Peter Huber Kältemaschinenbau Unichiller 020Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 020Tw-H

Peter Huber Kältemaschinenbau Unichiller 020Tw-H — Recirculating chillers with Noise level 60 dB.

Water-cooled recirculating chiller with a 2 kW heater and a -20 to 100 °C working range, built for externally closed loops. The 'Tw' modification uses a cooling-water circuit (G1/2 male, 1–6 bar differential) instead of an air-cooled condenser, distinguishing it from the 015 siblings. It serves as a temperature-control source for processes requiring both heating and cooling capacity.

Noise level
60 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 020Tw-H Current 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Expansion volume 4 l
Heating capacity 2 kW
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 141 kg
Electrical, Power & Drive Specifications
Current consumption 16 A (max.)
Fuse rating 16 A
Power supply 230V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 90 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Control, installation and operation

Integration & Operating Notes

Process Limits
Closed-loop process only; not for open baths

The manual states the unit is designed for externally closed applications. Integrating it with an open tank or unsealed circuit would violate the intended hydraulic configuration and is outside the documented application.

Unichiller Operation Manual, Function Description, p. 34
Cooling & Media
Cooling water 5–20 °C, 1–6 bar differential

Supply water below 5 °C or above 20 °C falls outside the design intake range; below the condensation point, pipes must be insulated. Exceeding 6 bar risks damage to the cooling-water circuit, and less than 1 bar differential pressure will not sustain the required flow.

Unichiller 020Tw-H Datasheet, Technical Data, p. 1–2
Electrical
Voltage/frequency tolerance combinations are restricted

A ±5% voltage tolerance is permitted only together with ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are explicitly not allowed. Verify the site supply before connection; the 16 A fuse rating and 16 A max current define the branch circuit.

Unichiller 020Tw-H Datasheet, Technical Data, p. 2
Mounting
50–200 cm side clearance and indoor siting required

Free-standing installation needs 50 to 200 cm wall clearance on either side for maintenance access. The unit is indoor-only, requires at least 300 lx illuminance, and must not be installed in an ATEX zone because it is not explosion-protected.

Unichiller Operation Manual, Commissioning, p. 25 and p. 13
Mounting
Counterhold fittings and respect maximum tightening torques

Pump and cooling-water connections must be counterheld during hose installation, and the maximum tightening torques in the manual table must not be exceeded. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, Installation, p. 28
Operation
Shut off cooling water even for overnight stops

The cooling water supply must be closed for any downtime, including short overnight periods, to prevent water damage and corrosion in the idle circuit. Leveling feet must be activated and parking brakes applied before the unit is switched on.

Unichiller Operation Manual, Installation/Operation, p. 28 and p. 31
Control & I/O
Configure power-failure response before unattended runs

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For processes that must resume automatically after a mains interruption, Power Failure Auto must be selected; otherwise the loop remains stopped after power returns.

Unichiller Operation Manual, Normal Operation, p. 19
Safety
Process Safety mode applies full cooling at OT trigger

In Process Safety mode the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This behavior must be matched to the exothermic risk of the process and the external Pt100 sensor placement.

Unichiller Operation Manual, Normal Operation, p. 18
Maintenance
Refrigerant work restricted to approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge (GWP 1397, 0.6 t CO2 equivalent) and room volume must be checked at installation against the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, Maintenance, p. 17 and p. 15

Selection & Integration Q&A

When should I choose the 020Tw-H instead of the air-cooled Unichiller 015 siblings?

Choose the 020Tw-H when a cooling-water supply (5–20 °C, 1–6 bar differential, 90 l/h at 15 °C) is available and condenser heat must be rejected to water rather than room air; the 015 siblings are air-cooled and rated 63 dB(A) versus 60 dB(A) here.

What cooling-water supply and fittings must be provided for this chiller?

Provide cooling water at 5–20 °C with minimum 1 bar differential and maximum 6 bar, G1/2 male connections, and 90 l/h at 15 °C water/0 °C flow. The process loop connects to the G3/4 AG male pump outlet; use compatible materials (copper, 1.4401, EPDM, PTFE, PA, PPE).

What installation and duty-cycle limits does the manual impose?

Indoor use only, 5–40 °C ambient, ≥300 lx, up to 2000 m altitude, 50–200 cm side clearance, no ATEX zone. Leveling feet/parking brakes engaged before start-up; cooling water shut off even overnight; electrical lines below the unit protected from condensate.

What maintenance and delivery items must be planned beyond the base unit?

Refrigerant-circuit work requires approved contractors; the R-449A charge and room volume must be checked at installation. The country-specific power plug must be specified at order. Minimum filling is 2.5 l with 4 l expansion volume; no cables or fittings beyond the supplied power cable are documented.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 020Tw-H Datasheet
Datasheet ·63,8 КБ ·2 pp.
View PDF