Peter Huber Kältemaschinenbau Unichiller P012-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P012-H

Peter Huber Kältemaschinenbau Unichiller P012-H — Recirculating chillers with Noise level 59 dB.

The Unichiller P012-H is an air-cooled recirculating chiller with a 'P' designation for high-pressure-drop external circuits, rated -20 to 100 °C and delivering up to 25 l/min at 2.5 bar through a non-adjustable plastic pump. Its cooling capacity falls from 1.2 kW at +15 °C to 0.25 kW at -20 °C, so it suits closed-loop process cooling where the coldest duty point is checked against the load.

Noise level
59 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P012-H Current 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 59 dB(A)
Maximum delivery flow rate 25 l/min (pressure pump max; also 2, 5 bar)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 66 kg
Electrical, Power & Drive Specifications
Current consumption 15 A (max)
Fuse rating 16 A
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour

Control, installation and operation

Integration & Operating Notes

Electrical
Supply tolerance is only ±5 % voltage with ±2 % frequency

The 220–240 V 1~/2~ 50/60 Hz feed must stay within ±5 % voltage and ±2 % frequency simultaneously; certain combinations are explicitly not allowed. Verify site power quality before commissioning, and provide a 16 A fuse for the 15 A maximum current draw.

Unichiller P012-H Datasheet, Technical Data Table
Process Limits
Plastic pump is not resistant to acetone or polyglycols

The standard plastic pump must not be used with acetone or polyglycols (depending on manufacturer); a more resistant plastic must be ordered if these media are required. For sub-zero duty, mix water with glysantine or ethylene glycol for freeze protection.

Unichiller P012-H Datasheet, Technical Data Table
Mounting
Keep 50–200 cm clearance per side for maintenance access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side. The unit is air-cooled, so restricted airflow will reduce capacity; ambient must remain between 5 and 40 °C and the unit may only be used indoors at ≥300 lx and up to 2000 m altitude.

Unichiller Operation Manual, Section: Commissioning
Process Limits
Externally closed circuits only; 'P' suits high pressure drop

The unit is designed for externally closed applications. The 'P' model designation indicates particular suitability for applications requiring high pressure drops, so check the external circuit pressure drop against the pump curve (25 l/min max, 2.5 bar max) to confirm adequate flow.

Unichiller Operation Manual, Section: Function Description
Safety
Process Safety mode releases max cooling at OT trigger

In Process Safety mode the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. Select this mode for exothermic processes where cutting cooling would cause runaway temperatures; otherwise the standard OT cut-out applies.

Unichiller Operation Manual, Section: Setup Mode
Maintenance
Refrigerant circuit work only by approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge of 0.45 kg (0.6 t CO2e) must be checked against room size at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Section: Maintenance
Mounting
Do not install or start up inside an ATEX zone

The unit is not explosion-protected and must not be installed or started up within an ATEX zone. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation; shut off cooling water even for short downtimes such as overnight.

Unichiller Operation Manual, Section: Commissioning

Selection & Integration Q&A

When should I choose the P012-H over the Unichiller 015 siblings?

Pick the P012-H when the external circuit has high pressure drop (the 'P' designation) and 59 dB(A) is preferred over the 63 dB(A) of the 015 variants. The 015 models are the alternative when their capacity and noise level suit the duty; verify the coldest-point capacity (0.25 kW at -20 °C) against your load either way.

What supply and protection must I provide for the P012-H?

A 220–240 V 1~/2~ 50/60 Hz feed within ±5 % voltage and ±2 % frequency simultaneously, protected by a 16 A fuse; maximum current draw is 15 A. Single/two-phase units ship with a power cable and country-specific plug that must be specified at ordering.

What are the mounting and duty-cycle limits for this chiller?

Indoor use only, 5–40 °C ambient, ≥300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. Leveling feet must be activated and parking brakes applied before start-up. Not permitted in ATEX zones. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises.

What maintenance and accessories are required beyond the base unit?

Refrigerant circuit work must be done by approved refrigeration contractors; the R-449A charge (0.45 kg, 0.6 t CO2e) must be checked against room volume. Hoses and couplings for the G3/4 male pump connections are not included, and the heat transfer fluid (min. 3.8 l fill, 1.7 l expansion) must be specified separately, with glycol for sub-zero duty.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P012-H Datasheet
Datasheet ·155,7 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF