Peter Huber Kältemaschinenbau Unichiller P022-H OLÉ
Peter Huber Kältemaschinenbau Unichiller P022-H OLÉ — Recirculating chillers with Noise level 65 dB.
The Unichiller P022-H OLÉ is a recirculating chiller with a 'P' designation for high-pressure-drop circuits, rated -10 to 100 °C with 2.2 kW cooling at 15 °C falling to 1 kW at -10 °C. Its OLÉ controller, 25 l/min pump and G3/4 male connections suit closed-loop process cooling where the external circuit imposes significant pressure loss.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P022-H OLÉ Current | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 65 dB(A) (sound pressure level) |
| Maximum delivery flow rate | 25 l/min (max pump delivery capacity) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| Temperature range | -10...100 °C |
| Weight | 83 kg |
| Current consumption | 15 A (max) |
| Fuse rating | 16 A |
| Power supply | 208-230V 1~/2~ 60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C (permissible ambient temperature) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The P022-H OLÉ is an air-cooled recirculating chiller in the Unichiller series, distinguished from the 015 siblings by the 'P' model designation, which the manual states makes it particularly suitable for applications requiring high pressure drops. It is intended only for externally closed circuits. The pump delivers up to 25 l/min and 2.5 bar, with the published curve giving 21 l/min at 0.5 bar down to 6 l/min at 2.0 bar, so the external loop's pressure loss must be checked against this curve rather than the free-delivery figure. The pump is type B, non-adjustable in speed, with G3/4 AG male connections.
Thermal behavior
- Operating range: -10 to 100 °C, covering sub-ambient cooling and moderate heating.
- Cooling capacity is set-point dependent: 2.2 kW at 15 °C, 1.6 kW at 0 °C, 1 kW at -10 °C. Size the load against the actual cold set point, not the 15 °C figure.
- Rising ambient temperature may reduce cooling capacity; ambient must remain within 5 to 40 °C.
- Overtemperature protection is adjustable per DIN 12876, safety class I/NFL. In Process Safety mode the controller releases maximum cooling capacity when the OT trigger is reached instead of switching off all poles.
Control, installation and operation
The OLÉ controller provides an OLED display with 0.1 K resolution, USB (Device) and RS232 interfaces. Power-failure response is configurable (OFF/Standby, ON, Power Failure Auto). The unit is IP20, Class A Group 1 per EN55011, and must be installed indoors at ≥300 lx and up to 2000 m altitude, outside any ATEX zone. Free-standing installation requires 50 to 200 cm wall clearance per side for maintenance. Leveling feet or caster brakes must be set before start-up. Refrigerant is R-449A, 0.56 kg, GWP 1397 (0.8 t CO2e); charge versus room volume must be checked at installation.
Integration & Operating Notes
The 208-230 V 1~/2~ 60 Hz feed must satisfy ±5% voltage and ±2% frequency together; the datasheet explicitly disallows a -5% voltage with +2% frequency combination. Max current is 15 A with a 16 A fuse, so breaker and cable sizing must cover the 15 A figure, and the country-specific plug must be specified at ordering.
The 'P' designation targets high-pressure-drop applications, but the pump still falls from 21 l/min at 0.5 bar to 6 l/min at 2.0 bar against a 2.5 bar maximum. Calculate the external circuit's pressure loss at the required flow before committing; a loop that demands more than the curve allows at the set point will under-deliver coolant and distort the process temperature.
The pump is not resistant to acetone or polyglycols (manufacturer-dependent), so these fluids must be avoided or a more resistant pump specified. For sub-zero set points, mix water with glysantine or ethylene glycol for freeze protection rather than running pure water. Minimum system filling capacity is 3.8 l with a 1.7 l expansion vessel.
The unit is not explosion-protected and must not be installed or started in an ATEX zone. It is for indoor use only, at ≥300 lx illuminance and up to 2000 m altitude, with 50 to 200 cm wall clearance per side for maintenance. IP20 means a dry, protected location; electrical lines directly below a water-cooled unit must be shielded from condensate.
Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration and air-conditioning system contractors; the R-449A charge of 0.56 kg (GWP 1397, 0.8 t CO2e) also triggers room-volume checks at installation. Shut off cooling water supply even for overnight downtimes, and set leveling feet or caster parking brakes before start-up to prevent movement.
Selection & Integration Q&A
Pick the P022-H when the external circuit has high pressure drop (the 'P' designation) and the load needs up to 2.2 kW at 15 °C down to 1 kW at -10 °C. The 015 siblings are lower-capacity; choose them only if the loop's pressure loss and heat load fit their smaller envelope.
Check the loop's pressure loss against the pump curve (21 l/min at 0.5 bar to 6 l/min at 2.0 bar, 2.5 bar max) and provide at least 3.8 l system fill with a 1.7 l expansion vessel. The feed must be 208-230 V 1~/2~ 60 Hz, 15 A max, 16 A fuse, with ±5% V and ±2% f met simultaneously.
Indoors only, outside any ATEX zone, at ≥300 lx and up to 2000 m altitude, with ambient 5-40 °C. Allow 50-200 cm wall clearance per side for maintenance, keep it dry (IP20), and set leveling feet or caster brakes before start-up.
Refrigerant circuit work requires approved refrigeration contractors; the R-449A charge (0.56 kg, 0.8 t CO2e) needs a room-volume check. The power cable and country-specific plug must be specified at ordering. For sub-zero duty, plan glysantine or ethylene glycol freeze protection; acetone and polyglycols are excluded by the plastic pump.