Peter Huber Kältemaschinenbau Unichiller P025
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P025

Peter Huber Kältemaschinenbau Unichiller P025 — Recirculating chillers with Noise level 64 dB.

The Unichiller P025 is a single-phase recirculating chiller with a 'P' designation for high-pressure-drop loops, rated -10 to 40 °C and 2.5 kW at 15 °C. Its non-adjustable pump delivers 25 l/min max and 6 l/min at 2.0 bar, distinguishing it from lower-capacity siblings. It serves externally closed cooling circuits requiring G3/4 connections and Pilot ONE control.

Noise level
64 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P025 Current 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Maximum delivery flow rate 25 l/min (also 2, 5 bar pump pressure)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -10...40 °C
Dimensions, Weight & Mechanical Design
Weight 82 kg
Electrical, Power & Drive Specifications
Current consumption 9 A
Fuse rating 16 A
Power supply 220-240V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The Unichiller P025 sits in the Unichiller Bis 25 kW series alongside 015 variants (015, 015-H, 015-H OLÉ, 015 OLÉ). The 'P' in the model name marks it as particularly suitable for applications requiring high pressure drops, per the operation manual. It is a free-standing, indoor-only unit (IP20) with a plastic pump (type B, non-adjustable speed) and G3/4 male pump connections. The refrigerant is R-290 (A3, H220) with a 0.22 kg charge and GWP 0.02. The controller is Pilot ONE with Ethernet, USB host/device, and RS232 interfaces; an external Pt100 sensor can be connected.

Behavior under load and media

Cooling capacity is strongly temperature-dependent: 2.5 kW at 15 °C, 2.0 kW at 0 °C, and 1.2 kW at -10 °C, so sizing must use the capacity at the actual operating point. The pump curve gives 25 l/min max and 6 l/min at 2.0 bar; system pressure drop must be checked against this curve. The unit is intended only for externally closed applications. The plastic pump is not resistant to acetone or polyglycols. For freeze protection below 0 °C, water should be mixed with glysantine or ethylene glycol. Ambient temperature must stay between 5 and 40 °C, and cooling capacity may drop above 20 °C.

Installation, control and operation

Electrical supply is 220–240 V 1~ 50 Hz with 9 A max current and a 16 A fuse; voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed. The unit must not be installed in an ATEX zone. A wall clearance of 50 to 200 cm per side is required for maintenance. Leveling feet must be activated and parking brakes applied before start-up. The Pilot ONE allows configuration of power-failure response (OFF/Standby, ON, or Power Failure Auto) and a Process Safety mode that releases maximum cooling capacity when the overtemperature trigger is reached. Refrigerant circuit work must be performed by approved refrigeration/air-conditioning contractors.

Integration & Operating Notes

Process Limits
Cooling capacity falls by half from 15 °C to -10 °C

Sizing must use the capacity at the actual operating temperature: 2.5 kW at 15 °C, 2.0 kW at 0 °C, and 1.2 kW at -10 °C. Selecting based on the 15 °C figure will undersize the unit for low-temperature duty and risk insufficient cooling.

Unichiller P025 Datasheet, Technical Specifications
Process Limits
Plastic pump incompatible with acetone and polyglycols

The standard plastic pump (type B) is not resistant to acetone or polyglycols. If the process fluid contains either, a more resistant pump must be ordered or the fluid changed; otherwise premature pump failure is likely.

Unichiller P025 Datasheet, page 2
Process Limits
Freeze protection required below 0 °C

For operation near or below 0 °C, water should be mixed with glysantine or ethylene glycol to prevent freezing and damage. Pure water must not be used at low temperatures.

Unichiller P025 Datasheet, page 2
Mounting
Not permitted in ATEX zones

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Site classification must be checked before installation; flammable refrigerant R-290 (A3) adds further requirements.

Unichiller Operation Manual, page 13
Mounting
Maintain 50–200 cm wall clearance per side

A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit must be indoors with illuminance at least 300 lx and at altitude up to 2000 m.

Unichiller Operation Manual, page 25
Electrical
Voltage tolerance ±5% only with frequency ±2%

The supply must be 220–240 V 1~ 50 Hz with 9 A max current and a 16 A fuse. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed; verify site supply quality before connection.

Unichiller P025 Datasheet, page 2
Control & I/O
Configure power-failure response for process safety

The Pilot ONE allows power-failure response as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant work only by approved contractors

Work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-290 charge is 0.22 kg with GWP 0.02; leak reporting and environmental compliance are simplified but not eliminated.

Unichiller Operation Manual, page 17
Maintenance
Shut off cooling water even overnight

Cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the P025 over the 015 siblings?

Choose the P025 when the process loop has high pressure drop (the 'P' designation) and requires up to 2.5 kW cooling at 15 °C. The 015 variants offer lower capacity and are not specifically suited to high-pressure-drop applications.

What mating hardware and fluid volume does the P025 require?

Pump connections are G3/4 male, so hoses and fittings must be G3/4 female. Minimum filling capacity is 3.8 l with 1.7 l expansion volume; the loop must be filled accordingly and expansion space respected.

What maintenance and accessories are not included in base delivery?

The base unit does not include freeze-protection additives (glysantine or ethylene glycol) or G3/4 female hoses. Refrigerant circuit work requires approved contractors. Cooling water supply must be shut off even overnight.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P025 Datasheet
Datasheet ·30,6 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.