Peter Huber Kältemaschinenbau Unichiller P015w-H
Peter Huber Kältemaschinenbau Unichiller P015w-H — Recirculating chillers with Noise level 54 dB.
The Unichiller P015w-H is a water-cooled recirculating chiller with a plastic pump and Pilot ONE controller, rated for -20 to 100 °C. Its 'P' designation targets externally closed loops with high pressure drop, distinguishing it from the air-cooled Unichiller 015 siblings. It serves as a precision cooling source for laboratory and pilot-scale processes requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P015w-H Current | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 54 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 61 kg |
| Current consumption | 12 A (max.) |
| Fuse rating | 16 A |
| Power supply | 208-230V 1~/2~ 60Hz V, Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 36 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 48 л/час · at 15°C, flow 0°C: 60 l/h · at 15°C, flow 15°C: 68 l/h |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
- The P015w-H is a water-cooled variant within the Unichiller series; unlike the air-cooled Unichiller 015/015-H/015 OLÉ models (63 dB(A)), this unit is rated at 54 dB(A) and requires a facility cooling water loop.
- The 'P' in the model name denotes suitability for applications with high pressure drop, per the operation manual. The 'w' indicates water cooling; the 'H' suffix is not defined in the supplied facts.
- The refrigerant circuit uses R-449A (GWP 1397, charge 0.25 kg, 0.3 t CO2e) and is classified under PED Art. 4.3.
- The unit is intended only for externally closed applications and must not be installed in ATEX zones.
Hydraulic and Thermal Behavior
- Temperature range: -20 to 100 °C. The plastic pump delivers up to 25 l/min and 2.5 bar maximum pressure, dropping to 6 l/min at 2.0 bar; pump speed is not adjustable.
- Cooling water consumption depends on the required flow temperature: 60 l/h at 0 °C, 68 l/h at 15 °C, 48 l/h at -10 °C, and 36 l/h at -20 °C, all at 15 °C cooling water inlet.
- Published cooling capacities are valid only at 20 °C room temperature, 15 °C cooling water inlet, and 1 bar differential pressure; higher cooling water temperatures reduce capacity.
- The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM; only compatible water may be used. The plastic pump is not resistant to acetone or polyglycols; water with glysantine or ethylene glycol is recommended for freeze protection.
- Minimum filling capacity is 3.8 l; expansion volume is 1.7 l.
Installation, Control, and Operation
- Cooling water connection: G1/2 male; pump connection: G3/4 AG male. Cooling water supply must provide 1 bar minimum differential pressure and must not exceed 6 bar. Cooling water intake temperature must be between 5 °C and 20 °C; pipes must be insulated if below condensation point.
- Electrical: 208-230 V, 1~/2~, 60 Hz; maximum current 12 A; fuse rating 16 A; IP20; safety class III/FL. Voltage tolerance ±5 %, frequency ±2 %, but -5 % voltage with +2 % frequency is not allowed. Single/two-phase units are supplied with a country-specific plug (specify at ordering).
- Controller: Pilot ONE with Ethernet, USB (Host and Device), and RS232 interfaces; external Pt100 sensor connection. Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached.
- Installation: indoors only, illuminance ≥300 lx, altitude up to 2000 m, ambient 5–40 °C. Wall clearance 50–200 cm per side for maintenance. Leveling feet or parking brakes must be activated before operation. Electrical lines below the unit must be protected against condensate. Cooling water supply must be shut off even for short downtimes.
- Refrigerant work only by approved refrigeration/air-conditioning contractors.
Integration & Operating Notes
The facility cooling water loop must deliver at least 1 bar differential pressure for the chiller to perform, and must never exceed 6 bar or the unit may be damaged. Verify the supply pressure at the installation site before commissioning, and install a pressure reducer if the site pressure can exceed 6 bar.
The unit is designed for cooling water intake between 5 °C and 20 °C. If the facility water can fall below the condensation point, pipes must be insulated to prevent condensation. Published cooling capacities are referenced to 15 °C cooling water inlet; higher inlet temperatures reduce available cooling power, so derate capacity if site water is warmer.
Voltage tolerance is ±5 % and frequency tolerance is ±2 %, but the combination of -5 % voltage with +2 % frequency is explicitly not allowed. Check the site power quality against this asymmetric rule before connection; operation outside these limits may compromise safety and warranty.
Single/two-phase devices (100–240 V) are supplied with a power cable and country-specific plug; the plug type must be specified when ordering. Three-phase devices below 63 A come with cable but no plug, and the customer must provide the plug and final connection. This unit is 208–230 V 1~/2~, so confirm the plug requirement at purchase.
The unit must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude, with ambient temperature between 5 °C and 40 °C. A wall clearance of 50 to 200 cm per side is required for maintenance access. Leveling feet must be unscrewed or parking brakes applied before operation to prevent movement.
Because this is a water-cooled unit, condensation can form and drip. Electrical lines directly below the temperature control unit must be protected against liquid ingress. Route cables away from the unit or install drip protection to avoid electrical hazard.
The unit is designed only for externally closed applications and must not be installed or started up inside an ATEX zone because it is not explosion-protected. Ensure the process circuit is closed and non-hazardous. The 'P' model is particularly suitable for applications requiring high pressure drops.
The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, use water mixed with glysantine or ethylene glycol instead. Verify fluid compatibility with the cooling water circuit materials: copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM.
The Pilot ONE controller allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, consider Process Safety mode, which releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles. Choose the mode that matches the process risk.
The cooling water supply must be shut off even for shorter downtimes such as overnight to prevent water damage and corrosion during idle periods. Incorporate a manual or automatic shut-off valve in the cooling water supply line, and include this step in the site operating procedure.
Work on the coolant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with GWP 1397 and a charge of 0.25 kg (0.3 t CO2e), so F-gas reporting and leak-check obligations apply. Keep service records for compliance.
Selection & Integration Q&A
Choose the P015w-H when a facility cooling water loop is available and the process loop has high pressure drop, since the 'P' model is suited for that. Air-cooled 015 siblings avoid water connections but are louder (63 dB(A) vs 54 dB(A)) and lack the high-pressure-drop capability.
Cooling water: G1/2 male connection, 1–6 bar differential pressure, 5–20 °C inlet. Process loop: G3/4 AG male pump connection, minimum filling volume 3.8 l, expansion volume 1.7 l. Ensure the external circuit is closed and compatible with copper, stainless steel, PA, PPE, PTFE, and EPDM.
Install indoors only, at 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, with 50–200 cm side clearance for maintenance. Activate leveling feet or parking brakes before operation. Protect electrical lines below the unit from condensate. Do not install in ATEX zones.
Shut off cooling water even overnight. Refrigerant circuit work only by certified contractors; R-449A charge 0.25 kg requires F-gas records. The base delivery includes a power cable and country-specific plug for single/two-phase units (specify plug at ordering); three-phase units require customer-supplied plug. No process hoses or cooling water valves are mentioned as included.