Peter Huber Kältemaschinenbau Unichiller 060T
Peter Huber Kältemaschinenbau Unichiller 060T — Recirculating chillers.
The Unichiller 060T is an air-cooled, CFC/HCFC-free recirculating chiller with a C3 circulation pump (80 l/min, 5.6 bar max) and G1 1/4 male connections, rated for -20 to 40 °C. Within the Unichiller series it is the 6 kW-class modification (6 kW at 15 °C, falling to 1.4 kW at -20 °C), intended for externally closed process loops that reject heat to ambient air.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 060T Current | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Minimum fill volume | 5 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 228 kg |
| Current consumption | 10 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280; GWP 1397; CO2 equivalent 3.1 t), CFC- and HCFC-free.
- Fixed-speed C3 pump, not speed-adjustable; pump connection G1 1/4 AG male.
- Controller Pilot ONE with Ethernet, USB (host and device) and RS232; external Pt100 sensor input.
- Safety class I / NFL to DIN 12876; protection IP20; weight 228 kg.
- Siblings in the series (030T, 030Tw, 030Tw-H, 040T) are smaller-capacity units; the 060T is the larger 6 kW-class machine. Models with “P” in the name are the ones intended for high-pressure-drop loops — the 060T carries no P suffix.
Behavior under load and temperature
- Cooling capacity is strongly temperature-dependent: 6 kW at 15 °C, 5 kW at 0 °C, 2.8 kW at -10 °C, 1.4 kW at -20 °C. Capacity data are valid at 20 °C room temperature and drop as ambient rises.
- Permissible operating range -20 to 40 °C; ambient range 5 to 40 °C.
- Pump curve: max delivery 80 l/min at max delivery pressure 5.6 bar; flow falls as loop pressure rises, so the hydraulic design must be checked against the actual pressure drop.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles — relevant for exothermic processes.
Installation, connection and operation
- 400 V 3~ 50 Hz supply, max current 10 A, fusing 3x16 A; voltage tolerance ±5% and frequency ±2%, not to be combined in the unfavorable direction.
- Three-phase devices below 63 A are delivered with cable but without a plug; a suitable plug must be provided on site.
- Air-cooled: sufficient ambient air and space for heat rejection required; indoor use only, illuminance at least 300 lx, site altitude up to 2000 m.
- Wall clearance 50–200 cm per side for maintenance on free-standing models; leveling feet/parking brakes must be set before start-up.
- Power-failure response configurable via Pilot ONE: OFF/Standby, ON, or Power Failure Auto.
- Not explosion-protected: must not be installed or started in an ATEX zone. Intended only for externally closed applications.
Integration & Operating Notes
The unit is designed for externally closed applications, so the hydraulic circuit must be a sealed loop. Open tanks or unsealed vessels change the pressure regime and are outside the intended use.
Cooling capacity falls from 6 kW at 15 °C to 1.4 kW at -20 °C and is quoted at 20 °C room temperature. Sizing on the nominal figure will undersize the machine at low set points or in warm rooms.
Devices below 63 A are delivered with cable but no plug, so the site must provide a matching 400 V 3~ connector. Verify voltage ±5% and frequency ±2% are not simultaneously at the unfavorable limit.
Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access. Air cooling also needs unobstructed ambient air for heat rejection, and the unit is IP20 so it must sit in a dry, protected indoor location.
The unit is not explosion-protected and must not be installed or started inside an ATEX zone. It is for indoor use only, at illuminance of at least 300 lx and site altitude up to 2000 m.
Maximum tightening torques for the G1 1/4 male pump connections must not be exceeded, and the connection must be counterheld during hose fitting to avoid damaging the pump port or creating leaks.
The Pilot ONE allows OFF/Standby, ON, or Power Failure Auto after mains loss. The chosen behavior determines whether the process restarts automatically, which matters for unattended or exothermic processes.
In Process Safety mode the controller releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles, preventing runaway temperatures in exothermic processes.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a CO2 equivalent of 3.1 t, which triggers F-gas documentation and leak-check obligations.
Selection & Integration Q&A
Pick the 060T when the loop needs the 6 kW class (6 kW at 15 °C, 5 kW at 0 °C, 2.8 kW at -10 °C, 1.4 kW at -20 °C). The 030T/040T are lower-capacity units; if the duty point is near their limit at low temperature, the 060T gives headroom.
A sealed externally closed loop with G1 1/4 male mating fittings, designed against the C3 pump curve (80 l/min max, 5.6 bar max, flow falling with pressure). Power: 400 V 3~ 50 Hz, 10 A max, 3x16 A fusing; cable supplied without plug.
Indoor only, 5–40 °C ambient, ≥300 lx lighting, up to 2000 m altitude, IP20 dry location, no ATEX zone. Allow 50–200 cm clearance per side for service and free air for the air-cooled condenser; set leveling feet/parking brakes before start-up.
The three-phase cable comes without a plug, so the site must supply one. Refrigerant-circuit work is restricted to approved contractors; the R-449A charge (CO2 equivalent 3.1 t) carries F-gas leak-check and documentation duties.