Peter Huber Kältemaschinenbau Unichiller 017T
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 017T

Peter Huber Kältemaschinenbau Unichiller 017T — Recirculating chillers with Noise level 64 dB.

The Unichiller 017T is an air-cooled, externally closed recirculating chiller rated -10 to 40 °C with a Pilot ONE controller and a fixed-speed pump delivering up to 25 l/min at 3 bar. It is the 1.7 kW-class modification in the Unichiller series, distinguished from the 015 siblings by its higher cooling capacity and 64 dB(A) noise level. It serves closed process loops requiring stable cooling down to -10 °C.

Noise level
64 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 017T Current 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Expansion volume 2 l
Maximum delivery flow rate 25 l/min (pressure pump max; also 3 bar)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -10...40 °C
Electrical, Power & Drive Specifications
Power supply 230V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and temperature

Installation, control and operation

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for airflow and service

The unit is air-cooled and rejects heat to the room; insufficient clearance reduces cooling capacity and can cause overheating. Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access, and the stated space requirements around the housing must be respected.

Unichiller Operation Manual, Installation; Unichiller 017T Technical Data, page 2
Electrical
Verify ±5% voltage with ±2% frequency before commissioning

The supply must be 230 V 1~ 50 Hz. Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed. Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

Unichiller 017T Technical Data, page 2
Process Limits
External circuit must be closed and matched to the pump curve

The chiller is designed only for externally closed applications with an atmospheric open expansion tank. The pump delivers 25 l/min max and 3 bar max, with flow falling to 5 l/min at 3 bar; the external loop pressure drop must be checked against this curve. Models with 'P' in the model name are the ones suited to high-pressure-drop applications, not the 017T.

Unichiller Operation Manual, Function Description; Unichiller 017T Technical Data, page 1
Cooling & Media
Size the heat load at the lowest operating temperature

Cooling capacity is 1.7 kW at 15 °C, 0.9 kW at 0 °C and 0.4 kW at -10 °C, rated at 20 °C room temperature. Capacity drops as ambient temperature rises, so the required heat load must be checked at the lowest set point and at the actual site ambient, not at the maximum rating.

Unichiller 017T Technical Data, page 1–2
Safety
Configure power-failure and Process Safety response before use

The Pilot ONE power-failure response can be set to OFF/Standby, ON or Power Failure Auto, determining whether the process restarts automatically after an outage. Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles, which matters for exothermic processes where cutting cooling would cause runaway temperatures.

Unichiller Operation Manual, Normal Operation and Safety
Maintenance
Refrigerant circuit work restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280, GWP 1397), so refrigerant handling and leak reporting obligations apply; the charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Maintenance and Installation
Environment
Indoor installation only, 5–40 °C ambient, not in ATEX zones

The unit is IP20 and must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude, with ambient temperature between 5 and 40 °C. It must not be installed or started up inside an ATEX zone because it is not explosion-protected. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, Installation; Unichiller 017T Technical Data, page 1

Selection & Integration Q&A

When should I choose the Unichiller 017T rather than the 015/015-H/015 OLÉ siblings?

Pick the 017T when the heat load requires the higher 1.7 kW at 15 °C capacity and 64 dB(A) is acceptable; the 015 siblings are 63 dB(A) but lower capacity. If the loop has high pressure drop, a 'P' model is required instead, since the 017T is not a high-pressure-drop variant.

What must the external hydraulic circuit and pump connections satisfy for the 017T?

The circuit must be externally closed with an atmospheric open expansion tank. Pump connections are G3/4 AG male; maximum tightening torques must not be exceeded and connections must be counterheld. Match loop pressure drop to the pump curve: 25 l/min max, 3 bar max, falling to 5 l/min at 3 bar.

What site and duty-cycle limits apply to installing the 017T?

Install indoors only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, never in an ATEX zone. Air-cooled housing needs 50–200 cm side clearance for maintenance. Leveling feet/parking brakes must be set before start-up, and cooling water supply (if applicable) shut off even overnight.

What maintenance and accessories are required beyond the base delivery?

Refrigerant circuit work must be done by approved refrigeration/air-conditioning contractors; R-449A (GWP 1397) handling and leak reporting apply. A country-specific plug must be specified at ordering. The external Pt100 sensor and process hoses are not part of the base unit and must be provided for the loop.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 017T Technical Data
Datasheet ·93,8 КБ ·2 pp.
View PDF