Peter Huber Kältemaschinenbau Unichiller P007 OLÉ
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P007 OLÉ

Peter Huber Kältemaschinenbau Unichiller P007 OLÉ — Recirculating chillers with Noise level 58 dB.

The Unichiller P007 OLÉ is a compact recirculating chiller with a plastic pump (type B, non-adjustable) and G3/4 AG male connections, rated for -20 to 40 °C. Its 'P' designation marks it for high-pressure-drop loops, where the pump still delivers 6 l/min at 2.0 bar. It serves externally closed cooling circuits requiring 0.2–0.7 kW of heat removal.

Noise level
58 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P007 OLÉ Current 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 58 dB(A)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 59 kg
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Thermal and hydraulic behavior

Installation, control and operation

Integration & Operating Notes

Process Limits
Loop pressure drop must be checked against the pump curve

Flow collapses from 21 l/min at 0.5 bar to 6 l/min at 2.0 bar. Although the 'P' designation targets high-pressure-drop applications, the actual circuit resistance must be calculated and matched to the curve, otherwise the required cooling capacity at the set point cannot be transferred.

Unichiller P007 OLÉ Technical Data, Specifications
Electrical
Do not combine -5% voltage with +2% frequency

Voltage and frequency tolerances must be applied simultaneously, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against both limits before commissioning, and size upstream protection between 10 A and 16 A.

Unichiller P007 OLÉ Technical Data, page 2
Mounting
Keep 50–200 cm wall clearance per side for maintenance

Free-standing installation requires a wall clearance of 50 to 200 cm on either side. The unit is IP20 and must be used indoors only, at illuminance of at least 300 lx and up to 2000 m altitude, within an ambient window of 5 to 40 °C.

Unichiller Operation Manual, Installation
Mounting
Not permitted inside an ATEX zone

The unit is not explosion-protected and must not be installed or started up within an ATEX zone. Refrigerant charge (0.3 kg R-449A) and room size must also be checked at installation to stay within the permitted refrigerant quantity per m³ of ambient air.

Unichiller Operation Manual, page 13 and page 15
Process Limits
Plastic pump excludes acetone and polyglycols

The type B plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these fluids must not be used as heat transfer media. For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.

Unichiller P007 OLÉ Technical Data, page 2
Control & I/O
Configure power-failure response before unattended runs

The power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE controller. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, Normal Operation and Safety
Maintenance
Refrigerant circuit work restricted to approved contractors

Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit is classified to DIN 12876 with NFL/FL identification depending on whether the thermal fluid is combustible; the rating is stated on the data sheet.

Unichiller Operation Manual, page 17
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the P007 OLÉ instead of a Unichiller 015 variant?

Pick the P007 OLÉ for lower heat loads (0.2–0.7 kW depending on set point) and high-pressure-drop loops, where its 'P' designation applies. The 015/015-H siblings are louder (63 vs 58 dB(A)) and belong to a different capacity class; select them when the load exceeds the P007 envelope.

What must I verify in the hydraulic circuit and power supply before connecting this chiller?

Check loop pressure drop against the pump curve (21 l/min at 0.5 bar down to 6 l/min at 2.0 bar) and provide at least 3.8 l filling volume with 1.7 l expansion allowance. Supply must be 208–240 V, 1~ or 2~, 50/60 Hz, 6.5 A max, fused 10–16 A; specify the country-specific plug at ordering.

What site and duty-cycle limits does the manual impose on installation?

Indoors only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. Not permitted in ATEX zones. Leveling feet or parking brakes must be engaged before start-up, and cooling water shut off even overnight.

What maintenance and consumables are not included in the base delivery?

The base unit ships with a power cable and country-specific plug only if specified at ordering. Heat transfer fluid (water with glysantine or ethylene glycol for freeze protection) and the external loop are not included. Refrigerant circuit service requires approved contractors; no routine interval is stated in the supplied digests.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P007 OLÉ Technical Data
Datasheet ·111,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF