4 AG maleE2I / NFL-20–40400V 3~ 50Hz
Unichiller P100w-H 58G1 1/4 AG maleE2III / FL-20–100400V 3~ 50Hz

Full specifications

Core Performance & Functional Specifications
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 20 kW (at 5 °C)
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling water inlet must stay between 5 and 20 °C

Operating with cooling water below 5 °C or above 20 °C is outside the design range; lower inlet temperatures require insulated cooling water pipes to prevent condensation on the circuit and possible electrical hazard.

Unichiller GL 050w Technical Data Sheet, page 2
Electrical
Do not combine -5% voltage with +2% frequency tolerance

The supply must remain within ±5% voltage and ±2% frequency, but the datasheet explicitly forbids simultaneous worst-case combinations such as -5% voltage with +2% frequency; verify the site supply against both limits before connection.

Unichiller GL 050w Technical Data Sheet, page 2
Electrical
Provide UL 489 branch protection and a plug or hardwire

External branch circuit protection according to UL 489 is mandatory. Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the installer must supply a suitable plug or hardwire the connection.

Unichiller GL 050w Technical Data Sheet, page 2
Mounting
Keep 50–200 cm clearance per side for maintenance

Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance access. The unit must be installed indoors at illuminance of at least 300 lx and at altitudes up to 2000 m, and must not be placed in an ATEX zone.

Unichiller Operation Manual, page 25
Mounting
Protect cables below the unit from condensation

Because the unit is water-cooled, condensate can form; electrical lines running directly below the chiller must be protected against liquid ingress to avoid shock hazard.

Unichiller Operation Manual, page 28
Process Limits
Process circuit must be externally closed

The chiller is designed only for externally closed applications; open baths or circuits that expose the thermal fluid to atmosphere are not permitted and would invalidate the intended hydraulic design.

Unichiller Operation Manual, page 34
Control & I/O
Configure power-failure response before unattended runs

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant work only by approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-744 (CO2) under pressure (GHS H280), so handling requires qualified personnel.

Unichiller Operation Manual, page 17
Maintenance
Shut off cooling water even for overnight stops

The cooling water supply must be shut off during any downtime, including short overnight stops, to prevent water damage and corrosion in the circuit.

Unichiller Operation Manual, page 28
Mounting
Set leveling feet or parking brakes before start-up

Before the unit is put into operation, the parking brakes at the casters (if any) must be activated and/or the leveling feet unscrewed/activated to prevent the unit from moving and causing crushing injuries.

Unichiller Operation Manual, page 31

Selection & Integration Q&A

When should I choose the Unichiller GL 050w instead of the air-cooled 030T or 040T siblings?

Choose the GL 050w when you need 20 kW cooling at 5 °C and have a cooling water supply (G1/2, 1–6 bar differential). The 030T/040T are air-cooled with published noise levels and lower capacity; they suit sites without cooling water but add fan noise and reject heat to the room.

What cooling water supply and quality does the GL 050w require from adjacent plant?

Cooling water at 5–20 °C inlet, differential pressure 1–6 bar, G1/2 male connections. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the water must be compatible; lower inlet temperatures require pipe insulation.

What are the installation environment and duty-cycle limits for this chiller?

Indoor use only, ambient 5–40 °C, illuminance ≥300 lx, altitude ≤2000 m, no ATEX zone. Keep 50–200 cm clearance per side for maintenance. Shut off cooling water during any downtime, including overnight. Set leveling feet or parking brakes before start-up.

What maintenance and accessories are required beyond the base delivery?

Refrigerant circuit work only by approved refrigeration contractors. External UL 489 branch protection and a plug or hardwire are installer-supplied (cable included, plug not). No process hoses or fittings are listed as included; provide G1 1/4 male mating connections and G1/2 cooling water fittings.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 050w Technical Data Sheet
Datasheet ·30,4 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF