Peter Huber Kältemaschinenbau Unichiller P022 OLÉ
Peter Huber Kältemaschinenbau Unichiller P022 OLÉ — Recirculating chillers with Noise level 65 dB.
The Unichiller P022 OLÉ is a recirculating chiller with a 'P' designation for high-pressure-drop loops, rated -10 to 40 °C and 2.2 kW at 15 °C. Its OLÉ controller and 21 l/min pump at 0.5 bar distinguish it from 015 siblings. It serves externally closed cooling circuits in dry indoor sites.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P022 OLÉ Current | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 65 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| Temperature range | -10...40 °C |
| Weight | 81 kg |
| Current consumption | 9 A |
| Power supply | 220-240V 1~ 50Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The P022 OLÉ is the high-pressure-drop variant of the Unichiller series: the 'P' in the model name marks it as particularly suitable for applications requiring high pressure drops (Operation Manual, p. 34).
- It is an externally closed-circuit chiller only; open baths or vented loops are outside its intended use (Operation Manual, p. 34).
- Within the series, the 015 siblings are the lower-capacity neighbors; the P022 OLÉ carries the OLÉ controller and a higher noise figure of 65 dB(A) versus 63 dB(A) for the 015 models.
- The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the charge and room volume must be checked at installation (Datasheet, p. 1; Operation Manual, p. 15).
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 2.2 kW at 15 °C, 1.6 kW at 0 °C and 1 kW at -10 °C, so selection must be made at the actual process temperature, not at nominal (Datasheet, p. 1).
- The pump curve falls with backpressure: 21 l/min at 0.5 bar, 17 l/min at 1.0 bar, 11 l/min at 1.5 bar and 6 l/min at 2.0 bar; the process loop pressure drop must be matched to this curve (Datasheet, p. 1).
- The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified (Datasheet, p. 2).
- For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure (Datasheet, p. 2).
- Cooling capacity data are valid at 20 °C room temperature and may drop as ambient temperature rises (Datasheet, p. 2).
Installation, control and operation
- Supply is 220-240 V single-phase 50 Hz with a minimum 10 A fuse; voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed (Datasheet, p. 1-2).
- The pump connection is G3/4 male; maximum tightening torques must not be exceeded and connections must be counterheld during hose installation (Datasheet, p. 1; Operation Manual, p. 28).
- The OLÉ controller provides an OLED display with 0.1 K resolution, USB (Device) and RS232 interfaces (Datasheet, p. 1).
- The power-failure response is configurable as OFF/Standby, ON or Power Failure Auto via the Pilot ONE (Operation Manual, p. 19).
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles (Operation Manual, p. 18).
- The unit is IP20 and must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude (Datasheet, p. 1; Operation Manual, p. 25).
- A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models (Operation Manual, p. 25).
- Leveling feet must be unscrewed/activated and parking brakes applied before switching on (Operation Manual, p. 31).
- The unit must not be installed or started up inside an ATEX zone because it is not explosion-protected (Operation Manual, p. 13).
Integration & Operating Notes
The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar. A high-pressure-drop loop will see strongly reduced flow, so calculate the circuit pressure drop and verify the resulting flow against the curve before commissioning.
The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). Using these fluids risks pump failure; specify an alternative pump or choose a different heat transfer fluid.
The datasheet permits ±5% voltage tolerance only with a simultaneous ±2% frequency tolerance, and certain combinations are not allowed. Verify the site supply quality before connection; a generator or weak grid may exceed these limits.
Single-phase devices are supplied with a power cable and country-specific plug that must be stated when ordering. Confirm the plug type for the installation country to avoid a missing or wrong connector on delivery.
Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access. Plan the installation footprint accordingly; a tight enclosure will block service access to the refrigerant circuit and pump.
The unit is not explosion-protected and must not be installed or started up within an ATEX zone. Additionally, the R-290 refrigerant charge and room volume must be checked to avoid exceeding the refrigerant limit per m³ of ambient air.
The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must stay within 5 to 40 °C; capacity data are valid at 20 °C room temperature and may drop if the room is warmer.
The power-failure response can be set to OFF/Standby, ON or Power Failure Auto via the Pilot ONE. For exothermic processes, choose the mode that restarts cooling automatically; for safe processes, OFF/Standby prevents unexpected restart.
In Process Safety mode the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is relevant for exothermic reactions where cutting cooling would cause runaway temperatures.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-290 (A3, H220) flammable refrigerant charge also requires checking the room volume at installation.
The cooling water supply must be shut off even for shorter downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Choose the P022 OLÉ for high-pressure-drop loops (the 'P' designation) and when the 2.2 kW at 15 °C capacity is needed. The 015 siblings are lower-capacity neighbors at 63 dB(A); the P022 OLÉ is rated 65 dB(A). Select the 015 when its capacity covers the load and high pressure drop is not a constraint.
The pump connection is G3/4 male, so mating hoses and couplings must match this thread and be counterheld at the maximum tightening torque. The supply is 220-240 V single-phase 50 Hz with a minimum 10 A fuse; voltage tolerance is ±5% only with ±2% frequency tolerance. The loop must be externally closed and its pressure drop matched to the pump curve.
Install indoors only, at 5 to 40 °C ambient, at least 300 lx illuminance and up to 2000 m altitude, with 50 to 200 cm clearance per side for maintenance. The unit is IP20 and not permitted in an ATEX zone. Activate leveling feet and parking brakes before operation; check the R-290 charge against room volume.
Refrigerant circuit work must be done by approved refrigeration/air-conditioning contractors. The heat transfer fluid must be prepared with glysantine or ethylene glycol for freeze protection; acetone and polyglycols must be avoided with the plastic pump. The power cable and country-specific plug must be specified at ordering; the manual does not list a spare-parts kit in base delivery.