Peter Huber Kältemaschinenbau Unichiller P015
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P015

Peter Huber Kältemaschinenbau Unichiller P015 — Recirculating chillers with Noise level 63 dB.

The Unichiller P015 is a recirculating chiller with a 'P' designation for high-pressure-drop applications, operating from -20 to +40 °C using R-290 refrigerant. It delivers 1.5 kW at +15 °C but only 0.3 kW at -20 °C, distinguishing it from standard 015 siblings. It serves externally closed industrial cooling loops requiring elevated pump pressure.

Noise level
63 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P015 Current 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 63 dB(A)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 61 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 220-240V 1~/2~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and Series Positioning

The Unichiller P015 is a free-standing recirculating chiller in the Unichiller series (up to 25 kW class). Its 'P' suffix explicitly denotes suitability for applications requiring high pressure drops, per the operation manual. Unlike the standard Unichiller 015, 015-H, or OLÉ variants, the P015 is configured with a pump delivering up to 2.5 bar and 25 l/min, making it the choice for restrictive circuits. The unit uses a plastic pump (type B, non-adjustable speed) with G3/4 male connections and a minimum filling capacity of 3.8 l. Refrigerant is R-290 (A3, H220) with a charge of only 0.16 kg (GWP 0.02).

Thermal Performance and Load Behavior

Cooling capacity is strongly temperature-dependent: 1.5 kW at +15 °C, 1.0 kW at 0 °C, 0.7 kW at -10 °C, and 0.3 kW at -20 °C, all rated at 20 °C room temperature. Capacity may fall if ambient temperature rises above 20 °C. The temperature range is -20 to +40 °C. For freeze protection below 0 °C, water must be mixed with glysantine or ethylene glycol. The plastic pump is not resistant to acetone or polyglycols, so these fluids require a different pump option.

Control, Installation, and Operation

The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 external sensor connection. Electrical supply is 220–240 V, 1~/2~, 50 Hz, with max current 8.5 A and a 16 A fuse. Voltage and frequency tolerances are ±5% and ±2% but must not be combined unfavorably. Single/two-phase units ship with a country-specific plug (must be specified at order). The unit is IP20, EMC Class A Group 1 (industrial environments only), and must be installed indoors at 5–40 °C ambient, at least 300 lx illuminance, and below 2000 m altitude. A wall clearance of 50–200 cm per side is required for maintenance. Leveling feet or parking brakes must be engaged before operation. The power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles.

Integration & Operating Notes

Process Limits
Cooling capacity collapses at low temperatures

Sizing on nominal capacity will undersize the unit at low temperatures: capacity drops from 1.5 kW at +15 °C to 0.3 kW at -20 °C. The required load must be checked at the actual operating temperature, and ambient above 20 °C further reduces available cooling.

Datasheet, Specifications
Electrical
Supply tolerances must not combine unfavorably

Voltage tolerance is ±5% and frequency tolerance ±2%, but the manual explicitly forbids combining -5% voltage with +2% frequency. Verify supply quality before commissioning to avoid compressor and pump damage and warranty voidance.

Datasheet, Specifications
Mounting
A3 refrigerant requires room-size assessment

R-290 is flammable (A3). The refrigerant charge (0.16 kg) and room volume must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air. The unit must not be installed in an ATEX zone.

Operation Manual, Installation; Datasheet, Specifications
Mounting
Indoor location with defined clearance and lighting

Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, and below 2000 m altitude. A wall clearance of 50–200 cm per side is required for maintenance access. Leveling feet or parking brakes must be engaged before start-up.

Operation Manual, Installation
Process Media
Plastic pump limits fluid compatibility

The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection below 0 °C, mix water with glysantine or ethylene glycol. Minimum filling capacity is 3.8 l; pump connections are G3/4 male and must be counterheld at maximum torque.

Datasheet, Specifications; Operation Manual, Installation
Control & I/O
Configure power-failure response for process safety

The Pilot ONE allows power-failure response as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles. Select the mode that matches the process risk.

Operation Manual, Normal Operation
Maintenance
Refrigerant work restricted to certified contractors

Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning contractors. The small A3 charge (0.16 kg) still requires leak-check and service procedures consistent with flammable refrigerant regulations.

Operation Manual, Maintenance; Datasheet, Specifications
Electrical
Country-specific plug must be specified at order

Single/two-phase devices (100–240 V) are supplied with a power cable and country-specific plug that must be specified when ordering. Three-phase devices below 63 A come without a plug. Missing specification delays commissioning.

Datasheet, Specifications

Selection & Integration Q&A

When should I choose the Unichiller P015 over the standard Unichiller 015 or the 015-H?

Choose the P015 when the process loop has high pressure drop: the 'P' model provides up to 2.5 bar and 25 l/min, while standard 015 variants target lower-resistance circuits. The 015-H adds a heater; the P015 is cooling-only. Select the P015 for restrictive cooling loops without heating demand.

What are the requirements for connecting the P015 to my process loop and electrical supply?

The pump connection is G3/4 male; hoses must be counterheld and torques not exceeded. Minimum filling capacity is 3.8 l. Electrical supply is 220–240 V, 1~/2~, 50 Hz, 8.5 A max, 16 A fuse, with a country-specific plug specified at order. Voltage/frequency tolerances must not combine unfavorably.

What installation environment and duty-cycle limits apply to the P015?

Install indoors at 5–40 °C ambient, at least 300 lx, below 2000 m altitude, with 50–200 cm wall clearance per side. IP20 and EMC Class A mean industrial environments only. A3 refrigerant requires room-volume assessment. Leveling feet or parking brakes must be engaged before operation.

What maintenance and accessories are required, and what is not included in base delivery?

Refrigerant circuit work requires approved contractors. Freeze protection requires glysantine or ethylene glycol; acetone and polyglycols are incompatible with the plastic pump. A country-specific plug must be specified at order. Hoses, expansion volume, and external Pt100 sensor are not included and must be planned separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P015
Datasheet ·168,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF