Peter Huber Kältemaschinenbau Unichiller 015-H OLÉ
Peter Huber Kältemaschinenbau Unichiller 015-H OLÉ — Recirculating chillers with Noise level 63 dB.
The Unichiller 015-H OLÉ is a single-phase, air-cooled recirculating chiller with an immersion pump and a rated operating window of -20 to 100 °C. Its 1.5 kW cooling capacity at 15 °C derates to 0.3 kW at -20 °C, so low-temperature duty must be sized against the load. It serves externally closed process loops where a compact, fixed-speed hydraulic circuit is sufficient.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 015-H OLÉ Current | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 63 dB(A) |
| Maximum delivery flow rate | 29 l/min (1 bar) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | Immersion pump |
| Temperature range | -20...100 °C |
| Weight | 61 kg |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~ 50Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series position
- The 015-H OLÉ is the OLÉ-controller variant of the 015-H frame; the sibling 015-H carries the same 63 dB(A) noise figure, while the 015w is the quieter 54 dB(A) member of the family.
- Hydraulics are built around a fixed-speed immersion pump (not speed-adjustable) with a G3/4 AG male pump connection; maximum delivery is 29 l/min at 1 bar.
- The unit is rated IP20 and weighs 61 kg, so it is a floor-standing indoor machine rather than a panel-mount component.
Thermal behavior under load
- Cooling capacity is 1.5 kW at 15 °C, 1 kW at 0 °C, 0.7 kW at -10 °C and 0.3 kW at -20 °C; the curve is steep, so a loop that is comfortable at mid-range temperatures may be marginal near the cold limit.
- Published capacity is valid at 20 °C room temperature; the datasheet warns that capacity may drop as ambient temperature rises toward the 40 °C maximum.
- The refrigerant is R-449A (A1, H280), GWP 1397, 0.35 kg charge, 0.5 t CO2 equivalent.
Control, connection and operation
- The OLÉ controller provides an OLED display with 0.1 K resolution, USB (Device) and RS232 interfaces, and a configurable power-failure response (OFF/Standby, ON, or Power Failure Auto).
- Electrical supply is 220–240 V 1~ 50 Hz with 13.5 A maximum current and a 16 A fuse; voltage tolerance ±5% and frequency tolerance ±2% apply but must not be combined in opposite directions.
- The manual restricts use to externally closed applications and forbids installation in an ATEX zone; models with “P” in the name are the ones intended for high pressure-drop loops, which this model is not.
Integration & Operating Notes
Free-standing installation requires this side clearance so maintenance can be performed; the datasheet also directs the installer to respect the space requirements for airflow, since air-cooled units lose capacity if the condenser is starved.
The unit must be immobilized before it is switched on: activate the caster parking brakes and/or unscrew the leveling feet. An unsecured 61 kg chiller can move during operation and cause crushing injury.
Single-phase devices in the 100–240 V class are delivered with a power cable and a country-specific plug; the plug type is not universal and must be stated when ordering to avoid rework at commissioning.
The manual states the unit is designed for externally closed applications. Open baths or circuits that expose the fluid to atmosphere fall outside the intended use and should not be connected to this chiller.
The standard plastic pump is not resistant to acetone or polyglycols depending on manufacturer. For sub-zero duty, mix water with glysantine or ethylene glycol for freeze protection rather than running pure water.
When attaching hoses to the G3/4 AG male pump connection, the maximum tightening torque from the manual table must not be exceeded and the connection must be counterheld to avoid transmitting torque into the pump housing.
The OLÉ controller can be configured for OFF/Standby, ON, or Power Failure Auto after mains loss. The datasheet also lists an autostart function, so the chosen mode determines whether the process restarts automatically or stays off after recovery.
An optional Pt100 sensor connection (#10519) can display, but not control, process temperature. It is available only factory fitted at additional charge and cannot be retrofitted in the field, so it must be decided at ordering.
The unit is IP20 and may only be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. It must not be installed or started inside an ATEX zone because it is not explosion-protected.
With 0.35 kg of R-449A (GWP 1397, 0.5 t CO2 equivalent), the refrigerant limit per m³ of ambient air must be checked against the room size at installation to avoid exceeding the permissible concentration.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Plan service access accordingly, since the 50–200 cm side clearance is what makes this work possible.
Selection & Integration Q&A
Pick the 015-H OLÉ when you need the OLÉ controller with OLED display, USB and RS232 and can accept 63 dB(A). The 015w is the quieter 54 dB(A) sibling for noise-sensitive rooms; the 015-H is the same frame without the OLÉ controller.
The country-specific plug must be specified at order, and the Pt100 process sensor connection (#10519) is a factory-fit option only, not retrofittable. Hoses and mating fittings for the G3/4 AG male pump connection are not part of the chiller.