Peter Huber Kältemaschinenbau Unichiller 020Tw
Peter Huber Kältemaschinenbau Unichiller 020Tw — Recirculating chillers with Noise level 60 dB.
The Unichiller 020Tw is a water-cooled recirculating chiller with a 4 l expansion volume and a fixed-speed pump delivering up to 25 l/min at 3 bar, operating from -20 to 40 °C. It is designed for externally closed circuits and requires a cooling water supply of 1–6 bar differential pressure. Its Pilot ONE controller and IP20 rating suit indoor industrial cooling tasks.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 020Tw Current | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 60 dB(A) |
| Expansion volume | 4 l |
| Maximum delivery flow rate | 25 l/min (pressure pump max.) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 133 kg |
| Current consumption | 11 A (max.) |
| Fuse rating | 16 A |
| Power supply | 230V 1~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 90 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The 020Tw is a water-cooled (not air-cooled) recirculating chiller; heat is rejected to a cooling water circuit via G1/2 male connections.
- Within the Unichiller series, this model is distinguished by its water-cooled condenser and 4 l expansion volume; sibling models 015/015-H/015-H OLÉ/015 OLÉ are air-cooled variants with higher noise (63 dB(A) vs 60 dB(A) here).
- The pump is type B, non-adjustable speed, with G3/4 AG male process connections; maximum delivery is 25 l/min at 3 bar.
- The refrigerant is R-449A (A1, H280), charge 0.4 kg (0.6 t CO2e); work on the refrigerant circuit must be performed by approved refrigeration contractors.
Behavior under load and media
- Temperature range is -20 to 40 °C; the unit is intended only for externally closed applications (manual, p. 34).
- Cooling water intake temperature must be between 5 and 20 °C; higher intake temperatures reduce cooling power and may require increased cooling water flow (datasheet, p. 2).
- Cooling water consumption is 90 l/h at 15 °C intake and 0 °C flow; minimum differential pressure 1 bar, maximum 6 bar.
- The cooling water circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) dictate compatible water quality; condensation may form and pipes must be insulated if it occurs.
- Minimum filling capacity of the external circuit is 2.5 l; expansion volume is 4 l.
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces; an external Pt100 sensor can be connected.
- Power supply is 230 V 1~ 50 Hz; maximum current 11 A, fuse 16 A; voltage tolerance ±5% with simultaneous frequency tolerance ±2% (certain combinations not allowed). Single-phase devices are supplied with a country-specific plug (specify when ordering).
- The unit is IP20 and must be used indoors at ≥300 lx illuminance and up to 2000 m altitude; ambient temperature 5–40 °C.
- A wall clearance of 50–200 cm per side is required for maintenance on free-standing models; leveling feet must be extended and parking brakes applied before start-up.
- The power-failure response is configurable (OFF/Standby, ON, Power Failure Auto); Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached.
- Cooling water supply must be shut off even for short downtimes (e.g., overnight). Electrical lines directly below the unit must be protected against liquid ingress.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access to the refrigerant circuit and components. Insufficient clearance can prevent servicing and may cause overheating or restricted airflow around the unit.
Because the chiller is water-cooled, condensation can form on the cooling water pipes and drip onto electrical lines routed directly below. Such lines must be protected against liquid ingress to avoid short circuits or shock hazard.
The unit is designed for cooling water intake temperatures of 5 to 20 °C. Higher intake temperatures reduce cooling capacity and may require increased cooling water flow; lower temperatures can cause condensation and must be managed by insulating pipes.
The cooling water circuit must be supplied with a minimum differential pressure of 1 bar and must never exceed 6 bar. Overpressure can damage the condenser or connections; ensure a pressure regulator or relief is installed if site pressure is higher.
The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280) with a charge of 0.4 kg (0.6 t CO2e); leak checks and handling must follow applicable regulations.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. This also avoids unnecessary water consumption and potential condensation when the unit is not running.
Selection & Integration Q&A
Choose the 020Tw when you need water-cooled heat rejection (no air-cooled condenser) and a 4 l expansion volume; it is quieter (60 dB(A)) than the 015 siblings (63 dB(A)). The 015 models are air-cooled and may suit sites without cooling water supply.
Provide cooling water at 5–20 °C, 1–6 bar differential pressure, and 90 l/h at 15 °C intake. Circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so use compatible water; insulate pipes if condensation occurs.
Install indoors only, at ≥300 lx and ≤2000 m altitude, ambient 5–40 °C, IP20. Maintain 50–200 cm side clearance for maintenance. Extend leveling feet and apply parking brakes before start-up. Shut off cooling water even for overnight downtime.
The base unit includes a power cable with country-specific plug (specify when ordering). Refrigerant circuit work must be by certified contractors; no routine maintenance intervals are specified in the digest. External Pt100 sensor and process hoses are not included and must be sourced separately.