Peter Huber Kältemaschinenbau Unichiller 055T-H
Peter Huber Kältemaschinenbau Unichiller 055T-H — Recirculating chillers with Noise level 74 dB.
The Unichiller 055T-H is an air-cooled recirculating chiller for externally closed loops, rated -20 to 100 °C with 5.5 kW cooling at 15 °C and 2 kW heating. Its C3 pump and G1 1/4 male connections distinguish it within the 030T/040T series, serving process cooling where load varies with temperature.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 055T-H Current | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Noise level | 74 dB(A) |
| Heating capacity | 2 kW |
| Minimum fill volume | 5 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 283 kg |
| Current consumption | 12 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled condenser using natural refrigerant R-290 (A3, H220); intended only for externally closed applications.
- C3 pump with fixed speed (not adjustable) and G1 1/4 AG male pump connections; minimum fill 5 l, expansion volume 7.5 l.
- Compared with siblings Unichiller 030T (69 dB(A)) and 030Tw/030Tw-H (61 dB(A)), this 055T-H is the larger-capacity modification; noise is 74 dB(A).
- Controller: Pilot ONE; interfaces Ethernet, USB host/device, RS232; external Pt100 sensor input.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 5.5 kW at 15 °C, 2.3 kW at 0 °C, 0.8 kW at -10 °C, 0.2 kW at -20 °C (datasheet, valid at 20 °C room temperature).
- Heating capacity 2 kW; working range -20 to 100 °C.
- Capacity may drop when ambient temperature rises above 20 °C; allowable ambient 5–40 °C.
- Safety class III / FL per DIN 12876; Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles.
Installation, connection and control
- Requires 400 V 3~ 50 Hz supply, 12 A current consumption, 3x16 A fuse; three-phase devices below 63 A are supplied with cable but without plug.
- IP20 protection; indoor use only, illuminance at least 300 lx, altitude up to 2000 m; not for ATEX zones.
- Wall clearance 50–200 cm per side for maintenance; leveling feet/parking brakes must be secured before start-up.
- Power-failure response configurable via Pilot ONE (OFF/Standby, ON, Power Failure Auto).
Integration & Operating Notes
The chiller is designed exclusively for externally closed applications, so the user loop must be sealed and able to withstand the pump pressure. Open vessels or baths are not permitted and would invalidate the intended use.
The unit draws 12 A from a 400 V 3~ 50 Hz supply and requires a 3x16 A fuse. Three-phase devices below 63 A are delivered with cable but without a plug, so the installer must provide a suitable plug and verify voltage/frequency tolerances (±5% / ±2%, not all combinations allowed).
A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Before start-up, activate parking brakes and/or unscrew leveling feet to prevent movement. The unit weighs 283 kg and measures 640x937x1606 mm, so floor loading and handling equipment must be planned.
The unit must not be installed in an ATEX zone because it is not explosion-protected. It is for indoor use at illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must stay within 5–40 °C; above 20 °C cooling capacity may drop. IP20 requires protection from water and dust.
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant. At installation, check the refrigerant charge on the rating plate against room size to avoid exceeding the limit per m³ of ambient air. Work on the refrigerant circuit must be performed only by approved refrigeration/air-conditioning contractors.
The Pilot ONE allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles; choose the mode that matches the risk of the process.
Selection & Integration Q&A
Choose the 055T-H when you need higher cooling capacity (5.5 kW at 15 °C) and heating up to 2 kW over -20 to 100 °C. The 030T/030Tw siblings are smaller-capacity units; the 030Tw variants are quieter (61 dB(A)) but not a substitute for the 055T-H's capacity.
A 400 V 3~ 50 Hz supply with a 3x16 A fuse; current consumption is 12 A. The unit comes with cable but no plug (three-phase below 63 A), so provide a suitable plug. Verify voltage/frequency tolerances of ±5% / ±2%, noting not all combinations are allowed.
Install indoors only, at 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, and not in an ATEX zone. Keep 50–200 cm clearance per side for maintenance, secure leveling feet/parking brakes, and protect from water/dust (IP20). Cooling capacity may drop above 20 °C ambient.
Refrigerant circuit work must be done by approved contractors; check R-290 charge against room volume. The base delivery excludes a power plug and process hoses/fittings for the G1 1/4 male connections; minimum fill is 5 l with 7.5 l expansion volume.