Peter Huber Kältemaschinenbau Unichiller 017T-H
Peter Huber Kältemaschinenbau Unichiller 017T-H — Recirculating chillers with Noise level 64 dB.
The Unichiller 017T-H is an air-cooled, single-phase recirculating chiller with a 2 kW heater and a -10 to 100 °C working range, controlled by the Pilot ONE. It is the 15.5 A / 16 A-fuse variant of the Unichiller series, distinguished from the 015 siblings by its higher current draw and 64 dB(A) noise level. It serves externally closed cooling circuits in industrial and laboratory processes.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 017T-H Current | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 64 dB(A) |
| Expansion volume | 2 l |
| Heating capacity | 2 kW |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 152 kg |
| Fuse rating | 16 A |
| Power supply | 230V 1~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and positioning
- Air-cooled condenser rejecting heat to the room; refrigerant R-449A (A1, H280, GWP 1397).
- Single-phase 230 V 1~ 50 Hz supply, 16 A fuse, maximum current 15.5 A.
- Pump type B with G3/4 AG male connections, non-adjustable speed; 25 l/min maximum, 5 l/min at 3.0 bar.
- 2 l expansion volume, minimum filling capacity 2.5 l.
- IP20, safety class I / NFL, 152 kg, 64 dB(A).
- Within the series, the 017T-H sits above the 015/015-H/015 OLÉ siblings (63 dB(A)); the 'P' suffix models are reserved for high-pressure-drop applications and are not this unit.
Behavior under load and media
- Cooling capacity is rated at 20 °C room temperature and falls at lower setpoints: 0.4 kW at -10 °C.
- Heating capacity 2 kW; temperature range -10 to 100 °C.
- Permitted ambient 5 to 40 °C; capacity may drop as ambient rises.
- Intended only for externally closed applications; external loop pressure drop must be checked against the pump curve (5 l/min at 3 bar).
Installation, control and operation
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor input.
- Power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Process Safety mode releases maximum cooling when the overtemperature trigger is reached.
- Leveling feet/parking brakes must be set before start-up; cooling water supply (where applicable) shut off during downtimes.
- Wall clearance 50–200 cm per side for maintenance; indoor use only, illuminance ≥300 lx, altitude up to 2000 m.
Integration & Operating Notes
The unit draws up to 15.5 A on a 230 V single-phase supply protected by a 16 A fuse, leaving minimal margin. The site circuit must be dedicated and rated for continuous load; shared sockets or long extension runs risk nuisance tripping.
The datasheet explicitly prohibits a -5% voltage combined with +2% frequency. Verify the site supply quality against the permitted tolerance envelope before commissioning; an out-of-tolerance combination can stress the compressor and electronics.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side. The air-cooled condenser also needs unobstructed airflow; insufficient clearance or high ambient temperature will reduce cooling capacity below the rated 20 °C figures.
The unit is designed for externally closed applications. The pump delivers 25 l/min maximum but only 5 l/min at 3.0 bar, so a restrictive external loop will reduce flow and heat transfer. Size the circuit and fluid charge (minimum 2.5 l, 2 l expansion) accordingly.
Published capacity is valid at 20 °C room temperature and drops at lower setpoints. Size the process heat load against the capacity at the lowest required temperature, not the nominal rating, and derate further if ambient exceeds 20 °C.
The Pilot ONE allows OFF/Standby, ON or Power Failure Auto after a mains interruption. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles; select the mode that matches the process risk.
IP20 offers no water protection, so washdown or damp areas are excluded. The unit is Class A, Group 1 for electromagnetic disturbance and may interfere with sensitive laboratory equipment; place it accordingly. Ambient must stay within 5–40 °C.
The R-449A charge (1.2 kg, GWP 1397, H280) requires leak checks and handling by approved refrigeration/air-conditioning contractors. Shut off cooling water supply even for overnight downtimes where a water-cooled configuration applies, and protect electrical lines below the unit from condensate.
Selection & Integration Q&A
The 017T-H draws up to 15.5 A on a 16 A fuse and is rated 64 dB(A), versus 63 dB(A) for the 015 family. Choose it when the process heat load requires the higher-capacity frame; choose a 015 variant when a lower current draw and slightly quieter operation suffice. The 'P' models are for high-pressure-drop loops and are not this unit.
A dedicated 230 V 1~ 50 Hz circuit with a 16 A fuse, since maximum current is 15.5 A. Voltage and frequency tolerances must be met simultaneously; -5% voltage with +2% frequency is not allowed. A country-specific plug must be specified at order time.
Indoor use only, illuminance ≥300 lx, altitude up to 2000 m, ambient 5–40 °C. Free-standing installation needs 50–200 cm wall clearance per side for maintenance and airflow. Leveling feet or parking brakes must be set before start-up; IP20 rules out damp or washdown areas.
Refrigerant R-449A (1.2 kg, GWP 1397) service must be by approved refrigeration contractors; leak checks apply. Minimum fluid charge is 2.5 l with 2 l expansion volume. The power cable and country-specific plug must be specified at order; heat transfer fluid and external circuit hardware are not part of the base unit.