Peter Huber Kältemaschinenbau Unichiller 015
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 015

Peter Huber Kältemaschinenbau Unichiller 015 — Recirculating chillers with Noise level 63 dB.

The Unichiller 015 is a compact air-cooled recirculating chiller with an immersion pump and Pilot ONE controller, rated 1 kW at 0 °C over a -20…40 °C range. It is the base model of the Unichiller 015 series, distinguished from the -H, OLÉ and w siblings by its standard configuration and 63 dB(A) noise level. It serves externally closed industrial process cooling loops.

Noise level
63 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 015 Current 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 63 dB(A)
Pump connection G3/4 AG male
Pump speed adjustable Yes
Pump type Immersion pump
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 56 kg
Electrical, Power & Drive Specifications
Current consumption 8 A (max.)
Fuse rating 16 A
Power supply 220-240V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 1 kW (at 0 °C)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The Unichiller 015 is an air-cooled, single-stage recirculating chiller using R-290 (A3, H220) refrigerant with a charge of 0.16 kg. It is built for externally closed applications only. Within the Unichiller 015 family, this base model shares the 63 dB(A) noise level with the -H, -H OLÉ and OLÉ variants, while the 015w sibling is quieter at 54 dB(A). The unit uses a plastic immersion pump with adjustable speed and G3/4 AG male connections; this pump is not resistant to acetone or polyglycols. Cooling is via an air-cooled condenser, so capacity depends on ambient temperature and airflow.

Thermal and hydraulic behavior

Cooling capacity is strongly temperature-dependent: 1.5 kW at +15 °C, 1 kW at 0 °C and 0.3 kW at -20 °C. The process load must therefore be checked at the actual working temperature, not at nominal capacity. The immersion pump delivers a maximum of 29 l/min at 1 bar and 21 l/min at 0.5 bar, so the pressure drop of the external circuit must be compared against this curve. The temperature range is -20…40 °C. For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection. Minimum filling capacity is 3.8 l with 1.7 l expansion volume; the external loop volume must be included when charging.

Electrical, control and installation

Electrical supply is 220-240 V single-phase 50 Hz with 8 A maximum current and a 16 A fuse. Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in opposite directions. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, a Pt100 external sensor connection, and configurable power-failure response (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. The unit is IP20 and must be installed indoors at 5…40 °C ambient with at least 300 lx illuminance and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance. It must not be installed in an ATEX zone. Leveling feet must be activated and parking brakes applied before startup. Electrical lines directly below the unit must be protected against liquid ingress. EMC classification is Class A, Group 1, intended for industrial environments.

Integration & Operating Notes

Process Limits
Cooling capacity must be checked at actual working temperature

Capacity falls from 1.5 kW at +15 °C to 0.3 kW at -20 °C, so sizing on the 1 kW at 0 °C figure will undersize the unit for low-temperature duty. Verify the process heat load at the lowest intended set point before selection.

Unichiller 015 Technical Data, Specifications
Cooling & Media
Plastic pump incompatible with acetone and polyglycols

The standard immersion pump is not resistant to acetone or polyglycols. These media require a more resistant pump option or must be avoided entirely to prevent chemical attack and pump failure.

Unichiller 015 Technical Data, Specifications
Cooling & Media
Freeze protection required for sub-zero set points

For operation below 0 °C, water should be mixed with glysantine or ethylene glycol rather than used pure, to avoid freezing and damage in the cooling circuit. Minimum filling capacity is 3.8 l with 1.7 l expansion volume.

Unichiller 015 Technical Data, Specifications
Electrical
Supply tolerances must not be combined in opposite directions

Voltage tolerance is ±5 % and frequency tolerance ±2 %, but a -5 % voltage combined with +2 % frequency is not allowed. Verify supply quality at the installation site before commissioning to stay within the permitted envelope.

Unichiller 015 Technical Data, Specifications
Mounting
Unit must not be installed in an ATEX zone

The chiller is not explosion-protected and uses R-290 (A3 flammable) refrigerant. The installation location must be assessed for flammable-refrigerant safety and ventilation, and refrigerant charge versus room size must be checked against the limit per m³ of ambient air.

Unichiller Operation Manual, Section: Installation
Mounting
Wall clearance of 50 to 200 cm per side required

Free-standing models need this clearance on either side for maintenance access. Insufficient clearance also degrades air-cooled condenser performance and may cause overtemperature trips. The unit must be used indoors at 5…40 °C ambient with at least 300 lx illuminance.

Unichiller Operation Manual, Section: Installation
Mounting
Leveling feet and parking brakes must be set before startup

Parking brakes must be activated at the casters (if any) and/or leveling feet unscrewed/activated before the unit is put into operation, to prevent movement and crushing injuries during operation.

Unichiller Operation Manual, Section: Commissioning
Control & I/O
Power-failure response must be configured for the process

The Pilot ONE allows OFF/Standby, ON, or Power Failure Auto. The choice determines whether the process restarts automatically after a power outage, which is critical for unattended or exothermic processes.

Unichiller Operation Manual, Section: Normal Operation
Control & I/O
Process Safety mode releases maximum cooling on OT trigger

In Process Safety mode, maximum cooling capacity is released when the overtemperature trigger value is detected, instead of switching off all poles. This is critical for exothermic processes where cutting cooling would cause runaway temperatures.

Unichiller Operation Manual, Section: Safety
Maintenance
Refrigerant circuit work only by approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant charge is 0.16 kg with GWP 0.02, so leak-check and record-keeping obligations are minimal but still apply.

Unichiller Operation Manual, Section: Maintenance

Selection & Integration Q&A

When should I choose the Unichiller 015 base model instead of the -H, OLÉ or 015w siblings?

Choose the base 015 for standard externally closed cooling where 63 dB(A) is acceptable. The 015w is quieter at 54 dB(A) for noise-sensitive sites. The -H and OLÉ variants share the 63 dB(A) level; select them only when their specific configuration is required, as they do not differ in noise from the base model.

What supply and circuit protection does the Unichiller 015 require at the installation site?

A 220-240 V single-phase 50 Hz supply with 8 A maximum current and a 16 A fuse. Voltage tolerance is ±5 % and frequency ±2 %, but these must not be combined in opposite directions (e.g. -5 % voltage with +2 % frequency is not allowed). Verify supply quality before commissioning.

Can the Unichiller 015 be installed in a hazardous area or outdoors?

No. It must not be installed or started up inside an ATEX zone because it is not explosion-protected, and it uses R-290 (A3 flammable) refrigerant. It is for indoor use only at 5…40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50 to 200 cm wall clearance per side.

What maintenance and service requirements apply, and what is not included in base delivery?

Refrigerant circuit work must be performed only by approved refrigeration/air-conditioning contractors. The refrigerant charge is 0.16 kg (GWP 0.02). Heat transfer fluid is not included; minimum filling capacity is 3.8 l with 1.7 l expansion volume, and freeze protection (glysantine or ethylene glycol) must be added for sub-zero operation. Leveling feet and parking brakes must be set before startup.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 015 Technical Data
Datasheet ·167,9 КБ ·2 pp.
View PDF