Peter Huber Kältemaschinenbau Unichiller 017T
Peter Huber Kältemaschinenbau Unichiller 017T — Recirculating chillers with Noise level 64 dB.
The Unichiller 017T is an air-cooled, externally closed recirculating chiller rated -10 to 40 °C with a Pilot ONE controller and a fixed-speed pump delivering up to 25 l/min at 3 bar. It is the 1.7 kW-class modification in the Unichiller series, distinguished from the 015 siblings by its higher cooling capacity and 64 dB(A) noise level. It serves closed process loops requiring stable cooling down to -10 °C.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 017T Current | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 64 dB(A) |
| Expansion volume | 2 l |
| Maximum delivery flow rate | 25 l/min (pressure pump max; also 3 bar) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...40 °C |
| Power supply | 230V 1~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280, GWP 1397); IP20 enclosure; 230 V 1~ 50 Hz supply.
- Fixed-speed pump type B, not speed-adjustable, with G3/4 AG male connections; 2 l expansion volume.
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; Pt100 external sensor input.
- Within the Unichiller series, the 017T sits above the 015/015-H/015 OLÉ siblings (63 dB(A)); it is the higher-capacity, 64 dB(A) variant. The 'T' suffix does not denote a high-pressure-drop 'P' model, so it is not the choice for high-pressure-drop loops.
Behavior under load and temperature
- Cooling capacity is strongly temperature-dependent: 1.7 kW at 15 °C, 0.9 kW at 0 °C and 0.4 kW at -10 °C, all rated at 20 °C room temperature; capacity falls further as ambient rises.
- Pump curve: maximum 25 l/min and 3 bar; flow falls to 5 l/min at 3 bar, so the external circuit pressure drop must be matched to the pump curve.
- Intended only for externally closed applications with an atmospheric open expansion tank; not for open-loop circuits.
Installation, control and operation
- Requires 230 V single-phase 50 Hz; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not permitted.
- EMC class A group 1 to EN55011: intended for industrial environments; may cause radio interference in residential areas.
- Power-failure response configurable via Pilot ONE as OFF/Standby, ON or Power Failure Auto; Process Safety mode releases maximum cooling capacity at the OT trigger instead of switching off all poles.
- Air-cooled housing needs clearance for airflow; wall clearance of 50–200 cm per side required for maintenance on free-standing models. Leveling feet/parking brakes must be set before start-up.
Integration & Operating Notes
The unit is air-cooled and rejects heat to the room; insufficient clearance reduces cooling capacity and can cause overheating. Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access, and the stated space requirements around the housing must be respected.
The supply must be 230 V 1~ 50 Hz. Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed. Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
The chiller is designed only for externally closed applications with an atmospheric open expansion tank. The pump delivers 25 l/min max and 3 bar max, with flow falling to 5 l/min at 3 bar; the external loop pressure drop must be checked against this curve. Models with 'P' in the model name are the ones suited to high-pressure-drop applications, not the 017T.
Cooling capacity is 1.7 kW at 15 °C, 0.9 kW at 0 °C and 0.4 kW at -10 °C, rated at 20 °C room temperature. Capacity drops as ambient temperature rises, so the required heat load must be checked at the lowest set point and at the actual site ambient, not at the maximum rating.
The Pilot ONE power-failure response can be set to OFF/Standby, ON or Power Failure Auto, determining whether the process restarts automatically after an outage. Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles, which matters for exothermic processes where cutting cooling would cause runaway temperatures.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280, GWP 1397), so refrigerant handling and leak reporting obligations apply; the charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
The unit is IP20 and must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude, with ambient temperature between 5 and 40 °C. It must not be installed or started up inside an ATEX zone because it is not explosion-protected. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Pick the 017T when the heat load requires the higher 1.7 kW at 15 °C capacity and 64 dB(A) is acceptable; the 015 siblings are 63 dB(A) but lower capacity. If the loop has high pressure drop, a 'P' model is required instead, since the 017T is not a high-pressure-drop variant.
The circuit must be externally closed with an atmospheric open expansion tank. Pump connections are G3/4 AG male; maximum tightening torques must not be exceeded and connections must be counterheld. Match loop pressure drop to the pump curve: 25 l/min max, 3 bar max, falling to 5 l/min at 3 bar.
Install indoors only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, never in an ATEX zone. Air-cooled housing needs 50–200 cm side clearance for maintenance. Leveling feet/parking brakes must be set before start-up, and cooling water supply (if applicable) shut off even overnight.
Refrigerant circuit work must be done by approved refrigeration/air-conditioning contractors; R-449A (GWP 1397) handling and leak reporting apply. A country-specific plug must be specified at ordering. The external Pt100 sensor and process hoses are not part of the base unit and must be provided for the loop.