Peter Huber Kältemaschinenbau Unichiller 017Tw-H
Peter Huber Kältemaschinenbau Unichiller 017Tw-H — Recirculating chillers with Noise level 64 dB.
The Unichiller 017Tw-H is a water-cooled recirculating chiller with a Pilot ONE controller, rated for -10 to 100 °C and 208 V 2~ 60 Hz. It is the 'Tw' (water-cooled) modification of the 017 platform, distinguished from the air-cooled 015 siblings by its G1/2 cooling-water circuit. It serves externally closed process loops requiring heat rejection to a facility cooling-water supply.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 017Tw-H Current | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 64 dB(A) |
| Expansion volume | 4 l |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Power supply | 208V 2~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 25 л/час (at 15°C, flow -10°C) at 15°C, flow 0°C: 40 l/h · at 15°C, flow 15°C: 53 l/h |
| Cooling water differential pressure | 1 bar |
| Cooling Water Pressure | 6 bar |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (Tw) variant of the Unichiller 017 platform; heat is rejected to a facility cooling-water loop via G1/2 male connections rather than to room air.
- Sibling models in the series (Unichiller 015, 015-H, 015-H OLÉ, 015 OLÉ) are air-cooled units at 63 dB(A); this model is rated 64 dB(A).
- Pump connection G3/4 AG male; pump type B, fixed speed (not adjustable).
- Internal circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; expansion volume 4 l, minimum filling capacity 2.5 l.
- Safety class III / FL per DIN 12876; protection rating IP20.
Working range and behavior
- Temperature range -10 to 100 °C; ambient rating 5 to 40 °C.
- Cooling-water consumption depends on the process flow temperature at 15 °C water: 25 l/h at -10 °C flow, 40 l/h at 0 °C flow, 53 l/h at 15 °C flow.
- Cooling capacity falls and cooling-water flow demand rises as the cooling-water intake temperature increases; intake must be 5–20 °C.
- No digitized performance curves were supplied; capacity at intermediate points must be taken from the datasheet.
Control, installation and operation
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching all poles off.
- Intended only for externally closed applications; not explosion-protected, so not for ATEX zones.
- Cooling-water supply must be shut off even for overnight downtimes.
Integration & Operating Notes
The manual restricts this chiller to externally closed applications, so the hydraulic circuit must be designed as a sealed loop with the process. Open-tank or atmospheric bath arrangements are outside the intended use and invalidate the thermal-fluid and safety-class assumptions.
The facility loop must deliver intake water between 5 and 20 °C with at least 1 bar differential pressure and no more than 6 bar at the G1/2 connection. Below 1 bar the unit cannot reject rated capacity; above 6 bar the cooling-water circuit is overstressed. Insulate pipes where water can fall below dew point.
Even short downtimes such as overnight require closing the cooling-water supply. Leaving it open invites condensation on the water-cooled assembly and corrosion or water damage in the enclosure, and the manual treats this as a standing operating rule rather than an optional measure.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to permit maintenance access. The unit is indoor-only, needs at least 300 lx illuminance and must be sited below 2000 m altitude; check refrigerant charge against room volume to stay within the limit per m³ of ambient air.
Because the unit is water-cooled, condensation can drip from the cooling-water circuit. Electrical lines routed directly below the chiller must be protected against liquid ingress, and the IP20 enclosure itself must be kept clear of water and dust. Plan cable trays accordingly.
The ±5% voltage tolerance applies only together with a ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are not permitted. Verify the site supply against this coupled window before connection; three-phase units under 63 A arrive with a cable but no plug, so a suitable plug must be provided on site.
The Pilot ONE can be set to OFF/Standby, ON or Power Failure Auto after a mains loss. For exothermic processes, Process Safety mode instead releases maximum cooling when the overtemperature trigger is reached rather than switching all poles off. Decide the mode during commissioning, since it defines restart behavior of the process loop.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning contractors. The refrigerant is R-449A (A1, H280, GWP 1397), so handling and ventilation precautions apply and service planning must account for certified personnel and refrigerant reporting obligations.
Selection & Integration Q&A
Pick the 017Tw-H when a facility cooling-water loop is available and room heat rejection is undesirable; it needs 5–20 °C water at 1–6 bar differential. Choose the 015 air-cooled siblings where no water loop exists, accepting their 63 dB(A) fan noise versus 64 dB(A) here.
Cooling water at 5–20 °C, minimum 1 bar differential and maximum 6 bar at G1/2 male connections; flow demand is 25–53 l/h depending on flow temperature. Process side uses G3/4 AG male pump connections with matching couplings and counterheld tightening.
Indoor use only, 5–40 °C ambient, at least 300 lx, below 2000 m, 50–200 cm side clearance, no ATEX zone. Leveling feet or parking brakes must be set before start-up, and cooling water must be shut off even overnight.
Refrigerant-circuit service only by approved contractors (R-449A, GWP 1397). Three-phase units under 63 A ship with cable but no plug, so a site plug is required; external Pt100 sensor and hose couplings for G3/4 and G1/2 are not included and must be sourced separately.