Peter Huber Kältemaschinenbau Unichiller 025Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 025Tw-H

Peter Huber Kältemaschinenbau Unichiller 025Tw-H — Recirculating chillers.

The Unichiller 025Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a fixed-speed pump, covering -10 to 100 °C. It is the 025 water-cooled modification of the Unichiller series, distinguished from the 015 siblings by its larger capacity class. It serves externally closed process loops that need both cooling and active heating.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 025Tw-H Current —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Expansion volume 4 l
Heating capacity 2 kW
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -10...100 °C
Electrical, Power & Drive Specifications
Current consumption 15 A
Fuse rating 16 A
Power supply 230V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 120 l/h (at 15°C, flow 15°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Installation, connection and control

Integration & Operating Notes

Process Limits
Externally closed loops only; not for open baths

The manual states the unit is designed for externally closed applications. Hydraulic circuit design must therefore avoid open tanks or any path that exposes the fluid to atmosphere, since the unit is not configured for that duty.

Unichiller Operation Manual, Function Description / Selection
Cooling & Media
Cooling water must stay within 5–20 °C inlet

The datasheet specifies cooling water intake between 5 and 20 °C. Facility water outside this window invalidates the rated cooling capacity and may require a mixing or booster loop; below the condensation point the internal cooling water pipes must be insulated.

Unichiller 025Tw-H Technical Data Sheet, Specifications
Cooling & Media
Shut off cooling water even for overnight stops

The manual requires shutting off the cooling water supply even for short downtimes such as overnight. This prevents water damage and corrosion during idle periods and should be part of the site's shutdown procedure.

Unichiller Operation Manual, Installation
Mounting
Keep 50–200 cm clearance per side for service

Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access. The installation footprint must be planned around this, not just the unit envelope.

Unichiller Operation Manual, Installation
Mounting
Indoor, dry, ≥300 lx, below 2000 m altitude

The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude. With IP20 protection it must be in a dry location protected from water ingress; electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, Installation
Electrical
Voltage and frequency tolerances are coupled

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed. Supply quality must be checked against this combined envelope, not each parameter independently.

Unichiller 025Tw-H Technical Data Sheet, Specifications
Electrical
Specify country-specific plug at ordering

Single/two-phase devices (100 V to 240 V) are supplied with a power cable and country-specific plug that must be specified when ordering. Three-phase devices below 63 A come with cable but without plug, leaving final connection to the site.

Unichiller 025Tw-H Technical Data Sheet, Specifications
Control & I/O
Configure power-failure response before commissioning

The Pilot ONE allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles; the chosen logic must match the process risk assessment.

Unichiller Operation Manual, Normal Operation / Safety
Maintenance
Refrigerant circuit work only by approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains 0.5 kg R-449A (GWP 1397, CO2 equivalent 0.7 t), so refrigerant handling and ventilation requirements apply and leak-reporting thresholds should be checked.

Unichiller Operation Manual, Maintenance / Technical Data Sheet
Mounting
Do not install in an ATEX zone

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Room size and refrigerant charge must also be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Foreword / Installation

Selection & Integration Q&A

When should I choose the 025Tw-H instead of the 015 siblings?

Pick the 025Tw-H when the process needs the larger 025 capacity class and water cooling; the 015 siblings are air-cooled and rated at 63 dB(A). If facility cooling water is unavailable, a 015 air-cooled variant is the alternative.

What does the facility need to provide for the cooling water and electrical connection?

Cooling water at 5–20 °C inlet, 120 l/h at 15 °C, minimum 1 bar differential and maximum 6 bar, connected via G1/2 male. Electrical: 230 V 1~ 50 Hz, 15 A, 16 A fuse, with a country-specific plug specified at ordering.

What are the site and duty-cycle limits from the manual?

Indoors only, 5–40 °C ambient, ≥300 lx, below 2000 m, IP20 dry location, 50–200 cm side clearance for service, no ATEX zone. Shut off cooling water even overnight; activate leveling feet and parking brakes before start-up.

What maintenance and accessories are not included in the base delivery?

Refrigerant circuit service must be done by approved contractors; the unit holds 0.5 kg R-449A. The power cable and country-specific plug must be specified at ordering, and cooling water insulation is a site responsibility when water is below the condensation point.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 025Tw-H Technical Data Sheet
Datasheet ·63,7 КБ ·2 pp.
View PDF